The Aachen-keratoprosthesis - a flexible KPro that permits intraocular pressure measurement

被引:6
作者
Krug, A [1 ]
Kompa, S [1 ]
Schrage, NF [1 ]
机构
[1] Univ Aachen, Dept Ophthalmol, D-5100 Aachen, Germany
关键词
tonometry; implant; keratoprosthesis; glaucoma; cornea;
D O I
10.1177/039139880202500311
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Postoperative glaucoma is one of the major complications observed in patients with keratoprostheses (KPro). The optical components of the majority of KPros are rigid and inflexible, which prevents indirect tonometry by common methods. This study confirms that the Aachen-KPro allows measurements of intraocular pressure (IOP) due to its flexible optical part with a modified Schiotz tonometer. Methods: The Aachen-KPro was placed in a special chamber where pressure can be generated and monitored by a transducer. Measurements were taken by common ophthalmologic methods, The results were compared to the initial pressure values in the chamber Results: With the Goldmann tonometer, the Tono-Pen and the topography system (Technomed), no significant pressure changes could be observed. The results of the Schiotz tonometer were promising. It was slightly modified and standard curves with different weights could be obtained. Conclusions: With a modified Schiotz tonometer, it is possible to detect elevated IOPs. Modifying this common instrument is inexpensive and can be easily performed. This advantage of the Aachen-KPro permits early management of postoperative high tension glaucoma.
引用
收藏
页码:238 / 242
页数:5
相关论文
共 11 条
[1]   A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma [J].
Ayyala, RS ;
Zurakowski, D ;
Smith, JA ;
Monshizadeh, R ;
Netland, PA ;
Richards, DW ;
Layden, WE .
OPHTHALMOLOGY, 1998, 105 (10) :1968-1976
[2]  
CARDONA H, 1962, 19 CONC OPHTHALMOL, V11, P1211
[3]   PROSTHOKERATOPLASTY [J].
DOHLMAN, CH ;
SCHNEIDE.HA ;
DOANE, MG .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1974, 77 (05) :694-700
[4]   Aachen-Keratoprosthesis as temporary implant. Case report on first clinical application [J].
Kompa, S ;
Langefeld, S ;
Kirchhof, B ;
Brenman, K ;
Schrage, N .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2000, 23 (05) :345-348
[5]   The Type II Aachen-Keratoprosthesis in humans: case report of the first prolonged application [J].
Kompa, S ;
Redbrake, C ;
Langefeld, S ;
Brenman, K ;
Schrage, N .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2001, 24 (02) :110-114
[6]  
Langefeld S, 2000, GRAEF ARCH CLIN EXP, V238, P722
[7]   Functionally adapted surfaces on a silicone keratoprosthesis [J].
Langefeld, S ;
Völcker, N ;
Kompa, S ;
von Fischern, T ;
Bienert, H ;
Klee, D ;
Höcker, H ;
Reim, M ;
Schrage, NF .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1999, 22 (04) :235-241
[8]   Glaucoma associated with keratoprosthesis [J].
Netland, PA ;
Terada, H ;
Dohlman, CH .
OPHTHALMOLOGY, 1998, 105 (04) :751-757
[9]  
NUSSBAUM JN, 1853, DTSCH KLIN, V34, P367
[10]   STRAMPELLI OSTEO-ODONTO-KERATOPROSTHESIS - CLINICAL AND HISTOLOGICAL LONG-TERM FEATURES OF 3 PROSTHESES [J].
RICCI, R ;
PECORELLA, I ;
DELLAROCCA, C ;
DITONDO, U ;
MARCHI, V .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1992, 76 (04) :232-234