Outcome of penile prosthesis implantation for treating erectile dysfunction: experience with 504 procedures

被引:148
作者
Minervini, A
Ralph, DJ
Pryor, JP
机构
[1] St Peters Hosp, London WC2A 2EX, England
[2] Inst Urol, London, England
关键词
penis; penile prosthesis; impotence; erectile dysfunction; satisfaction; infection;
D O I
10.1111/j.1464-410X.2005.05907.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE To evaluate the outcome of penile prosthesis surgery for different types of prosthesis. PATIENTS AND METHODS The notes of 447 men who had 504 penile prosthesis implanted between August 1975 and December 2000 were evaluated. Of the prostheses inserted, 393 were malleable, 81 were three-piece inflatable and 30 were self-contained hydraulic prostheses. The mean (range) age of the men was 52 (21-78) years; 404 men had primary implants and 43 had revision surgery after operations at other institutions. The mean follow-up was 50 (1-297) months. RESULTS Of the 447 men, 22 were lost to follow-up immediately after surgery. The most serious postoperative complications were infection (8%) and erosion (5%), which was more common in diabetic patients (10%) and after pelvic trauma with a urethral injury (21%). Of 482 prostheses, 21 failed mechanically (4%) and revision surgery was needed for 5% of the prostheses inserted (24/482). Overall, 89% (377/425) of men could have sexual intercourse and 344 (81%) were satisfied with the results. CONCLUSIONS Of the men implanted with a penile prosthesis, 81% were satisfied with the outcome and an even higher proportion were satisfied with the inflatable prostheses. Dissatisfaction was mainly due to complications that resulted in removal of the prosthesis.
引用
收藏
页码:129 / 133
页数:5
相关论文
共 23 条
[1]   USE OF GLYCOSYLATED HEMOGLOBIN TO IDENTIFY DIABETICS AT HIGH-RISK FOR PENILE PERIPROSTHETIC INFECTIONS [J].
BISHOP, JR ;
MOUL, JW ;
SIHELNIK, SA ;
PEPPAS, DS ;
GORMLEY, TS ;
MCLEOD, DG .
JOURNAL OF UROLOGY, 1992, 147 (02) :386-388
[2]   Efficacy, safety and patient satisfaction outcomes of the AMS 700CX inflatable penile prosthesis: Results of a long-term multicenter study [J].
Carson, CC ;
Mulcahy, JJ ;
Govier, FE .
JOURNAL OF UROLOGY, 2000, 164 (02) :376-380
[3]   10 years of experience with penile prosthesis implantation in Taiwanese patients [J].
Chiang, HS ;
Wu, CC ;
Wen, TC .
JOURNAL OF UROLOGY, 2000, 163 (02) :476-480
[4]   TREATMENT OF ORGANIC IMPOTENCE [J].
CUMMING, J ;
PRYOR, JP .
BRITISH JOURNAL OF UROLOGY, 1991, 67 (06) :640-643
[5]   Long-term mechanical reliability of multicomponent inflatable penile prosthesis: Comparison of device survival [J].
Dubocq, F ;
Tefilli, MV ;
Gheiler, EL ;
Li, HK ;
Dhabuwala, CB .
UROLOGY, 1998, 52 (02) :277-281
[6]  
Evans C, 1998, BRIT J UROL, V81, P591
[7]   Does surgical approach affect the incidence of inflatable penile prosthesis infection? [J].
Garber, BB ;
Marcus, SM .
UROLOGY, 1998, 52 (02) :291-293
[8]   Inflatable penile prosthesis: Results of 150 cases [J].
Garber, BB .
BRITISH JOURNAL OF UROLOGY, 1996, 78 (06) :933-935
[9]   Safety and efficacy outcome of mentor alpha-1 inflatable penile prosthesis implantation for impotence treatment [J].
Goldstein, I ;
Newman, L ;
Baum, N ;
Brooks, M ;
Chaikin, L ;
Goldberg, K ;
McBride, A ;
Krane, RJ .
JOURNAL OF UROLOGY, 1997, 157 (03) :833-839
[10]   Mechanical reliability, surgical complications, and patient and partner satisfaction of the modern three-piece inflatable penile prosthesis [J].
Govier, FE ;
Gibbons, RP ;
Correa, RJ ;
Pritchett, TR ;
Kramer-Levien, D .
UROLOGY, 1998, 52 (02) :282-286