Modeling the temporal-spatial nature of the readout of an electronic portal imaging device (EPID)

被引:2
|
作者
Abbasian, Parandoush [1 ]
McCowan, Peter M. [1 ,2 ]
Rickey, Daniel W. [1 ,2 ,3 ]
Van Uytven, Eric [2 ,3 ]
McCurdy, Boyd M. C. [1 ,2 ,3 ]
机构
[1] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[2] CancerCare Manitoba, Med Phys Dept, 675 McDermot Ave, Winnipeg, MB R3E 0V9, Canada
[3] Univ Manitoba, Dept Radiol, 820 Sherbrook St, Winnipeg, MB R3A 1R9, Canada
关键词
banding artifacts; dosimetry; EPID; radiotherapy; real-time dosimetry; QUALITY-ASSURANCE; RADIATION-THERAPY; DOSIMETRY; IMRT; IMAGES; ERRORS;
D O I
10.1002/mp.14440
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In real-time electronic portal imaging device (EPID) dosimetry applications where on-treatment measured transmission images are compared to an ideal predicted image, ideally a tight tolerance should be set on the quantitative image comparison in order to detect a wide variety of possible delivery errors. However, this is currently not possible due to the appearance of banding artifacts in individual frames of the measured EPID image sequences. The purpose of this work was to investigate simulating banding artifacts in our cine-EPID predicted image sequences to improve matching of individual image frames to the acquired image sequence. Increased sensitivity of this method to potential treatment delivery errors would represent an improvement in patient safety and treatment accuracy. Methods A circuit board was designed and built to capture the target current (TARG-I) and forward power signals produced by the linac to help model the discrete beam-formation process of the linac. To simulate the temporal-spatial nature of the EPID readout, a moving read out mask was applied with the timing of the application of the readout mask synchronized to the TARG-I pulses. Since identifying the timing of the first TARG-I pulse affected the location of the banding artifacts throughout the image sequence, and furthermore the first several TARG-I pulses at the beginning of "beam on" are not at full height yet (i.e., dose rate is ramping up), the forward-power signal was also used to assist in reliable detection of the first radiation pulse of the beam delivery. The predicted EPID cine-image sequence obtained using a comprehensive physics-based model was modified to incorporate the discrete nature of the EPID frame readout. This modified banding predicted EPID (MBP-EPID) image sequence was then compared to its corresponding measured EPID cine-image sequence on a frame-by-frame basis. The EPID was mounted on a Clinac 2100ix linac (Varian Medical Systems, Palo Alto, CA). The field size was set to 21.4 x 28.6 cm(2)with no MLC modulation, beam energy of 6 MV, dose rate of 600 MU/min, and 700 MU were delivered for each clockwise (CW) and counter-clockwise (CCW) arc. No phantoms were placed in the beam. Results The dose rate ramp up effect was observed at the beginning irradiations, and the identification and timing of the radiation pulses, even during the dose rate ramp up, were able to be quantified using the TARG-I and forward power signals. The approach of capturing individual dose pulses and synchronizing with the mask image applied to the original predicted EPID image sequence was demonstrated to model the actual EPID readout. The MBP-EPID image sequences closely reproduced the location and magnitude of the banding features observed in the acquired (i.e., measured) image sequence, for all test irradiations examined here. Conclusions The banding artifacts observed in the measured EPID cine-frame sequences were reproduced in the predicted EPID cine-frames by simulating the discrete temporal-spatial nature of the EPID read out. The MBP-EPID images showed good agreement qualitatively to the corresponding measured EPID frame sequence of a simple square test field, without any phantom in the beam. This approach will lead to improved image comparison tolerances for real-time patient dosimetry applications.
引用
收藏
页码:5301 / 5311
页数:11
相关论文
共 50 条
  • [1] Reliability analysis of an Electronic Portal Imaging Device (EPID) of a linear accelerator
    Gouveia, Clesia
    Sobral, Jose
    2023 IEEE 7TH PORTUGUESE MEETING ON BIOENGINEERING, ENBENG, 2023, : 52 - 55
  • [2] Transit Dosimetry for Verification of Dose Delivery Using Electronic Portal Imaging Device (EPID)
    Baek, T.
    Chung, E.
    Lee, S.
    Yoon, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 301 - 301
  • [3] An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool
    Ma, Yangguang
    Wang, Xuemin
    Mai, Rizhen
    Wang, Tao
    Pei, Yuntong
    Liu, Shuaipeng
    Guo, Yuexin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 45 - 58
  • [4] Development of an Electronic Portal Imaging Device Dosimetry Method
    Zhang, Jun
    Fan, Ziting
    Zhang, Xile
    Yang, Ruijie
    Wen, Junhai
    DIAGNOSTICS, 2021, 11 (09)
  • [5] The effect of respiratory motion on electronic portal imaging device dosimetry
    Fielding, Andrew L.
    Mendieta, Jessica Benitez
    Maxwell, Sarah
    Jones, Catherine
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (03): : 45 - 55
  • [6] Transit dosimetry with an electronic portal imaging device (EPID) for 115 prostate cancer patients
    Pasma, KL
    Kroonwijk, M
    Quint, S
    Visser, AG
    Heijmen, BJM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (05): : 1297 - 1303
  • [7] A Clinical Objective IMRT QA Method Based on Portal Dosimetry and Electronic Portal Imager Device (EPID) Measurement
    Zhang, Miao
    Qin, Songbing
    Chen, Ting
    Kim, Sung
    Jabbour, Salma
    Haffty, Bruce
    Yue, Ning J.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2013, 12 (02) : 145 - 150
  • [8] The linear accelerator mechanical and radiation isocentre assessment with an electronic portal imaging device (EPID)
    Liu G.
    van Doorn T.
    Bezak E.
    Australasian Physics & Engineering Sciences in Medicine, 2004, 27 (3): : 111 - 117
  • [9] Verification of radiosurgery target point alignment with an electronic portal imaging device (EPID)
    Dong, L
    Shiu, A
    Tung, S
    Boyer, A
    MEDICAL PHYSICS, 1997, 24 (02) : 263 - 267
  • [10] Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance
    Huang, Yen-Cho
    Yeh, Chien-Yi
    Yeh, Jih-Hsiang
    Lo, Ching-Jung
    Tsai, Ping-Fang
    Hung, Chih-Hung
    Tsai, Chieh-Sheng
    Chen, Chen-Yuan
    MEDICAL DOSIMETRY, 2013, 38 (01) : 35 - 41