Cell analysis in a microfluidic cytorneter applying a DVD pickup head

被引:22
作者
Kostner, S. [1 ]
Vellekoop, M. J. [1 ]
机构
[1] Vienna Univ Technol, Inst Sensor & Actuator Syst, A-1040 Vienna, Austria
关键词
single cell; particle; cytometry; DVD; pickup;
D O I
10.1016/j.snb.2007.11.038
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper we present an optical detection system for single biological cells that utilizes a standard DVD pickup head. In a feasibility study we have shown the detection for different particles and cells (yeast) suspended on a platinum mirror. For those experiments the integrated magnetic actuator of the pickup was used to simulate the particle movement and we found a remarkable sensitivity. Then a microfluidic flow cell with an integrated mirror was designed to be able to measure particles in a cytometric setup. The mirror at the bottom of the flow channel reflects the focused laser beam back to the integrated photodiodes. The advantage of using a DVD pickup head is not only in the low cost but also in the ability to use the integrated magnetic actuators to adjust for alignment errors between flow channel and pickup. Measurement results are given that show the high sensitivity and good repeatability for polystyrene beads and erythrocytes from cattle. In a two-parameter plot the cytometric separation of a mixture of counting beads and erythrocytes into two sub-populations is demonstrated. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:512 / 517
页数:6
相关论文
共 21 条
[1]   Expansion channel for microchip flow cytometers [J].
Bang, Hyunwoo ;
Yun, Hoyoung ;
Lee, Won Gu ;
Park, Junha ;
Lee, Jooruno ;
Chung, Seok ;
Cho, Keunchang ;
Chung, Chanil ;
Han, Dong-Chul ;
Chang, Jun Kenn .
LAB ON A CHIP, 2006, 6 (10) :1381-1383
[2]   Impedance spectroscopy flow cytometry: On-chip label-free cell differentiation [J].
Cheung, K ;
Gawad, S ;
Renaud, P .
CYTOMETRY PART A, 2005, 65A (02) :124-132
[3]   Microfluidic impedance spectroscopy flow cytometer: Particle size calibration [J].
Cheung, K ;
Gawad, S ;
Renaud, P .
MEMS 2004: 17TH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2004, :343-346
[4]   Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing [J].
Gawad, S ;
Schild, L ;
Renaud, P .
LAB ON A CHIP, 2001, 1 (01) :76-82
[5]  
Givan, 2001, FLOW CYTOMETRY 1 PRI, DOI 10.1002/0471223948
[6]   Microfluidics for flow cytometric analysis of cells and particles [J].
Huh, D ;
Gu, W ;
Kamotani, Y ;
Grotberg, JB ;
Takayama, S .
PHYSIOLOGICAL MEASUREMENT, 2005, 26 (03) :R73-R98
[7]   Design and fabrication of a micromachined Coulter counter [J].
Koch, M ;
Evans, AGR ;
Brunnschweiler, A .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1999, 9 (02) :159-161
[8]  
KOSTNER S, 2007, P 14 INT C SOL STAT, P2123
[9]  
Lee DW, 2005, PROC IEEE MICR ELECT, P678
[10]   The hydrodynamic focusing effect inside rectangular microchannels [J].
Lee, Gwo-Bin ;
Chang, Chih-Chang ;
Huang, Sung-Bin ;
Yang, Ruey-Jen .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2006, 16 (05) :1024-1032