Construction of Au nanoparticles on choline chloride modified glassy carbon electrode for sensitive detection of nitrite

被引:154
作者
Wang, Po [2 ]
Mai, Zhibin [2 ]
Dai, Zong [2 ]
Li, Yongxin [1 ]
Zou, Xiaoyong [2 ]
机构
[1] Anhui Normal Univ, Coll Chem & Mat Sci, Wuhu 241000, Peoples R China
[2] Sun Yat Sen Univ, Sch Chem & Chem Engn, Guangzhou 510275, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Nitrite; Electrocatalytic oxidation; Au nanoparticles; Choline chloride; Sensor; GOLD NANOPARTICLES; ELECTROCATALYTIC OXIDATION; AMPEROMETRIC SENSOR; ELECTROCHEMICAL OXIDATION; GLUCOSE-OXIDASE; BIOSENSOR; REDUCTION; NANOTUBES; NITRATE; NANOCLUSTERS;
D O I
10.1016/j.bios.2009.04.006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A promising electrochemical sensor for sensitive determination of nitrite was fabricated by construction of Au nanoparticles on the surface of choline chloride (Ch) modified glassy carbon electrode (GCE). Field emission scanning electron microscope, powder X-ray diffraction, X-ray photoelectron spectroscopy and electrochemical techniques were used for the surface characterization of the modified electrode. It was demonstrated that Ch was covalently immobilized onto the GCE surface forming a planted Ch monolayer, which could provide a suitable supporting material for the construction of Au nanoparticles. As a result, the Au nanoparticles with average size of about 110 nm were assembled to form a flowerlike structure on the surface of Ch monolayer. Moreover, the uniform nano-Au/Ch film exhibited remarkable electrocatalytic activity towards the oxidation of nitrite with obvious reduction of overpotential. Under the optimum conditions, the linear range for the detection of nitrite was 4.0 x 10(-7) to 7.5 x 10(-4) M with a high sensitivity of 0.354 mu A mu M-1, and a low detection limit of 1.0 x 10(-7) M. The proposed method was successfully applied in the detection of nitrite in water samples and sausage samples, and the results were consistent with those obtained by ion chromatography and UV-visible spectrophotometric methods. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:3242 / 3247
页数:6
相关论文
共 48 条
[1]   Biosensing nitrite using the system nitrite redutase/Nafion/methyl viologen - A voltammetric study [J].
Almeida, M. Gabriela ;
Silveira, Celia M. ;
Moura, Jose J. G. .
BIOSENSORS & BIOELECTRONICS, 2007, 22 (11) :2485-2492
[2]  
Bard A.J., 2022, Electrochemical Methods: Fundamentals and Applications
[3]   Electrochemical oxidation of nitrite and the oxidation and reduction of NO2 in the room temperature ionic liquid [C2mim][NTf2] [J].
Broder, Tessa L. ;
Silvester, Debbie S. ;
Aldous, Leigh ;
Hardacre, Christopher ;
Compton, Richard G. .
JOURNAL OF PHYSICAL CHEMISTRY B, 2007, 111 (27) :7778-7785
[5]   Nitrate and nitrite electrocatalytic reduction on Rh-modified pyrolytic graphite electrodes [J].
Brylev, Oleg ;
Sarrazin, Mathieu ;
Roue, Lionel ;
Belanger, Daniel .
ELECTROCHIMICA ACTA, 2007, 52 (21) :6237-6247
[6]  
CAI Q, 2001, ANAL SCI, V17, P1917
[7]   Nitrite and nitrosyl compounds in food preservation [J].
Cammack, R ;
Joannou, CL ;
Cui, XY ;
Martinez, CT ;
Maraj, SR ;
Hughes, MN .
BIOCHIMICA ET BIOPHYSICA ACTA-BIOENERGETICS, 1999, 1411 (2-3) :475-488
[8]   Electrocatalytic determination of nitrite on a rigid disk electrode having cobalt phthalocyanine prepared in situ [J].
Canevari, Thiago C. ;
Luz, Rita C. S. ;
Gushikem, Yoshitaka .
ELECTROANALYSIS, 2008, 20 (07) :765-770
[9]   An electropolymerized Nile Blue sensing film-based nitrite sensor and application in food analysis [J].
Chen, Xiaowei ;
Wang, Fang ;
Chen, Zilin .
ANALYTICA CHIMICA ACTA, 2008, 623 (02) :213-220
[10]   A novel nitrite biosensor based on the direct electron transfer of hemoglobin immobilized on CdS hollow nanospheres [J].
Dai, Zhihui ;
Bai, Hongyan ;
Hong, Mei ;
Zhu, Yinyan ;
Bao, Jianchun ;
Shen, Jian .
BIOSENSORS & BIOELECTRONICS, 2008, 23 (12) :1869-1873