Electrochemical impedance spectroscopy biosensor with interdigitated electrode for detection of human immunoglobulin A

被引:109
作者
Ohno, Ryuzo [1 ]
Ohnuki, Hitoshi [1 ]
Wang, Huihui [1 ]
Yokoyama, Takuya [1 ]
Endo, Hideaki [2 ]
Tsuya, Daiju [3 ]
Izumi, Mitsuru [1 ]
机构
[1] Tokyo Univ Marine Sci & Technol, Div Marine Technol, Koto Ku, Tokyo 1358533, Japan
[2] Tokyo Univ Marine Sci & Technol, Div Marine Sci, Minato Ku, Tokyo 1088477, Japan
[3] Natl Inst Mat Sci, Tsukuba, Ibaraki 3050047, Japan
关键词
Electrochemical impedance spectroscopy; Interdigitated microelectrode; Immunosensor; Human immunoglobulin A; IMMUNOSENSOR;
D O I
10.1016/j.bios.2012.07.052
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Interdigitated electrodes (IDEs) that have a series of parallel microband electrodes with alternating microbands connected together were utilized in electrochemical impedance spectroscopy (EIS) to build a label-free human immunoglobulin A (IgA) immunosensor. Anti-human IgA (anti-IgA) was employed as an IgA receptor and was covalently immobilized on the IDE surface through a self-assembled monolayer, as confirmed by atomic force microscopy. EIS measurements revealed that the specific adsorption of IgA onto the immobilized anti-IgA gave rise to a clear increase in the value of interfacial charge transfer resistance (R-ct). A linear relationship between Delta R-ct and the logarithm of IgA concentration was found for the concentration range of 0.01-100 ng/mL. No modulation of R-ct was detected by immersing the sensor in solutions of other proteins such as human immunoglobulin G or bovine serum albumin, which confirmed a high selectivity of this immunosensor for IgA. These results demonstrated that the anti-IgA receptor simply immobilized on the IDE surface can provide a sensitive biosensor. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:422 / 426
页数:5
相关论文
共 17 条
[1]   Antibody functionalized interdigitated μ-electrode (IDμE) based impedimetric cortisol biosensor [J].
Arya, Sunil K. ;
Chornokur, Ganna ;
Venugopal, Manju ;
Bhansali, Shekhar .
ANALYST, 2010, 135 (08) :1941-1946
[2]   Challenges of Electrochemical Impedance Spectroscopy in Protein Biosensing [J].
Bogomolova, A. ;
Komarova, E. ;
Reber, K. ;
Gerasimov, T. ;
Yavuz, O. ;
Bhatt, S. ;
Aldissi, M. .
ANALYTICAL CHEMISTRY, 2009, 81 (10) :3944-3949
[3]   Electrochemical immunosensor based on colloidal carbon sphere array [J].
Chen, Xiaojun ;
Zhang, Kui ;
Zhou, Jinjun ;
Xuan, Jie ;
Yan, Wei ;
Jiang, Li-Ping ;
Zhu, Jun-Jie .
BIOSENSORS & BIOELECTRONICS, 2010, 25 (05) :1130-1136
[4]   Label-free impedance biosensors: Opportunities and challenges [J].
Daniels, Jonathan S. ;
Pourmand, Nader .
ELECTROANALYSIS, 2007, 19 (12) :1239-1257
[5]  
Gosling J.P., 1997, PRINCIPLES PRACTICE, P349
[6]  
Hermanson GT, 2008, BIOCONJUGATE TECHNIQUES, 2ND EDITION, P1, DOI 10.1016/B978-0-12-370501-3.00001-1
[7]   Increases in serum immunoglobulins to age-related normal levels in children with IgA and/or IgG subclass deficiency [J].
Kutukculer, Necil ;
Karaca, Neslihan Edeer ;
Demircioglu, Ozlem ;
Aksu, Guzide .
PEDIATRIC ALLERGY AND IMMUNOLOGY, 2007, 18 (02) :167-173
[8]   Planar electrochemical sensors for biomedical applications [J].
Laschi, Serena ;
Mascini, Marco .
MEDICAL ENGINEERING & PHYSICS, 2006, 28 (10) :934-943
[9]   Enzyme Immunoassay (EIA)/Enzyme-Linked Immunosorbent Assay (ELISA) [J].
Lequin, RM .
CLINICAL CHEMISTRY, 2005, 51 (12) :2415-2418
[10]   Electrochemical detection of single-nucleotide mismatches using an electrode microarray [J].
Li, Xiaohong ;
Lee, Jeremy S. ;
Kraatz, Heinz-Bernhard .
ANALYTICAL CHEMISTRY, 2006, 78 (17) :6096-6101