Is TKA Using Patient-specific Instruments Comparable to Conventional TKA? A Randomized Controlled Study of One System

被引:93
作者
Roh, Yoon Whan [1 ]
Kim, Tae Woo [2 ]
Lee, Sahnghoon [2 ]
Seong, Sang Cheol [2 ]
Lee, Myung Chul [2 ]
机构
[1] Seoul NOW Hosp, Joint Reconstruct Ctr, Kyonggi Do, South Korea
[2] Seoul Natl Univ Hosp, Dept Orthoped Surg, Seoul 110744, South Korea
关键词
TOTAL KNEE ARTHROPLASTY; COMPUTER-ASSISTED SURGERY; CORONAL ALIGNMENT; REPLACEMENT; GUIDES; MALALIGNMENT; METAANALYSIS; COMPONENT; FAILURE;
D O I
10.1007/s11999-013-3206-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Patient-specific CT-based instrumentation may reduce implant malpositioning and improve alignment in TKA. However, it is not known whether this innovation is an advance that benefits patients. We evaluated (1) the precision of patient-specific TKA by comparing the incidence of outliers in postoperative alignment between TKAs using patient-specific instruments and TKAs using conventional instruments, and (2) the reliability of patient-specific instruments by intraoperatively investigating whether the surgery could be completed with patient-specific instruments alone. In this randomized controlled trial, we compared patient-specific TKA instruments from one manufacturer (n = 50) with conventional TKA instruments (n = 50). Postoperative hip-knee-ankle angles, femoral component rotation, and coronal and sagittal alignments of each component were measured. The validity of the patient-specific instrument was examined using cross-checking procedures with conventional instruments during the surgeries. When the procedure could not be completed accurately with patient-specific instruments, the procedure was converted to TKA using conventional instruments, and the frequency of this occurrence was tallied. Outliers in the hip-knee-ankle angle were comparable between groups (12% in the patient-specific instrument group and 10% in the conventional instrument group). Other parameters such as sagittal alignment and femoral component rotation did not differ in terms of outliers. Patient-specific guides were abandoned in eight knees (16%) during the surgery because of malrotation of the femoral components and decreased slope of the tibia. Accuracy was comparable between TKAs done with patient-specific instruments and those done with conventional instruments. However, the patient-specific instrument procedures had to be aborted frequently, incurring expenses that did not benefit patients. Level II, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
引用
收藏
页码:3988 / 3995
页数:8
相关论文
共 29 条
  • [1] Alignment in total knee arthroplasty -: A comparison of computer-assisted surgery with the conventional technique
    Bäthis, H
    Perlick, L
    Tingart, M
    Lüring, C
    Zurakowski, D
    Grifka, J
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (05): : 682 - 687
  • [2] Bäthis H, 2004, ORTHOPEDICS, V27, P476
  • [3] Navigated total knee replacement - A meta-analysis
    Bauwens, Kai
    Matthes, Gerrit
    Wich, Michael
    Gebhard, Florian
    Hanson, Beate
    Ekkernkamp, Axel
    Stengel, Dirk
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (02) : 261 - 269
  • [4] Computer-Assisted Techniques Versus Conventional Guides for Component Alignment in Total Knee Arthroplasty
    Blakeney, William G.
    Khan, Riaz J. K.
    Wall, Simon J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (15) : 1377 - 1384
  • [5] Femoral shaft fracture resulting from femoral tracker placement in navigated TKA
    Bonutti, Peter
    Dethmers, Daniel
    Stiehl, James B.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (06) : 1499 - 1502
  • [6] A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement
    Chareancholvanich, K.
    Narkbunnam, R.
    Pornrattanamaneewong, C.
    [J]. BONE & JOINT JOURNAL, 2013, 95B (03) : 354 - 359
  • [7] Delp SL, 1998, CLIN ORTHOP RELAT R, P49
  • [8] Gap Balancing versus Measured Resection Technique for Total Knee Arthroplasty
    Dennis, Douglas A.
    Komistek, Richard D.
    Kim, Raymond H.
    Sharma, Adrija
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) : 102 - 107
  • [9] Kinematically Versus Mechanically Aligned Total Knee Arthroplasty
    Dossett, H. Gene
    Swartz, George J.
    Estrada, Nicolette A.
    LeFevre, George W.
    Kwasman, Bertram G.
    [J]. ORTHOPEDICS, 2012, 35 (02) : E160 - E169
  • [10] Rotational malalignment of the femoral component in total knee arthroplasty
    Fehring, TK