Motion Artifact Removal using Cascade Adaptive Filtering for Ambulatory ECG Monitoring System

被引:0
作者
Kim, Hyejung [1 ]
Kim, Sunyoung [1 ]
Van Helleputte, Nick [1 ]
Berset, Torfinn [2 ]
Geng, Di [2 ]
Romero, Inaki [2 ]
Penders, Julien [2 ]
Van Hoof, Chris [1 ,2 ]
Yazicioglu, Refet Firat [1 ]
机构
[1] IMEC, Kapeldreef 75, B-3000 Louvain, Belgium
[2] Imec nl Holst Ctr, Eindhoven, Netherlands
来源
2012 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS): INTELLIGENT BIOMEDICAL ELECTRONICS AND SYSTEM FOR BETTER LIFE AND BETTER ENVIRONMENT | 2012年
关键词
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A motion artifact removal method with a two-stage cascade LMS adaptive filter is proposed for an ambulatory ECG monitoring system. The first LMS stage consisting of analog feedback prevents the signal saturation to reduce the input dynamic range. An adaptive step-size LMS algorithm is introduced and employed for the second LMS stage. The adaptive step-size algorithm can achieve fast convergence to track large sudden motion artifact quickly, while preventing the distortion of the ECG component. The filtering performance is evaluated by the heart beat detection, measured by sensitivity (Se) and positive predictive value (+p), and the performance is increased by 9.8% and 6.48%, respectively, compared to the unfiltered signal at the worst case with -25dB SNR. The proposed motion artifact method is implemented on an ambulatory ECG monitoring module, and the real-time measurement shows a significant performance improvement.
引用
收藏
页码:160 / 163
页数:4
相关论文
共 7 条
  • [1] [Anonymous], 2012, Digest of ISSCC
  • [2] A comparative analysis of principal component and independent component techniques for electrocardiograms
    Chawla, M. P. S.
    [J]. NEURAL COMPUTING & APPLICATIONS, 2009, 18 (06) : 539 - 556
  • [3] A Family of Normalized LMS Algorithms
    Douglas, Scott C.
    [J]. IEEE SIGNAL PROCESSING LETTERS, 1994, 1 (03) : 49 - 51
  • [4] Comparison of methods for adaptive removal of motion artifact
    Hamilton, PS
    Curley, MG
    Aimi, RM
    Sae-Hau, C
    [J]. COMPUTERS IN CARDIOLOGY 2000, VOL 27, 2000, 27 : 383 - 386
  • [5] Kim S., 2010, IEEE S VLSI CIRC VLS, P219
  • [6] Motion Artifact Removal for Functional Near Infrared Spectroscopy: A Comparison of Methods
    Robertson, F. C.
    Douglas, T. S.
    Meintjes, E. M.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (06) : 1377 - 1387
  • [7] Romero I., 2011, P C WIR HLTH