Design and evaluation of a new bioelectrical impedance sensor for micro-surgery: application to retinal vein cannulation

被引:15
作者
Schoevaerdts, Laurent [1 ]
Esteveny, Laure [1 ]
Gijbels, Andy [1 ]
Smits, Jonas [1 ]
Reynaerts, Dominiek [1 ]
Vander Poorten, Emmanuel [1 ]
机构
[1] Univ Leuven, KU Leuven, Dept Mech Engn, Louvain, Belgium
关键词
Vitreoretinal surgery; Sensorised instruments; Bio-impedance sensor; Robot-assisted surgery; TISSUE-PLASMINOGEN ACTIVATOR; OCCLUSION; NEEDLE; BIOIMPEDANCE; INJECTION;
D O I
10.1007/s11548-018-1850-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose Nowadays, millions of people suffer from retinal vein occlusion, a blind-making eye disease. No curative treatment currently exists for this vascular disorder. However, a promising treatment consists in injecting a thrombolytic drug directly inside the affected retinal vessel. Successfully puncturing miniature vessels with diameters between 50 and 400m remains a real challenge, amongst others due to human hand tremor, poor visualisation and depth perception. As a consequence, there is a significant risk of double-puncturing the targeted vessel. Sub-surfacic injection of thrombolytic agent could potentially lead to severe retinal damage.Methods A new bio-impedance sensor has been developed to visually display the instant of vessel puncture. The physical working principle of the sensor has been analysed, and a representative electrical model has been derived. Based on this model, the main design parameters were derived to maximise the sensor sensitivity. A detailed characterisation and experimental validation of this concept were conducted.Results Stable, repeatable and robust impedance measurements were obtained. In an experimental campaign, 35 puncture attempts on ex vivo pig eyes vessels were conducted. A confusion matrix shows a detection accuracy of 80% if there is a puncture, a double puncture or no puncture. The 20% of inaccuracy most probably comes from the limitations of the employed eye model and the experimental conditions.Conclusions The developed bio-impedance sensor has shown great promise to help in avoiding double punctures when cannulating retinal veins. Compared to other puncture detection methods, the proposed sensor is simple and therefore potentially more affordable. Future research will include validation in an in vivo situation involving vitreoretinal surgeons.
引用
收藏
页码:311 / 320
页数:10
相关论文
共 22 条
  • [1] [Anonymous], J UROLOGY
  • [2] [Anonymous], IEEE INT C BIOM ROB
  • [3] [Anonymous], 2018 IEEE INT C ROB
  • [4] [Anonymous], MEDICALXPRESS
  • [5] [Anonymous], P 22 INT C IEEE ENG
  • [6] [Anonymous], EV PROC MATT
  • [7] [Anonymous], IEEE SENSORS
  • [8] A New Venous Entry Detection Method Based on Electrical Bio-impedance Sensing
    Cheng, Zhuoqi
    Davies, Brian L.
    Caldwell, Darwin G.
    Mattos, Leonardo S.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2018, 46 (10) : 1558 - 1567
  • [9] Retinal endovascular lysis in ischemic central retinal vein occlusion - One-year results of a pilot study
    Feltgen, Nicolas
    Junker, Bernd
    Agostini, Hansjuergen
    Hansen, Lutz L.
    [J]. OPHTHALMOLOGY, 2007, 114 (04) : 716 - 723
  • [10] Gijbels A, 2015, IEEE INT CONF ROBOT, P2270, DOI 10.1109/ICRA.2015.7139500