A preliminary analysis of LET effects in the dosimetry of proton beams using PRESAGE™ and optical CT

被引:29
作者
Al-Nowais, S. [1 ]
Doran, S. [1 ]
Kacperek, A. [2 ]
Krstajic, N. [3 ]
Adamovics, J. [4 ]
Bradley, D. [1 ]
机构
[1] Univ Surrey, Dept Phys, Fac Elect & Phys Sci, Guildford GU2 7XH, Surrey, England
[2] Clatterbridge Ctr Oncol, Douglas Cyclotron Unit, Wirral, Merseyside, England
[3] Univ Sheffield, Dept Comp Sci, Sheffield S14DP, S Yorkshire, England
[4] Heuris Pharma, Skillman, NJ USA
关键词
RADIATION-DOSE DISTRIBUTIONS; POLYMER GEL DOSIMETERS; POLYACRYLAMIDE-GELS; TOMOGRAPHY; SCANNER; CODE;
D O I
10.1016/j.apradiso.2008.06.032
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
PRESAGE (TM) is a solid dosimeter based on a clear polyurethane matrix doped with radiochromic components (leuco dyes). on exposure to ionizing radiation a colour change is generated in the dosimeter, and hence an optical absorption or optical density change that can be read out by optical CT. The main focus of present investigations has been to investigate the possible LET dependence of PRESAGE (TM) to the dose deposited at the Bragg maxima using proton beam absorbed dose measurements, and the linearity of response of the dosimeter. Proton irradiations were performed using the proton beam facility at the Douglas Cyclotron, Clatterbridge Centre for Oncology (CCO) using a configuration that approximates the one routinely used in treatment of patients with ocular tumours. The samples were irradiated with both monoenergetic and modulated proton beams. Optical tomography measurements were carried out with Our in-house CCD-based optical-CT system. Initial results for monoenergetic beams show that in PRESAGE (TM) the measured ratio of the Bragg peak dose to entrance dose is approximately 2:1 whereas the true Value measured at CCO is approximately 5:1. For range-modulated proton beams, the absorbed dose close to the end of the proton range, i.e. at the Bragg peak, is underestimated by approximately 20% compared to the corresponding diode measurement. Further investigations are necessary to understand and quantify the effect of LET on PRESAGE (TM), and to measure the uncertainties related to our optical CT. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:415 / 418
页数:4
相关论文
共 23 条
[1]  
Adamovics J., 2004, Journal of Physics: Conference Series, V3, P172, DOI 10.1088/1742-6596/3/1/020
[2]  
ADAMOVICS JA, 2006, COMMUNICATION
[3]  
Bäck SÅJ, 1999, PHYS MED BIOL, V44, P1983, DOI 10.1088/0031-9155/44/8/310
[4]   Fourier transform Raman spectroscopy of polyacrylamide gels (PAGs) for radiation dosimetry [J].
Baldock, C ;
Rintoul, L ;
Keevil, SF ;
Pope, JM ;
George, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) :3617-3627
[5]  
De Deene Y., 2004, Journal of Physics: Conference Series, V3, P34, DOI 10.1088/1742-6596/3/1/006
[6]  
Doran S., 2006, Journal of Physics: Conference Series, V56, P231, DOI 10.1088/1742-6596/56/1/036
[7]   A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results [J].
Doran, SJ ;
Koerkamp, KK ;
Bero, MA ;
Jenneson, P ;
Morton, EJ ;
Gilboy, WB .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (12) :3191-3213
[8]   MEASUREMENT OF RADIATION-DOSE DISTRIBUTIONS BY NUCLEAR MAGNETIC-RESONANCE (NMR) IMAGING [J].
GORE, JC ;
KANG, YS ;
SCHULZ, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (10) :1189-1197
[9]   Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels .1. Development of an optical scanner [J].
Gore, JC ;
Ranade, M ;
Maryanski, MJ ;
Schulz, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12) :2695-2704
[10]   Linear energy transfer dependence of a normoxic polymer gel dosimeter investigated using proton beam absorbed dose measurements [J].
Gustavsson, H ;
Bjäck, SAJ ;
Medin, J ;
Grusell, E ;
Olsson, LE .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17) :3847-3855