Convergence of Dissipation and Impedance Analysis of Quartz Crystal Microbalance Studies

被引:34
|
作者
Zhang, Yaozhong [2 ]
Du, Binyang [1 ]
Chen, Xinan [2 ]
Ma, Hongwei [2 ]
机构
[1] Zhejiang Univ, Dept Polymer Sci & Engn, Minist Educ, Key Lab Macromol Synth & Functionalizat, Hangzhou 310027, Peoples R China
[2] Peking Univ, Coll Engn, Dept Biomed Engn, Beijing 100871, Peoples R China
关键词
TRANSFER RADICAL POLYMERIZATION; SURFACE-PLASMON RESONANCE; SHEAR MODE RESONATORS; VISCOELASTIC PROPERTIES; POLY(N-ISOPROPYLACRYLAMIDE) CHAINS; EQUIVALENT-CIRCUIT; THIN-FILMS; THICKNESS; BRUSHES; ADSORPTION;
D O I
10.1021/ac8019762
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A quartz crystal microbalance (QCM) consists of a resonator, which measures the resonance frequency of the quartz slab. When coupled with a network analyzer or coupled with impulse excitation technology, QCM gives additional impedance or dissipation information, respectively. Ibis report provides a set of equations that bring the QCM community a convergence of the dissipation and impedance analysis. Equations derived from the complex frequency shift were applied to quantitatively analyze the dissipation data of polymer brushes obtained from QCM-D. The obtained viscoelastic properties of polymer brushes were then compared with those obtained by the Voigt model method. We believe that these equations will be useful in quantitative studies of interfacial phenomena accompanied with mass or viscoelasticity changes.
引用
收藏
页码:642 / 648
页数:7
相关论文
共 50 条
  • [1] Quartz Crystal Microbalance with Dissipation
    不详
    TENSIDE SURFACTANTS DETERGENTS, 2009, 46 (02) : 121 - 121
  • [2] A nanocell for quartz crystal microbalance and quartz crystal microbalance with dissipation-monitoring sensing
    Ohlsson, Gabriel
    Langhammer, Christoph
    Zoric, Igor
    Kasemo, Bengt
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (08):
  • [3] Impedance characterization of a quartz crystal microbalance
    Vanysek, P
    Delia, LA
    ELECTROANALYSIS, 2006, 18 (04) : 371 - 377
  • [4] Optimisation of an electrochemical impedance spectroscopy aptasensor by exploiting quartz crystal microbalance with dissipation signals
    Formisano, Nello
    Jolly, Pawan
    Bhalla, Nikhil
    Cromhout, Mary
    Flanagan, Shane P.
    Fogel, Ronen
    Limson, Janice L.
    Estrela, Pedro
    SENSORS AND ACTUATORS B-CHEMICAL, 2015, 220 : 369 - 375
  • [5] A combined reflectometry and quartz crystal microbalance with dissipation setup for surface interaction studies
    Wang, Guoliang
    Rodahl, Michael
    Edvardsson, Malin
    Svedhem, Sofia
    Ohlsson, Gabriel
    Hoeoek, Fredrik
    Kasemo, Bengt
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (07):
  • [6] Open Source Quartz Crystal Microbalance with dissipation monitoring
    Mista, C.
    Zalazar, M.
    Penalva, A.
    Martina, M.
    Reta, J. M.
    20TH ARGENTINEAN BIOENGINEERING SOCIETY CONGRESS (XX ARGENTINE BIOENGINEERING CONGRESS AND IX CONFERENCE OF CLINICAL ENGINEERING), (SABI 2015), 2016, 705
  • [7] Probing Bitumen Liberation by a Quartz Crystal Microbalance with Dissipation
    Xiang, Bailin
    Liu, Qingxia
    Long, Jun
    ENERGY & FUELS, 2018, 32 (07) : 7451 - 7457
  • [8] On dissipation of quartz crystal microbalance as a mechanical spectroscopy tool
    Voinova, MV
    Jonson, M
    Kasemo, B
    SPECTROSCOPY-AN INTERNATIONAL JOURNAL, 2004, 18 (04): : 537 - 544
  • [9] Aluminium Nanoparticles Released by Joint Implants - Quartz Crystal Microbalance with Dissipation Analysis
    Pekala, Jakub
    Pekala, Przemyslaw
    Tomaszewska, Iwona
    Santos-Martinez, Maria
    Radomski, Marek
    Paziewski, Mateusz
    Pasieka, Pawel
    Tomaszewski, Krzysztof
    FASEB JOURNAL, 2020, 34
  • [10] Analysis of interpenetrating polymer networks via quartz crystal microbalance with dissipation monitoring
    Irwin, EF
    Ho, JE
    Kane, SR
    Healy, KE
    LANGMUIR, 2005, 21 (12) : 5529 - 5536