Effect of respiratory motion on cardiac defect contrast in myocardial perfusion SPECT: a physical phantom study

被引:5
作者
Kortelainen, Matti J. [1 ,2 ]
Koivumaki, Tuomas M. [3 ]
Vauhkonen, Marko J. [1 ]
Hakulinen, Mikko A. [1 ,2 ]
机构
[1] Univ Eastern Finland, Dept Appl Phys, Kuopio, Finland
[2] Kuopio Univ Hosp, Diagnost Imaging Ctr, Kuopio, Finland
[3] Cent Finland Cent Hosp, Dept Radiat Therapy, Jyvaskyla, Finland
基金
芬兰科学院;
关键词
SPECT; Myocardial perfusion imaging; Respiratory motion; Phantom; POSITRON-EMISSION-TOMOGRAPHY; ORDERED-SUBSETS; PET; RECONSTRUCTION; HEART; RESOLUTION; 3D;
D O I
10.1007/s12149-019-01335-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveCorrection for respiratory motion in myocardial perfusion imaging requires sorting of emission data into respiratory windows where the intra-window motion is assumed to be negligible. However, it is unclear how much intra-window motion is acceptable. The aim of this study was to determine an optimal value of intra-window residual motion.MethodsA custom-designed cardiac phantom was created and imaged with a standard dual-detector SPECT/CT system using Tc-99m as the radionuclide. Projection images were generated from the list-mode data simulating respiratory motion blur of several magnitudes from 0 (stationary phantom) to 20mm. Cardiac defect contrasts in six anatomically different locations, as well as myocardial perfusion of apex, anterior, inferior, septal and lateral walls, were measured at each motion magnitude. Stationary phantom data were compared to motion-blurred data. Two physicians viewed the images and evaluated differences in cardiac defect visibility and myocardial perfusion.ResultsSignificant associations were observed between myocardial perfusion in the anterior and inferior walls and respiratory motion. Defect contrasts were found to decline as a function of motion, but the magnitude of the decline depended on the location and shape of the defect. Defects located near the cardiac apex lost contrast more rapidly than those located on the anterior, inferior, septal and lateral wall. The contrast decreased by less than 5% at every location when the motion magnitude was 2mm or less. According to a visual evaluation, there were differences in myocardial perfusion if the magnitude of the motion was greater than 1mm, and there were differences in the visibility of the cardiac defect if the magnitude of the motion was greater than 9mm.ConclusionsIntra-window respiratory motion should be limited to 2mm to effectively correct for respiratory motion blur in myocardial perfusion SPECT.
引用
收藏
页码:305 / 316
页数:12
相关论文
共 40 条
[11]  
Faber TL, 1999, J NUCL MED, V40, P650
[12]   EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology [J].
Hesse, B ;
Tägil, K ;
Cuocolo, A ;
Anagnostopoulos, C ;
Bardiés, M ;
Bax, J ;
Bengel, F ;
Sokole, EB ;
Davies, G ;
Dondi, M ;
Edenbrandt, L ;
Franken, P ;
Kjaer, A ;
Knuuti, J ;
Lassmann, M ;
Ljungberg, M ;
Marcassa, C ;
Marie, PY ;
McKiddie, F ;
O'Connor, M ;
Prvulovich, E ;
Underwood, R ;
van Eck-Smit, B .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 (07) :855-897
[13]  
HIGLEY B, 1993, J NUCL MED, V34, P30
[14]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[15]   Data-driven respiratory motion tracking and compensation in CZT cameras: A comprehensive analysis of phantom and human images [J].
Ko, Chi-Lun ;
Wu, Yen-Wen ;
Cheng, Mei-Fang ;
Yen, Ruoh-Fang ;
Wu, Wen-Chau ;
Tzen, Kai-Yuan .
JOURNAL OF NUCLEAR CARDIOLOGY, 2015, 22 (02) :308-318
[16]   Dual gated PET/CT imaging of small targets of the heart: Method description and testing with a dynamic heart phantom [J].
Kokki, Tommi ;
Sipila, Hannu T. ;
Teras, Mika ;
Noponen, Tommi ;
Durand-Schaefer, Nicolas ;
Klen, Riku ;
Knuuti, Juhani .
JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (01) :71-84
[17]   Respiratory motion reduction with a dual gating approach in myocardial perfusion SPECT: Effect on left ventricular functional parameters [J].
Kortelainen, Matti J. ;
Koivumaki, Tuomas M. ;
Vauhkonen, Marko J. ;
Hedman, Marja K. ;
Karkkainen, Satu T. J. ;
Quintero, Juanita Nino ;
Hakulinen, Mikko A. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2018, 25 (05) :1633-1641
[18]   Correction of heart motion due to respiration in clinical myocardial perfusion SPECT scans using respiratory gating [J].
Kovalski, Gil ;
Israel, Ora ;
Keidar, Zohar ;
Frenkel, Alex ;
Sachs, Jonathan ;
Azhari, Haim .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (04) :630-636
[19]   Organ Movement Reduction in PET/CT using Dual-Gated Listmode Acquisition [J].
Lang, Norbert ;
Dawood, Mohammad ;
Buether, Florian ;
Schober, Otmar ;
Schaefers, Michael ;
Schaefers, Klaus .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2006, 16 (01) :93-100
[20]   Recommendations for chamber quantification [J].
Lang, Roberto M. ;
Bierig, Michelle ;
Devereux, Richard B. ;
Flachskampf, Frank A. ;
Foster, Elyse ;
Pellikka, Patricia A. ;
Picard, Michael H. ;
Roman, Mary J. ;
Seward, James ;
Shanewise, Jack ;
Solomon, Scott ;
Spencer, Kirk T. ;
Sutton, Martin St. John ;
Stewart, William .
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2006, 7 (02) :79-108