Enzymatic but not compensated Jaffe methods reach the desirable specifications of NKDEP at normal levels of creatinine. Results of the French multicentric evaluation

被引:48
作者
Boutten, Anne [1 ]
Bargnoux, Anne-Sophie [2 ]
Carlier, Marie-Christine [3 ]
Delanaye, Pierre [4 ]
Rozet, Eric [5 ]
Delatour, Vincent [6 ]
Cavalier, Etienne [4 ]
Hanser, Anne-Marie [7 ]
Froissart, Marc [8 ]
Cristol, Jean-Paul [2 ]
Pieroni, Laurence [9 ]
机构
[1] CHU Bichat, APHP, Paris, France
[2] CHU Lapeyronie, Montpellier, France
[3] Hop Lyon Sud, Lyon, France
[4] Univ Liege, CHU Sart Tilman, Liege, Belgium
[5] Univ Liege, CIRM, Chem Lab, Liege, Belgium
[6] Lab Natl Metrol & Essais LNE, Dept Biomed & Organ Chem, Paris, France
[7] Hosp Civils, Colmar, France
[8] Hop Europeen Georges Pompidou, APHP, Paris, France
[9] Grp Hosp Pitie Salpetriere, APHP, F-75634 Paris, France
关键词
Compensated Jaffe method; Creatinine; Enzymatic method; IDMS-traceable; Multicentric evaluation; Standardization of methods; SERUM CREATININE; WORKING GROUP; ASSAYS; TIME;
D O I
10.1016/j.cca.2013.01.021
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The French Society of Clinical Biochemistry conducted this study to compare the accuracy and performances of the best creatinine enzymatic assays and the compensated Jaffe methods from the same manufacturers. Creatinine was measured in 3 serum pools with creatinine levels of 35.9 +/- 0.9 mu mol/L, 74.4 +/- 1.4 mu mol/L, and 97.9 +/- 1.7 mu mol/L (IDMS determination). The performances of the assays (total error that includes the contribution of bias and imprecision) were evaluated using Monte-Carlo simulations and compared against desirable NKDEP criteria. The enzymatic assays always fell within the desirable total Error of 7.6%. By contrast, this requirement was never obtained for the compensated Jaffe methods at the critical level of 74.4 +/- 1.4 mu mol/L Only the compensated Jaffe creatinine on Olympus analyzer reached this specification at 35.9 +/- 0.9 and 97.9 +/- 1.7 mu mol/L levels. This study demonstrates that, despite substantial improvement regarding traceability to the IDMS reference method and precision, compensated Jaffe creatinine methods, by contrast to enzymatic ones, do not reach the desirable specifications of NKDEP at normal levels of creatinine. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:132 / 135
页数:4
相关论文
共 11 条
  • [1] Prime Time for Enzymatic Creatinine Methods in Pediatrics
    Cobbaert, Christa M.
    Baadenhuijsen, Henk
    Weykamp, Cas W.
    [J]. CLINICAL CHEMISTRY, 2009, 55 (03) : 549 - 558
  • [2] Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement An international study in the framework of the EC4 creatinine standardization working group
    Delanghe, Joris R.
    Cobbaert, Christa
    Galteau, Marie-Madeleine
    Harmoinen, Aimo
    Jansen, Rob
    Kruse, Rolf
    Laitinen, Paeivi
    Thienpont, Linda M.
    Wuyts, Birgitte
    Weykamp, Cas
    Panteghini, Mauro
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (09) : 1319 - 1325
  • [3] Delatour V, 2011, REV FR METROLOGIE, V26, P21
  • [4] Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method Principles
    Greenberg, Neil
    Roberts, William L.
    Bachmann, Lorin M.
    Wright, Elizabeth C.
    Dalton, R. Neil
    Zakowski, Jack J.
    Miller, W. Greg
    [J]. CLINICAL CHEMISTRY, 2012, 58 (02) : 391 - 401
  • [5] Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffe method
    Junge, W
    Wilke, B
    Halabi, A
    Klein, G
    [J]. CLINICA CHIMICA ACTA, 2004, 344 (1-2) : 137 - 148
  • [6] Specimen materials, target values and commutability for external quality assessment (proficiency testing) schemes
    Miller, WG
    [J]. CLINICA CHIMICA ACTA, 2003, 327 (1-2) : 25 - 37
  • [7] Recommendations for improving serum creatinine measurement: A report from the laboratory working group of the National Kidney Disease Education Program
    Myers, GL
    Miller, WG
    Coresh, J
    Fleming, J
    Greenberg, N
    Greene, T
    Hostetter, T
    Levey, AS
    Panteghini, M
    Welch, M
    Eckfeldt, JH
    [J]. CLINICAL CHEMISTRY, 2006, 52 (01) : 5 - 18
  • [9] A multicentric evaluation of IDMS-traceable creatinine enzymatic assays
    Pieroni, Laurence
    Delanaye, Pierre
    Boutten, Anne
    Bargnoux, Anne-Sophie
    Rozet, Eric
    Delatour, Vincent
    Carlier, Marie-Christine
    Hanser, Anne-Marie
    Cavalier, Etienne
    Froissart, Marc
    Cristol, Jean-Paul
    [J]. CLINICA CHIMICA ACTA, 2011, 412 (23-24) : 2070 - 2075
  • [10] SIEKMANN L, 1985, J CLIN CHEM CLIN BIO, V23, P137