The influence of stem taper re-use upon the failure load of ceramic heads

被引:14
作者
Guehrs, Julian [1 ]
Krull, Annika [1 ]
Witt, Florian [1 ]
Morlock, Michael M. [1 ]
机构
[1] TUHH Hamburg Univ Technol, D-21073 Hamburg, Germany
关键词
Total hip replacement; Stem taper re-use; Ceramic bearings; Fracture load; Head-taper connection; TOTAL HIP-ARTHROPLASTY; FEMORAL HEADS; FRACTURE; REVISION; METAL; BEARINGS; CORROSION; STRENGTH; JUNCTION;
D O I
10.1016/j.medengphy.2015.03.010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ceramic components are frequently used in total hip replacement due to their good tribological properties. In revision of broken ceramic heads clinical uncertainties arise, whether the taper of the stem can be reused with a new ceramic head, especially if the stem is well fixed. Ceramic is a brittle material. Even small damage on the male stem taper can lead to stress concentrations causing premature failure of a new ceramic head. As a consequence, manufactures strictly prohibit stem taper re-use for ceramic heads. The aim of this study was to determine the fracture strength of ceramic heads assembled to re-used male stem tapers, which were subjected to prior head fracture. Five 12/14 Ti6Al4V male tapers and 15 Al2O3 ceramic heads (BIOLOX forte (R); circle divide 28 mm, L) were used for three consecutive fracture tests. Before and after every fracture test, all components were inspected visually and the surface geometry was analyzed. Mean fracture force (52.5 kN) did not decrease with the number of taper re-uses (p >= 0.77) but the range increased significantly from initially 4.1 kN to 31.8 kN for the first and 52.6 kN for the second re-use due to some components failing at very low loads. Visual inspection was not sufficient to predict the reduced failure loads. Ceramic heads should therefore not be put on used male tapers without metal adapter sleeves. 2015 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:545 / 552
页数:8
相关论文
共 35 条
  • [21] An assessment of in vivo failures of alumina ceramic total hip joint replacements
    Morrell, Roger
    Danzer, Robert
    Milosev, Ingrid
    Trebse, Rihard
    [J]. JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2012, 32 (12) : 3073 - 3084
  • [22] Pandorf T, 2010, MODERN TRENDS THA BE, P97
  • [23] Park J., 2009, Bioceramics: Properties, Characterizations, and Applications
  • [24] Fracture of a ceramic femoral head after a revision operation - A case report
    Pulliam, IT
    Trousdale, RT
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (01) : 118 - 121
  • [25] Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses
    Rehmer, Annelie
    Bishop, Nicholas E.
    Morlock, Michael M.
    [J]. CLINICAL BIOMECHANICS, 2012, 27 (01) : 77 - 83
  • [26] Tateiwa T, 2008, AM J ORTHOP, V37, P26
  • [27] Fracture of Ceramic Bearing Surfaces following Total Hip Replacement: A Systematic Review
    Traina, Francesco
    De Fine, Marcello
    Di Martino, Alberto
    Faldini, Cesare
    [J]. BIOMED RESEARCH INTERNATIONAL, 2013, 2013
  • [28] Are asymmetric metal markings on the cone surface of ceramic femoral heads an indication of entrapped debris?
    Valet, Sebastian
    Weisse, Bernhard
    Kuebler, Jakob
    Zimmermann, Martin
    Affolter, Christian
    Terrasi, Giovanni Pietro
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2014, 13
  • [29] Influence of contaminants in the stem-ball interface on the static fracture load of ceramic hip joint ball heads
    Weisse, B.
    Affolter, C.
    Stutz, A.
    Terrasi, G. P.
    Koebel, S.
    Weber, W.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) : 829 - 835
  • [30] Failure analysis of in vivo fractured ceramic femoral heads
    Weisse, B.
    Affolter, Ch.
    Terrasi, G. P.
    Piskoty, G.
    Koebel, S.
    [J]. ENGINEERING FAILURE ANALYSIS, 2009, 16 (04) : 1188 - 1194