A comparison of the Airtraq®, McGrath®, and Macintosh laryngoscopes for difficult paediatric intubation: A manikin study

被引:11
|
作者
Owada, Gen [1 ]
Mihara, Takahiro [1 ,2 ]
Inagawa, Gaku [3 ]
Asakura, Ayako [2 ]
Goto, Takahisa [2 ]
Ka, Koui [1 ]
机构
[1] Kanagawa Childrens Med Ctr, Dept Anaesthesiol, Yokohama, Kanagawa, Japan
[2] Yokohama City Univ, Grad Sch Med, Dept Anaesthesiol & Crit Care Med, Yokohama, Kanagawa, Japan
[3] Yokohama Municipal Citizens Hosp, Dept Anaesthesiol, Yokohama, Kanagawa, Japan
来源
PLOS ONE | 2017年 / 12卷 / 02期
关键词
TRACHEAL INTUBATION; MANAGEMENT; PERSONNEL; TRIAL;
D O I
10.1371/journal.pone.0171889
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background The efficacy of devices for difficult intubation in paediatric patients, especially with a Cormack-Lehane grade 4 view, has yet to be established. We compared intubating parameters among three devices (the Airtraq (R), McGrath (R), and Macintosh laryngoscopes). Methods This study is a randomised cross-over trial. Participants were 20 anaesthetists. Each device was tested three times using a paediatric manikin with a Cormack-Lehane grade 4 view. The order to use each device was randomised by a computer-generated random sequence. The primary endpoint was the rate of successful intubation. Secondary endpoints included the time taken to intubate, percentage of glottic opening score, and severity of potential dental trauma. Results The successful intubation rates of the Airtraq (R), McGrath (R), and Macintosh laryngoscopes were 100%, 72%, and 45%, respectively. The risk ratio of the success rates of Airtraq (R) compared with McGrath (R) and Macintosh laryngoscopes were 1.40 (95% CI; 1.19-1.64, P < 0.001) and 2.22 (95% CI; 1.68-2.94, P < 0.001), respectively. The modified Cormack-Lehane grade and percentage of the glottic opening score were better for the Airtraq (R) than for the other devices. The dental trauma score was lower for the Airtraq (R) than for the other devices. There were no significant differences in the intubation time among the groups. Conclusions The Airtraq (R) had higher success rate, had better visibility, and was associated with less dental trauma than the other devices in a difficult paediatric intubation model with a Cormack-Lehane grade 4 view.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Intubation Biomechanics Laryngoscope Force and Cervical Spine Motion during Intubation with Macintosh and Airtraq Laryngoscopes
    Hindman, Bradley J.
    Santoni, Brandon G.
    Puttlitz, Christian M.
    From, Robert P.
    Todd, Michael M.
    ANESTHESIOLOGY, 2014, 121 (02) : 260 - 271
  • [22] Comparison of the Macintosh, GlideScope®, Airtraq®, and King Vision™ laryngoscopes in routine airway management
    Al-Ghamdi, Abdulmohsen A.
    El Tahan, Mohamed R.
    Khidr, Alaa M.
    MINERVA ANESTESIOLOGICA, 2016, 82 (12) : 1278 - +
  • [23] Comparison of McGrath Series 3 and Macintosh Laryngoscopes for Tracheal Intubation in Patients With Normal Airway by Inexperienced Anesthetists A Randomized Study
    Liu, Zi-Jia
    Yi, Jie
    Guo, Wen-Juan
    Ma, Chao
    Huang, Yu-Guang
    MEDICINE, 2016, 95 (02)
  • [24] Comparison of Airtraq with Macintosh Laryngoscope for Endotracheal Intubation in Paediatric Patients: A Randomised Controlled Trial
    Dwivedi, Yogita
    Jain, Ankit
    Yadav, Vijay Pratap
    Agarwal, Atiharsh Mohan
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2022, 16 (03) : UC16 - UC18
  • [25] Comparison of the Pentax Airway Scope® and McGrath® Videolaryngoscope with the Macintosh laryngoscope in tracheal intubation by anaesthetists unfamiliar with videolaryngoscopes: a manikin study
    Sharma, D. -J.
    Weightman, W. M.
    Travis, A.
    ANAESTHESIA AND INTENSIVE CARE, 2010, 38 (01) : 39 - 42
  • [26] Comparison of the ETView Single Lumen and Macintosh laryngoscopes for endotracheal intubation in an airway manikin with immobilized cervical spine by novice paramedics A randomized crossover manikin trial
    Gawlowski, Pawel
    Smereka, Jacek
    Madziala, Marcin
    Cohen, Barak
    Ruetzler, Kurt
    Szarpak, Lukasz
    MEDICINE, 2017, 96 (16)
  • [27] Comparison of the C-MAC® videolaryngoscope with the Macintosh, Glidescope® and Airtraq® laryngoscopes in easy and difficult laryngoscopy scenarios in manikins reply
    McElwain, J.
    Laffey, J. G.
    ANAESTHESIA, 2010, 65 (09) : 955 - 956
  • [28] A simple fibreoptic assisted laryngoscope for paediatric difficult intubation: a manikin study
    Komiya, K.
    Inagawa, G.
    Nakamura, K.
    Kikuchi, T.
    Fujimoto, J.
    Sugawara, Y.
    Goto, T.
    ANAESTHESIA, 2009, 64 (04) : 425 - 429
  • [29] A comparison of the forces applied to a manikin during laryngoscopy with the GlideScope® and Macintosh laryngoscopes
    Russell, T.
    Lee, C.
    Firat, M.
    Cooper, R. M.
    ANAESTHESIA AND INTENSIVE CARE, 2011, 39 (06) : 1098 - 1102
  • [30] Comparison of Airtraq and Miller laryngoscopes for intubation in patients with inline manual neck stabilization
    Smereka, Jacek
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (08) : 1721 - 1721