A miniaturized transcutaneous system for continuous glucose monitoring

被引:30
作者
Croce, Robert A., Jr.
Vaddiraju, SanthiSagar [1 ,2 ]
Kondo, Jun
Wang, Yan [3 ]
Zuo, Liang [4 ]
Zhu, Kai [4 ]
Islam, Syed K. [4 ]
Burgess, Diane J. [3 ]
Papadimitrakopoulos, Fotios [2 ,5 ]
Jain, Faquir C. [1 ]
机构
[1] Univ Connecticut, Biorasis Inc, Technol Incubat Program, Storrs, CT 06269 USA
[2] Univ Connecticut, Nanomat Optoelect Lab, Polymer Program, Inst Mat Sci, Storrs, CT 06269 USA
[3] Univ Connecticut, Dept Pharmaceut Sci, Storrs, CT 06269 USA
[4] Univ Tennessee, Knoxville, TN USA
[5] Univ Connecticut, Dept Chem, Storrs, CT 06269 USA
基金
美国国家科学基金会;
关键词
Implantable sensors; CMOS circuits; Amperometric glucose sensors; Low-power microelectronics; In vivo monitoring; TELEMETRY-INSTRUMENTATION SYSTEM; IMPLANTED GLUCOSE; ELECTRON-TRANSFER; SENSOR; DEXAMETHASONE; MICROSPHERES; PERFORMANCE; TECHNOLOGY; BIOSENSORS; DELIVERY;
D O I
10.1007/s10544-012-9708-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable sensors for continuous glucose monitoring hold great potential for optimal diabetes management. This is often undermined by a variety of issues associated with: (1) negative tissue response; (2) poor sensor performance; and (3) lack of device miniaturization needed to reduce implantation trauma. Herein, we report our initial results towards constructing an implantable device that simultaneously address all three aforementioned issues. In terms of device miniaturization, a highly miniaturized CMOS (complementary metal-oxide-semiconductor) potentiostat and signal processing unit was employed (with a combined area of 0.665 mm(2)). The signal processing unit converts the current generated by a transcutaneous, Clark-type amperometric sensor to output frequency in a linear fashion. The Clark-type amperometric sensor employs stratification of five functional layers to attain a well-balanced mass transfer which in turn yields a linear sensor response from 0 to 25 mM of glucose concentration, well beyond the physiologically observed (2 to 22 mM) range. In addition, it is coated with a thick polyvinyl alcohol (PVA) hydrogel with embedded poly(lactic-co-glycolic acid) (PLGA) microspheres intended to provide continuous, localized delivery of dexamethasone to suppress inflammation and fibrosis. In vivo evaluation in rat model has shown that the transcutaneous sensor system reproducibly tracks repeated glycemic events. Clarke's error grid analysis on the as-obtained glycemic data has indicated that all of the measured glucose readings fell in the desired Zones A & B and none fell in the erroneous Zones C, D and E. Such reproducible operation of the transcutaneous sensor system, together with low power (140 mu W) consumption and capability for current-to-frequency conversion renders this a versatile platform for continuous glucose monitoring and other biomedical sensing devices.
引用
收藏
页码:151 / 160
页数:10
相关论文
共 56 条
[1]   A Wireless-Implantable Microsystem for Continuous Blood Glucose Monitoring [J].
Ahmadi, Mohammad Mahdi ;
Jullien, Graham A. .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2009, 3 (03) :169-180
[2]  
Bhardwaj Upkar, 2007, J Diabetes Sci Technol, V1, P8
[3]   PLGA/PVA hydrogel composites for long-term inflammation control following s.c. implantation [J].
Bhardwaj, Upkar ;
Sura, Radhakrishana ;
Papadimitrakopoulos, Fotios ;
Burgess, Diane J. .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2010, 384 (1-2) :78-86
[4]   Implantable ultra-low power DSP-based system for a miniature chemico-rheological biosensor [J].
Bolomey, L. ;
Meurville, E. ;
Ryser, P. .
PROCEEDINGS OF THE EUROSENSORS XXIII CONFERENCE, 2009, 1 (01) :1235-1238
[5]   Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor [J].
Boyne, MS ;
Silver, DM ;
Kaplan, J ;
Saudek, CD .
DIABETES, 2003, 52 (11) :2790-2794
[6]   Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients - Part 2. Superiority of the one-point calibration method [J].
Choleau, C ;
Klein, JC ;
Reach, G ;
Aussedat, B ;
Demaria-Pesce, V ;
Wilson, GS ;
Gifford, R ;
Ward, WK .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (08) :647-654
[7]   Amperometric glucose biosensor based on electrodeposition of platinum nanoparticles onto covalently immobilized carbon nanotube electrode [J].
Chu, Xia ;
Duan, Daxue ;
Shen, Guoli ;
Yu, Ruqin .
TALANTA, 2007, 71 (05) :2040-2047
[8]   DIRECT ELECTRICAL COMMUNICATION BETWEEN CHEMICALLY MODIFIED ENZYMES AND METAL-ELECTRODES .1. ELECTRON-TRANSFER FROM GLUCOSE-OXIDASE TO METAL-ELECTRODES VIA ELECTRON RELAYS, BOUND COVALENTLY TO THE ENZYME [J].
DEGANI, Y ;
HELLER, A .
JOURNAL OF PHYSICAL CHEMISTRY, 1987, 91 (06) :1285-1289
[9]   AMPEROMETRIC GLUCOSE SENSOR WITH FERROCENE AS AN ELECTRON-TRANSFER MEDIATOR [J].
DONG, SJ ;
WANG, BX ;
LIU, BF .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (03) :215-222
[10]   Raman spectroscopy for noninvasive glucose measurements [J].
Enejder, AMK ;
Scecina, TG ;
Oh, J ;
Hunter, M ;
Shih, WC ;
Sasic, S ;
Horowitz, GL ;
Feld, MS .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (03) :1-9