Biomechanical test comparing the load to failure of the biodegradable meniscus arrow versus meniscal suture

被引:58
作者
Song, EK [1 ]
Lee, KB [1 ]
机构
[1] Chonnam Univ Hosp, Dept Orthopaed, Kwangju, South Korea
来源
ARTHROSCOPY | 1999年 / 15卷 / 07期
关键词
meniscal repair; biodegradable meniscus arrow; biomechanical test; failure strength;
D O I
10.1016/S0749-8063(99)70004-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The biodegradable Meniscus Arrow (Bionx, Blue Bell, PA) is a newly introduced method for repair of meniscal injury. However, because of a lack of studies, there is still little understanding of its biomechanical behavior. Biomechanical test was carried out to evaluate the load to failure and failure modes of the Meniscus Arrow and to compare the results with those of three traditionally used methods. Thirty-five lateral menisci obtained from Yorkshire pigs were incised longitudinally, simulating a peripheral longitudinal tear, and were repaired by means of Meniscus Arrows (one-point and two-point fixation), as well as with the techniques of knot-end, horizontal, and vertical suture using No. 1 PDS II monofilament suture. Tensile strength testing was performed according to each repair method with the Instron tensometer (Model No. 5569; Instron, Canton, MA) and its results were analyzed using a computerized statistical program. The average maximal tensile strengths were 113.9 +/- 14.6 N in vertical suture, 75.1 +/- 18.4 N in horizontal suture, 53.9 +/- 6.4 N in knot-end suture, 38.3 +/- 4.3 N in one-point fixation Meniscus Arrow repair, and 56.5 +/- 3.5 N in two-paint fixation Meniscus Arrow repair. The initial failure strength for two-point fixation Meniscus Arrow repair is comparable to that of a knot-end suture. Following the results of this study, meticulous caution should be taken when meniscal tears are repaired with Meniscus Arrows.
引用
收藏
页码:726 / 732
页数:7
相关论文
共 21 条
[1]  
Albrecht-Olsen P, 1993, Knee Surg Sports Traumatol Arthrosc, V1, P104, DOI 10.1007/BF01565462
[2]   Failure strength of a new meniscus arrow repair technique: Biomechanical comparison with horizontal suture [J].
AlbrechtOlsen, P ;
Lind, T ;
Kristensen, G ;
Falkenberg, B .
ARTHROSCOPY, 1997, 13 (02) :183-187
[3]  
ALBRECHTOLSEN P, 1994, ACTA ORTHOP SCA S260, V17, P65
[4]  
ANNANDALE T, 1985, BRIT MED J, V17, P779
[5]   WINNER OF THE 1982 ODONOGHUE AWARD - THE MICROVASCULATURE OF THE MENISCUS AND ITS RESPONSE TO INJURY - AN EXPERIMENTAL-STUDY IN THE DOG [J].
ARNOCZKY, SP ;
WARREN, RF .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1983, 11 (03) :131-141
[6]   MENISCAL REPAIR - AN ARTHROSCOPIC TECHNIQUE [J].
BARBER, FA ;
STONE, RG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (01) :39-41
[7]  
Blasier R D, 1997, Instr Course Lect, V46, P531
[8]  
CABAUD EH, 1981, AM J SPORTS MED, V9, P129
[9]  
DEHAVEN KE, 1985, CLIN ORTHOP RELAT R, P31
[10]   Failure strengths of suture versus biodegradable arrow for meniscal repair: An in vitro study [J].
Dervin, GF ;
Downing, KJW ;
Keene, GCR ;
McBride, DG .
ARTHROSCOPY, 1997, 13 (03) :296-300