Efficacy of the active middle-ear implant in patients with sensorineural hearing loss

被引:17
作者
Butler, C. L. [1 ]
Thavaneswaran, P. [1 ]
Lee, I. H. [1 ]
机构
[1] Royal Australasian Coll Surg, Australian Safety & Efficacy Register New Interve, Adelaide, SA, Australia
关键词
Ear; Middle; Ossicular Prosthesis; Hearing Aids; Hearing Loss; Review; VIBRANT SOUNDBRIDGE; OSSICULAR STIMULATOR; DEVICE; IMPAIRMENT; AIDS; MULTICENTER; EXPERIENCE; CHILDREN; SYSTEM;
D O I
10.1017/S0022215113001151
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Introduction: This systematic review aims to advise on the effectiveness of the active middle-ear implant in patients with sensorineural hearing loss, compared with external hearing aids. Methods: A systematic search of several electronic databases, including PubMed and Embase, was used to identify relevant studies for inclusion. Results: Fourteen comparative studies were included. Nine studies reported on the primary outcome of functional gain: one found that the middle-ear implant was significantly better than external hearing aids (p < 0.001), while another found that external hearing aids were generally significantly better than middle-ear implants (p < 0.05). Six of the seven remaining studies found that middle-ear implants were better than external hearing aids, although generally no clinically significant difference (i.e. >= 10 dB) was seen. Conclusion: Generally, the active middle-ear implant appears to be as effective as the external hearing aid in improving hearing outcomes in patients with sensorineural hearing loss.
引用
收藏
页码:S8 / S16
页数:9
相关论文
共 50 条
  • [21] Application of an active middle ear implant in congenital middle ear malformations: A contemporary review
    da Silva, Vagner Antonio Rodrigues
    Pauna, Henrique Furlan
    Guimaraes, Guilherme Correa
    Lavinsky, Joel
    Linder, Thomas E.
    Castilho, Arthur Menino
    [J]. BRAZILIAN JOURNAL OF OTORHINOLARYNGOLOGY, 2025, 91 (03)
  • [22] Middle ear implant in conductive and mixed congenital hearing loss in children
    Roman, Stephane
    Denoyelle, Francoise
    Farinetti, Anne
    Garabedian, Erea-Noel
    Triglia, Jean-Michel
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2012, 76 (12) : 1775 - 1778
  • [23] Functional results with active middle ear implant or semi-implantable bone conduction device in patients with comparable hearing loss
    Spiegel, Jennifer L.
    Weiss, Bernhard G.
    Bertlich, Mattis
    Stoycheva, Ivelina
    Canis, Martin
    Ihler, Friedrich
    [J]. INTERNATIONAL JOURNAL OF AUDIOLOGY, 2022, 61 (10) : 859 - 867
  • [24] COMPARISON OF METHODS OF EVALUATING HEARING BENEFIT OF MIDDLE-EAR SURGERY
    TONER, JG
    SMYTH, GDL
    [J]. JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1993, 107 (01) : 4 - 5
  • [25] Prevalence of middle ear pathologies in children with bilateral sensorineural hearing loss
    Abou-Elhamd, Kamal-Eldin Ahmed
    Moussa, Abd-Elmateen
    Soltan, Mohammed Abd-Elkader
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2006, 70 (06) : 1081 - 1084
  • [26] Experience With the Active Middle Ear Implant in Patients With Moderate-to-Severe Mixed Hearing Loss: Indications and Results
    de Abajo, Jorge
    Sanhueza, Ignacio
    Giron, Laura
    Manrique, Manuel
    [J]. OTOLOGY & NEUROTOLOGY, 2013, 34 (08) : 1373 - 1379
  • [27] The totally implantable middle ear device "Esteem" for rehabilitation of severe sensorineural hearing loss
    Barbara, Maurizio
    Biagini, Michela
    Monini, Simonetta
    [J]. ACTA OTO-LARYNGOLOGICA, 2011, 131 (04) : 399 - 404
  • [28] Endoscopic middle ear exploration in pediatric patients with conductive hearing loss
    Carter, John M.
    Hoff, Stephen R.
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2017, 96 : 21 - 24
  • [29] Active Middle Ear Implantation for Patients With Sensorineural Hearing Loss and External Otitis: Long-Term Outcome in Patient Satisfaction
    Zwartenkot, Joost W.
    Hashemi, Javad
    Cremers, Cor W. R. J.
    Mulder, Jef J. S.
    Snik, Ad F. M.
    [J]. OTOLOGY & NEUROTOLOGY, 2013, 34 (05) : 855 - 861
  • [30] Round Window Application of an Active Middle Ear Implant: A Comparison With Hearing Aid Usage in Japan
    Iwasaki, Satoshi
    Usami, Shin-ichi
    Takahashi, Haruo
    Kanda, Yukihiko
    Tono, Tetsuya
    Doi, Katsumi
    Kumakawa, Kozo
    Gyo, Kiyofumi
    Naito, Yasushi
    Kanzaki, Sho
    Yamanaka, Noboru
    Kaga, Kimitaka
    [J]. OTOLOGY & NEUROTOLOGY, 2017, 38 (06) : E145 - E151