Comparison of ultra-high-resolution parallel-hole collimator materials based on the CdTe pixelated semiconductor SPECT system

被引:25
|
作者
Lee, Young-Jin [1 ]
Ryu, Hyun-Ju [1 ]
Lee, Seung-Wan [1 ]
Park, Su-Jin [1 ]
Kim, Hee-Joung [1 ]
机构
[1] Yonsei Univ, Coll Hlth Sci, Dept Radiol Sci, Wonju 220710, Gangwon Do, South Korea
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2013年 / 713卷
基金
新加坡国家研究基金会;
关键词
SPECT; Cadmium telluride (CdTe) pixelated semiconductor detector; Ultra-high-resolution parallel-hole collimator materials; Monte Carlo simulation; MONTE-CARLO SIMULATIONS; GAMMA-CAMERA; PINHOLE SPECT; GATE; PERFORMANCE; TOMOGRAPHY; VALIDATION; DESIGN; IMAGER;
D O I
10.1016/j.nima.2013.03.014
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Recently, many studies have sought to improve the sensitivity and spatial resolution of pixelated semiconductor detectors. Spatial resolution can be improved by using a pinhole or pixelated parallel-hole collimator with equal hole and pixel sizes. We compared a pinhole to a pixelated parallel-hole collimator and found that the pixelated parallel-hole collimator had higher sensitivity. Additionally, collimator materials with high absorption efficiency are often used because of their high spatial resolution. The purpose of this study was to compare the quality of images generated using a pixelated semiconductor single photon emission computed tomography (SPECT) system simulated with pixelated parallel-hole collimators of lead, tungsten, gold, and depleted uranium. We performed a simulation study of the PID 350 (Ajat Oy Ltd., Finland) CdTe pixelated semiconductor detector, which consists of small pixels (0.35 x 0.35 mm(2)), using a Geant4 Application for Tomographic Emission (GATE) simulation. Sensitivities and spatial resolutions were measured for the four collimator materials. To evaluate overall image performance, a hot-rod phantom was designed using GATE simulation. The results showed that with lead, sensitivity was 425%, 6.53%, and 1028% higher than with tungsten, gold, and depleted uranium, respectively. Spatial resolution using depleted uranium was 3.19%, 4.19%, and 8.01% better than that of gold, tungsten, and lead, respectively. Sensitivity and spatial resolution showed little difference among the four types of collimator materials tested. It was difficult to visually distinguish between the reconstructed images of the hot-rod phantom for different collimator materials. The results are promising for notable cost reductions in collimator manufacturing while avoiding impractical and rare materials. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:33 / 39
页数:7
相关论文
共 50 条
  • [11] Performance evaluation of high-resolution square parallel-hole collimators with a CZT room temperature pixelated semiconductor SPECT system: a Monte Carlo simulation study
    Lee, Y.
    Kang, W.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [12] Feasibility Study on an Ultra-High-Resolution SPECT With CdTe Detectors
    Ogawa, Koichi
    Muraishi, Masaaki
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 17 - 24
  • [13] Performance evaluation of a small CZT pixelated semiconductor gamma camera system with a newly designed stack-up parallel-hole collimator
    Lee, Youngjin
    Kim, Hee-Joung
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 794 : 54 - 61
  • [14] Comparison of a Multi-pinhole Stationary SPECT System and a Parallel-hole High Speed Rotational SPECT system
    Fujishiro, Yohei
    Murata, Kazumi
    Motomura, Nobutoku
    Ogawa, Koichi
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [15] Development of an ultra-high resolution SPECT system with a CdTe semiconductor detector
    Ogawa, Koichi
    Ohmura, Naoka
    Iida, Hirokazu
    Nakamura, Kayoko
    Nakahara, Tadaki
    Kubo, Atsushi
    ANNALS OF NUCLEAR MEDICINE, 2009, 23 (08) : 763 - 770
  • [16] Development of an ultra-high resolution SPECT system with a CdTe semiconductor detector
    Koichi Ogawa
    Naoka Ohmura
    Hirokazu Iida
    Kayoko Nakamura
    Tadaki Nakahara
    Atsushi Kubo
    Annals of Nuclear Medicine, 2009, 23 : 763 - 770
  • [17] Reduction in variability of dopamine transporter measures using SPECT and ultra-high-resolution fan beam collimators: comparison with parallel-hole collimators in healthy controls for a clinical reference database
    De Cristofaro, Maria Teresa R.
    Linguanti, Flavia
    Pepponi, Miriam
    Abenavoli, Elisabetta
    Ramat, Silvia
    Bessi, Valentina
    Passeri, Alessandro
    Alfarano, Vincenzo
    Sorbi, Sandro
    Sciagra, Roberto
    Berti, Valentina
    CLINICAL AND TRANSLATIONAL IMAGING, 2024, 12 (03) : 347 - 357
  • [18] High-resolution in vivo SPECT imaging of I-131 biodistributions using a rotating parallel-hole collimator.
    Trotter, DEG
    Bowsher, JE
    Stone, CD
    Jaszczak, RJ
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 34P - 35P
  • [19] A Numerical Study of Different Types of Collimators for a High-Resolution Preclinical CdTe Pixelated Semiconductor SPECT System
    Jeong, Hyun-Woo
    Kim, Jong Seok
    Bae, Se Young
    Seo, Kanghyen
    Kim, Seung Hun
    Kang, Seong Hyeon
    Shin, Dong Jin
    Lee, Chang-Lae
    Kim, Kyuseok
    Lee, Youngjin
    JOURNAL OF THE OPTICAL SOCIETY OF KOREA, 2016, 20 (06) : 663 - 668
  • [20] Development of ultra-high-resolution pinhole collimator for small-animal SPECT with CZT detector
    Nishi, Koudai
    Kobayashi, Masato
    Matsunari, Ichiro
    Miyazaki, Yoshiharu
    Shiba, Kazuhiro
    Kawai, Keiichi
    Kinuya, Seigo
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54