Efficacy of the active middle-ear implant in patients with sensorineural hearing loss

被引:17
作者
Butler, C. L. [1 ]
Thavaneswaran, P. [1 ]
Lee, I. H. [1 ]
机构
[1] Royal Australasian Coll Surg, Australian Safety & Efficacy Register New Interve, Adelaide, SA, Australia
关键词
Ear; Middle; Ossicular Prosthesis; Hearing Aids; Hearing Loss; Review; VIBRANT SOUNDBRIDGE; OSSICULAR STIMULATOR; DEVICE; IMPAIRMENT; AIDS; MULTICENTER; EXPERIENCE; CHILDREN; SYSTEM;
D O I
10.1017/S0022215113001151
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Introduction: This systematic review aims to advise on the effectiveness of the active middle-ear implant in patients with sensorineural hearing loss, compared with external hearing aids. Methods: A systematic search of several electronic databases, including PubMed and Embase, was used to identify relevant studies for inclusion. Results: Fourteen comparative studies were included. Nine studies reported on the primary outcome of functional gain: one found that the middle-ear implant was significantly better than external hearing aids (p < 0.001), while another found that external hearing aids were generally significantly better than middle-ear implants (p < 0.05). Six of the seven remaining studies found that middle-ear implants were better than external hearing aids, although generally no clinically significant difference (i.e. >= 10 dB) was seen. Conclusion: Generally, the active middle-ear implant appears to be as effective as the external hearing aid in improving hearing outcomes in patients with sensorineural hearing loss.
引用
收藏
页码:S8 / S16
页数:9
相关论文
共 50 条
[31]   Active Middle Ear Implantation for Patients With Sensorineural Hearing Loss and External Otitis: Long-Term Outcome in Patient Satisfaction [J].
Zwartenkot, Joost W. ;
Hashemi, Javad ;
Cremers, Cor W. R. J. ;
Mulder, Jef J. S. ;
Snik, Ad F. M. .
OTOLOGY & NEUROTOLOGY, 2013, 34 (05) :855-861
[32]   Round Window Application of an Active Middle Ear Implant: A Comparison With Hearing Aid Usage in Japan [J].
Iwasaki, Satoshi ;
Usami, Shin-ichi ;
Takahashi, Haruo ;
Kanda, Yukihiko ;
Tono, Tetsuya ;
Doi, Katsumi ;
Kumakawa, Kozo ;
Gyo, Kiyofumi ;
Naito, Yasushi ;
Kanzaki, Sho ;
Yamanaka, Noboru ;
Kaga, Kimitaka .
OTOLOGY & NEUROTOLOGY, 2017, 38 (06) :E145-E151
[33]   Standardized Active Middle-Ear Implant Coupling to the Short Incus Process [J].
Mlynski, Robert ;
Dalhoff, Ernst ;
Heyd, Andreas ;
Wildenstein, Daniela ;
Rak, Kristen ;
Radeloff, Andreas ;
Hagen, Rudolf ;
Gummer, Anthony W. ;
Schraven, Sebastian P. .
OTOLOGY & NEUROTOLOGY, 2015, 36 (08) :1390-1398
[34]   Piezoelectric Actuator with Frequency Characteristics for a Middle-Ear Implant [J].
Shin, Dong Ho ;
Cho, Jin-Ho .
SENSORS, 2018, 18 (06)
[35]   Clinical effect of an active transcutaneous bone-conduction implant on tinnitus in patients with ipsilateral sensorineural hearing loss [J].
Lee, Hyun-Jin ;
Kahinga, Aveline Aloyce ;
Moon, In Seok .
AURIS NASUS LARYNX, 2021, 48 (03) :394-399
[36]   Application of Active Middle Ear Implants in Patients With Severe Mixed Hearing Loss [J].
Verhaegen, Veronique J. O. ;
Mulder, Jef J. S. ;
Cremers, Cor W. R. J. ;
Snik, Ad F. M. .
OTOLOGY & NEUROTOLOGY, 2012, 33 (03) :297-301
[37]   Economic benefits of the cochlear implant for treating profound sensorineural hearing loss [J].
Penaranda, Augusto ;
Carlos Mendieta, Juan ;
Andres Perdomo, Jorge ;
Leonor Aparicio, Maria ;
Marcela Marin, Lina ;
Manuel Garcia, Juan ;
Baron, Clemencia .
REVISTA PANAMERICANA DE SALUD PUBLICA-PAN AMERICAN JOURNAL OF PUBLIC HEALTH, 2012, 31 (04) :325-331
[38]   Congenital defects of the middle ear - uncommon cause of pediatric hearing loss [J].
Sena Esteves, Sara Duarte ;
da Silva, Ana Pereira ;
Coutinho, Miguel Bebiano ;
Abrunhosa, Jose Manuel ;
Almeida e Sousa, Cecilia .
BRAZILIAN JOURNAL OF OTORHINOLARYNGOLOGY, 2014, 80 (03) :251-256
[39]   Systematic review to evaluate the safety, efficacy and economical outcomes of the Vibrant Soundbridge for the treatment of sensorineural hearing loss [J].
Bruchhage, Karl-Ludwig ;
Leichtle, Anke ;
Schonweiler, Rainer ;
Todt, Ingo ;
Baumgartner, Wolf-Dieter ;
Frenzel, Henning ;
Wollenberg, Barbara .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2017, 274 (04) :1797-1806
[40]   Coupling of an active middle-ear implant to the long process of the incus using an elastic clip attachment [J].
Schraven, Sebastian P. ;
Mlynski, Robert ;
Dalhoff, Ernst ;
Heyd, Andreas ;
Wildenstein, Daniela ;
Rak, Kristen ;
Radeloff, Andreas ;
Hagen, Rudolf ;
Gummer, Anthony W. .
HEARING RESEARCH, 2016, 340 :179-184