Two-dimensional tomography for gas concentration and temperature distributions based on tunable diode laser absorption spectroscopy

被引:102
作者
Wang, F. [1 ]
Cen, K. F. [1 ]
Li, N. [1 ]
Jeffries, Jay B. [2 ]
Huang, Q. X. [1 ]
Yan, J. H. [1 ]
Chi, Y. [1 ]
机构
[1] Zhejiang Univ, State Key Lab Clean Energy Utilizat, Hangzhou 310027, Zhejiang, Peoples R China
[2] Stanford Univ, Dept Mech Engn, High Temp Gasdynam Lab, Stanford, CA 94305 USA
关键词
tomography; gas concentration; temperature; tunable diode laser absorption spectroscopy; SENSOR; CO2;
D O I
10.1088/0957-0233/21/4/045301
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A tomography system is presented that uses wavelength-scanned direct absorption of two transitions of a target species (NH3 in the demonstration experiment) to determine the distributions of gas concentration and temperature. The absorption measurements are performed simultaneously from four platforms that each rotate a beam from a single laser through an 11 degrees arc, acquiring a data set from all four laser platforms in 100 ms to enable observation of dynamic flow events. The laser is wavelength scanned through two absorption transitions with different internal energy producing two sets of equations with species mole fraction and temperature as independent variables. The mole fraction and temperature distributions are reconstructed using the algebraic reconstruction technique (ART) for this set of incomplete projections. A numerical simulation is used to evaluate the measurement accuracy for measurements of an NH3 mixture escaping from an open pipe. This phantom distribution is then realized in the laboratory and the measurement strategy is demonstrated using a tunable diode laser absorption spectroscopy (TDLAS) measurement using a single laser near 1.5 mu m to scan adjacent transitions in NH3. The reconstruction of NH3 concentration and gas temperature is compared with independently determined values to illustrate the fidelity of the tomographically reconstructed distributions for the NH3 mole fraction assuming a fixed temperature and for unknown mole fraction and temperature. Potential extensions of this research in the future include evaluation of other reconstruction algorithms and investigation of the dynamic distribution of various gases for combustion diagnostics.
引用
收藏
页数:10
相关论文
共 15 条
  • [1] FAST OPTICAL-ABSORPTION TOMOGRAPHY
    BEITING, EJ
    [J]. OPTICS LETTERS, 1991, 16 (16) : 1280 - 1282
  • [2] CAREY SJ, 1998, 1 WORLD C IND PROC T, P480
  • [3] EXPERIMENTAL RECONSTRUCTION OF THE SPATIAL DENSITY DISTRIBUTION OF A NONREACTING FLOW WITH A SMALL NUMBER OF ABSORPTION-MEASUREMENTS
    CHUNG, KB
    GOULDIN, FC
    WOLGA, GJ
    [J]. APPLIED OPTICS, 1995, 34 (24): : 5492 - 5500
  • [4] Conklin J., 2006, Journal of Physics: Conference Series, V56, P211, DOI 10.1088/1742-6596/56/1/030
  • [5] Kohse-Hoinghaus K., 2002, APPL COMBUSTION DIAG
  • [6] Fast laser scanning optical-CT apparatus for 3D radiation dosimetry
    Krstajic, Nikola
    Doran, Simon J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (11) : N257 - N263
  • [7] Wavelength modulation absorption spectroscopy with 2f detection using multiplexed diode lasers for rapid temperature measurements in gaseous flows
    Liu, JTC
    Jeffries, JB
    Hanson, RK
    [J]. APPLIED PHYSICS B-LASERS AND OPTICS, 2004, 78 (3-4): : 503 - 511
  • [8] ANALYSIS OF THE HIGH-RESOLUTION SPECTRUM OF AMMONIA ((NH3)-N-14) IN THE NEAR-INFRARED REGION, 6400-6900-CM(-1)
    LUNDSBERGNIELSEN, L
    HEGELUND, F
    NICOLAISEN, FM
    [J]. JOURNAL OF MOLECULAR SPECTROSCOPY, 1993, 162 (01) : 230 - 245
  • [9] Numerical investigation of hyperspectral tomography for simultaneous temperature and concentration imaging
    Ma, Lin
    Cai, Weiwei
    [J]. APPLIED OPTICS, 2008, 47 (21) : 3751 - 3759
  • [10] Diode laser sensor for measurements of CO, CO2, and CH4 in combustion flows
    Mihalcea, RM
    Baer, DS
    Hanson, RK
    [J]. APPLIED OPTICS, 1997, 36 (33): : 8745 - 8752