Modeling the temporal-spatial nature of the readout of an electronic portal imaging device (EPID)

被引:2
|
作者
Abbasian, Parandoush [1 ]
McCowan, Peter M. [1 ,2 ]
Rickey, Daniel W. [1 ,2 ,3 ]
Van Uytven, Eric [2 ,3 ]
McCurdy, Boyd M. C. [1 ,2 ,3 ]
机构
[1] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[2] CancerCare Manitoba, Med Phys Dept, 675 McDermot Ave, Winnipeg, MB R3E 0V9, Canada
[3] Univ Manitoba, Dept Radiol, 820 Sherbrook St, Winnipeg, MB R3A 1R9, Canada
关键词
banding artifacts; dosimetry; EPID; radiotherapy; real-time dosimetry; QUALITY-ASSURANCE; RADIATION-THERAPY; DOSIMETRY; IMRT; IMAGES; ERRORS;
D O I
10.1002/mp.14440
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In real-time electronic portal imaging device (EPID) dosimetry applications where on-treatment measured transmission images are compared to an ideal predicted image, ideally a tight tolerance should be set on the quantitative image comparison in order to detect a wide variety of possible delivery errors. However, this is currently not possible due to the appearance of banding artifacts in individual frames of the measured EPID image sequences. The purpose of this work was to investigate simulating banding artifacts in our cine-EPID predicted image sequences to improve matching of individual image frames to the acquired image sequence. Increased sensitivity of this method to potential treatment delivery errors would represent an improvement in patient safety and treatment accuracy. Methods A circuit board was designed and built to capture the target current (TARG-I) and forward power signals produced by the linac to help model the discrete beam-formation process of the linac. To simulate the temporal-spatial nature of the EPID readout, a moving read out mask was applied with the timing of the application of the readout mask synchronized to the TARG-I pulses. Since identifying the timing of the first TARG-I pulse affected the location of the banding artifacts throughout the image sequence, and furthermore the first several TARG-I pulses at the beginning of "beam on" are not at full height yet (i.e., dose rate is ramping up), the forward-power signal was also used to assist in reliable detection of the first radiation pulse of the beam delivery. The predicted EPID cine-image sequence obtained using a comprehensive physics-based model was modified to incorporate the discrete nature of the EPID frame readout. This modified banding predicted EPID (MBP-EPID) image sequence was then compared to its corresponding measured EPID cine-image sequence on a frame-by-frame basis. The EPID was mounted on a Clinac 2100ix linac (Varian Medical Systems, Palo Alto, CA). The field size was set to 21.4 x 28.6 cm(2)with no MLC modulation, beam energy of 6 MV, dose rate of 600 MU/min, and 700 MU were delivered for each clockwise (CW) and counter-clockwise (CCW) arc. No phantoms were placed in the beam. Results The dose rate ramp up effect was observed at the beginning irradiations, and the identification and timing of the radiation pulses, even during the dose rate ramp up, were able to be quantified using the TARG-I and forward power signals. The approach of capturing individual dose pulses and synchronizing with the mask image applied to the original predicted EPID image sequence was demonstrated to model the actual EPID readout. The MBP-EPID image sequences closely reproduced the location and magnitude of the banding features observed in the acquired (i.e., measured) image sequence, for all test irradiations examined here. Conclusions The banding artifacts observed in the measured EPID cine-frame sequences were reproduced in the predicted EPID cine-frames by simulating the discrete temporal-spatial nature of the EPID read out. The MBP-EPID images showed good agreement qualitatively to the corresponding measured EPID frame sequence of a simple square test field, without any phantom in the beam. This approach will lead to improved image comparison tolerances for real-time patient dosimetry applications.
引用
收藏
页码:5301 / 5311
页数:11
相关论文
共 50 条
  • [21] Clinical Feasibility of Using an Electronic Portal Imaging Device for Position Verification during Conventional Radiotherapy
    Shen, Yuxiaotong
    Zhang, Jie
    Ge, Yun
    Chen, Ying
    Li, Haiwei
    Sun, Wei
    Ji, Mingxi
    Wei, Quanbo
    Cai, Jing
    Li, Bing
    2016 IEEE 10TH INTERNATIONAL CONFERENCE ON NANO/MOLECULAR MEDICINE AND ENGINEERING (NANOMED), 2016, : 99 - 103
  • [22] Development and testing of an improved dosimetry system using a backscatter shielded electronic portal imaging device
    King, Brian W.
    Morf, Daniel
    Greer, Peter B.
    MEDICAL PHYSICS, 2012, 39 (05) : 2839 - 2847
  • [23] Simple Proposal for Dosimetry with an Elekta iViewGT™ Electronic Portal Imaging Device (EPID) Using Commercial Software Modules
    Liebich, Janett
    Licher, Joerg
    Scherf, Christian
    Kara, Eugen
    Koch, Nadine
    Roedel, Claus
    Ramm, Ulla
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (05) : 316 - 321
  • [24] Comparison of pretreatment VMAT quality assurance with the integral quality monitor (IQM) and electronic portal imaging device (EPID)
    Ghafarian, Melissa
    Price, Michael
    Morales-Paliza, Manuel
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (03): : 166 - 175
  • [25] Evaluation of set-up errors and determination of set-up margin in pelvic radiotherapy by electronic portal imaging device (EPID)
    Noghreiyan, Vajiheh Vejdani
    Nasseri, Shahrookh
    Anvari, Kazem
    Naji, Maryam
    Momennezhad, Mehdi
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2020, 19 (02) : 150 - 156
  • [26] Imaging linear accelerator output using a high-speed scintillation based electronic portal imaging device (Hi-EPID)
    Ranade, Manisha
    Lynch, Bart
    Li, Jonathan
    Kim, Siyong
    Dempsey, James
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1811 - +
  • [27] In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion
    Kroonwijk, M
    Pasma, KL
    Quint, S
    Koper, PCM
    Visser, AG
    Heijmen, BJM
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) : 125 - 132
  • [28] Intensity modulated radiation therapy (IMRT) verification using the frame averaging feature of Varian Mark II electronic portal imaging device (EPID)
    Chang, J
    Mageras, G
    Ling, CC
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2059 - 2062
  • [29] An Efficient Way to Confirm the Coincidence of SRS Collimator and Radiation Beam Axes Using Electronic Portal Imaging Device(EPID)
    Kang, D. H.
    Huang, S. H. M.
    Deng, X. W.
    Zhou, Y. J.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1841 - +
  • [30] A radiation-tolerant electronic readout system for portal imaging
    Östling, J
    Brahme, A
    Danielsson, M
    Iacobaeus, C
    Peskov, V
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2) : 308 - 312