A bent Laue-Laue monochromator for a synchrotron-based computed tomography system

被引:13
作者
Ren, B [1 ]
Dilmanian, FA
Chapman, LD
Ivanov, I
Wu, XY
Zhong, Z
Huang, X
机构
[1] Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA
[2] IIT, Ctr Synchrotron Radiat Res & Instrumentat, Chicago, IL 60616 USA
[3] Brookhaven Natl Lab, Natl Synchrotron Light Source Dept, Upton, NY 11973 USA
关键词
bent crystal monochromator; beam smiling; harmonics rejection; Laue spot; CT;
D O I
10.1016/S0168-9002(99)00056-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We designed and tested a two-crystal bent Laue-Laue monochromator for wide, fan-shaped synchrotron X-ray beams for the program multiple energy computed tomography (MECT) at the National Synchrotron Light Source (NSLS). MECT employs monochromatic X-ray beams from the NSLS's X17B superconducting wiggler beamline for computed tomography (CT) with an improved image quality. MECT uses a fixed horizontal fan-shaped beam with the subject's apparatus rotating around a vertical axis. The new monochromator uses two Czochralski-grown Si[1 1 1] crystals, 0.7 and 1.4 mm thick, respectively, and with thick ribs on their upper and lower ends. The crystals are bent cylindrically, with the axis of the cylinder parallel to the fan beam, using 4-rod benders with two fixed rods and two movable ones. The bent-crystal feature of the monochromator resolved the difficulties we had had with the flat Laue-Laue design previously used in MECT, which included (a) inadequate beam intensity, (b) excessive fluctuations in beam intensity, and (c) instability of the shape of the beam's horizontal profile. Compared with that earlier monochromator, the bent Laue-Laue device tested at 42 and 108 keV showed about a 10-fold larger beam flux, about 5 times better beam stability, 10-fold less harmonic contamination, and a smaller energy bandwidth at certain bending radii. The present work gave us better understanding of the basis for the beam-smiling effect in bent-crystal monochromators, and allowed us to refine the theoretical method of estimating the beam-harmonic contamination in bent-crystal monochromators. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:528 / 550
页数:23
相关论文
共 15 条
  • [1] DYNAMICAL NEUTRON-DIFFRACTION BY CURVED CRYSTALS IN LAUE GEOMETRY
    ALBERTINI, G
    BOEUF, A
    KLAR, B
    LAGOMARSINO, S
    MAZKEDIAN, S
    MELONE, S
    PULITI, P
    RUSTICHELLI, F
    [J]. PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 1977, 44 (01): : 127 - 136
  • [2] GENERALIZATION OF THE PARALLEL POSITION IN DIFFRACTION USING ELASTICALLY CURVED CRYSTALS
    ARROTT, AS
    TEMPLETON, TL
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1985, 18 (DEC) : 388 - 395
  • [3] Single- and dual-energy CT with monochromatic synchrotron x-rays
    Dilmanian, FA
    Wu, XY
    Parsons, EC
    Ren, B
    Kress, J
    Button, TM
    Chapman, LD
    Coderre, JA
    Giron, F
    Greenberg, D
    Krus, DJ
    Liang, Z
    Marcovici, S
    Petersen, MJ
    Roque, CT
    Shleifer, M
    Slatkin, DN
    Thomlinson, WC
    Yamamoto, K
    Zhong, Z
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) : 371 - 387
  • [4] CT with monochromatic synchrotron x rays and its potential in clinical research
    Dilmanian, FA
    Wu, XY
    Ren, BR
    Button, TM
    Chapman, LD
    Dobbs, JM
    Huang, XL
    Nickoloff, EL
    Parsons, EC
    Petersen, MJ
    Thomlinson, WC
    Zhong, Z
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY, 1997, 3149 : 25 - 32
  • [5] X-RAY REFLECTIVITY OF BENT PERFECT CRYSTALS IN BRAGG AND LAUE GEOMETRY
    EROLA, E
    ETELANIEMI, V
    SUORTTI, P
    PATTISON, P
    THOMLINSON, W
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1990, 23 : 35 - 42
  • [6] ETELANIEMI V, 1989, 43247 BNL
  • [7] ON THE SAGITTAL FOCUSING OF SYNCHROTRON RADIATION WITH A DOUBLE CRYSTAL MONOCHROMATOR
    KUSHNIR, VI
    QUINTANA, JP
    GEORGOPOULOS, P
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 328 (03) : 588 - 591
  • [8] Penning P., 1961, PHILIPS RES REP, V16, P419
  • [9] INTERPRETATION OF BENT-CRYSTAL ROCKING CURVES
    POPOVICI, M
    STOICA, AD
    CHALUPA, B
    MIKULA, P
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1988, 21 (03) : 258 - 265
  • [10] REN B, 1997, THESIS STATE U NEW Y