Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides for SPECT

被引:71
作者
Brown, Saxby [1 ,2 ]
Bailey, Dale L. [1 ,3 ]
Willowson, Kathy [1 ,3 ]
Baldock, Clive [1 ]
机构
[1] Univ Sydney, Inst Med Phys, Sch Phys, Sydney, NSW 2006, Australia
[2] Royal N Shore Hosp, No Sydney Canc Ctr, Dept Radiat Oncol, Sydney, NSW 2065, Australia
[3] Royal N Shore Hosp, Dept Nucl Med, Sydney, NSW 2065, Australia
关键词
radionuclide; CT; SPECT; attenuation correction;
D O I
10.1016/j.apradiso.2008.01.002
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
This study has investigated the relationship between linear attenuation coefficients (mu) and Hounsfield units (HUs) for six materials covering the range of values found clinically. Narrow-beam mu values were measured by performing radionuclide transmission scans using Tc-99m, I-123, I-131, Tl-201 and In-111. The mu values were compared to published data. The relationships between mu and HU were determined. These relationships can be used to convert computed tomography (CT) images to mu-maps for single photon emission computed tomography (SPECT) attenuation correction. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1206 / 1212
页数:7
相关论文
共 14 条
[1]   A generalized model for the conversion from CT numbers to linear attenuation coefficients [J].
Bai, CY ;
Shao, L ;
Da Silva, AJ ;
Zhao, Z .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) :1510-1515
[2]   Transmission scanning in emission tomography [J].
Bailey, DL .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (07) :774-787
[3]   Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment [J].
Blankespoor, SC ;
Wu, X ;
Kalki, K ;
Brown, JK ;
Tang, HR ;
Cann, CE ;
Hasegawa, BH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (04) :2263-2274
[4]   Experimental study of attenuation properties of normoxic polymer gel dosimeters [J].
Brindha, S ;
Venning, AJ ;
Hill, B ;
Baldock, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (20) :N353-N361
[5]   PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients [J].
Burger, C ;
Goerres, G ;
Schoenes, S ;
Buck, A ;
Lonn, AHR ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :922-927
[6]   Role of apoptosis in ventricular remodeling. [J].
Chandrashekhar Y. .
Current Heart Failure Reports, 2005, 2 (1) :18-22
[8]  
GELDENHUYS EM, 1988, J NUCL MED, V29, P538
[9]  
JOHNS HE, 1983, PHYS RADIOLOGY, P137
[10]   INVESTIGATION OF THE USE OF X-RAY CT IMAGES FOR ATTENUATION COMPENSATION IN SPECT [J].
LACROIX, KJ ;
TSUI, BMW ;
HASEGAWA, BH ;
BROWN, JK .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (06) :2793-2799