Linearity of a silicon carbide photodiode in the deep-UV spectral region: implications on Doppler broadening thermometry

被引:5
|
作者
Dinesan, Hemanth [1 ]
Gravina, Stefania [1 ]
Clivati, Cecilia [2 ]
Castrillo, Antonio [1 ]
Levi, Filippo [2 ]
Gianfrani, Livio [1 ,3 ]
机构
[1] Univ Campania Luigi Vanvitelli, Dipartimento Matemat & Fis, Caserta, Italy
[2] Ist Nazl Ric Metrol, INRiM, Turin, Italy
[3] Ist Nazl Fis Nucl, Sez Napoli, Naples, Italy
关键词
silicon carbide detectors; linear responsivity; deep-UV radiation; FREQUENCY COMB; DETECTOR; RATIO;
D O I
10.1088/1681-7575/aba052
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The linearity of the responsivity of a silicon carbide photodetector in the deep-ultraviolet has been studied using a variant of the so-called intensity-ratio method in view of the implementation of low-uncertainty Doppler-broadening thermometry in mercury vapours. To this purpose, coherent radiation at the wavelength of 254 nm has been produced by means of non-linear mixing of two visible laser beams in a beta-barium borate crystal. The linearity has been quantified as a function of the exposure power, variable in the range between 50 and 300 nW. It is shown that possible deviations from linearity are within the uncertainty of 4 x 10(-5)(corresponding to one standard deviation). Such a performance makes the detector well suitable for the thermometric application.
引用
收藏
页数:6
相关论文
empty
未找到相关数据