Glaucoma diagnostic value of the total macular thickness and ganglion cell-inner plexiform layer thickness according to optic disc area

被引:6
作者
Yoon, Myung Hun [1 ,2 ]
Park, Sang Jun [1 ,2 ]
Kim, Chan Yun [3 ]
Chin, Hee Seung [1 ,2 ]
Kim, Na Rae [1 ,2 ]
机构
[1] Inha Univ, Sch Med, Dept Ophthalmol, Inchon, South Korea
[2] Inha Univ, Inha Vis Sci Lab, Inchon, South Korea
[3] Yonsei Univ, Coll Med, Dept Ophthalmol, Inst Vis Res, Seoul, South Korea
关键词
NERVE-FIBER LAYER; HEIDELBERG RETINA TOMOGRAPH; FOURIER-DOMAIN OCT; COHERENCE TOMOGRAPHY; DETECT GLAUCOMA; AXIAL LENGTH; ACCURACY; COMPLEX; MYOPIA; PROFILE;
D O I
10.1136/bjophthalmol-2013-303185
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims This study examined the diagnostic value of the perimacular ganglion cell-inner plexiform layer (GCIPL) thickness in the detection of glaucoma according to the optic disc area. Methods The participants underwent reliable standard automated perimetry tests and Cirrus optical coherence tomography. The area under the receiver operating characteristic curve (AUC) was used to determine the diagnostic ability of the GCIPL, total macular thickness, retinal nerve fibre layer (RNFL) and optic nerve head parameters. Subsequently, optic disc areas of <= 1.9, <= 2.4 and >2.4 mm(2) were chosen arbitrarily, and the AUCs and sensitivities at fixed specificities were compared for each. Results 87 normal subjects and 96 glaucoma patients were enrolled. The mean GCIPL thickness and RNFL showed similar ability to diagnose glaucoma (p=0.911). The mean GCIPL thickness had higher diagnostic ability than the central retinal subfield thickness (p=0.001), comparable with that of the retinal cube volume (p=0.840) and mean retinal cube thickness (p=0.840). The mean RNFL (0.866), the minimum GCIPL (0.840) and the inferior and inferotemporal (0.853) disc had the highest AUCs for small, medium and large discs, respectively. Conclusions The macular GCIPL thickness, total macular cube thickness, circumpapillary RNFL thickness and disc rim area showed similar performance in diagnosing glaucoma, irrespective of the optic disc size.
引用
收藏
页码:315 / 321
页数:7
相关论文
共 25 条
[1]   Influence of optic disc size on the diagnostic performance of macular ganglion cell complex and peripapillary retinal nerve fiber layer analyses in glaucoma [J].
Cordeiro, Daniela Valenca ;
Lima, Veronica Castro ;
Castro, Dinorah P. ;
Castro, Leonardo C. ;
Pacheco, Maria Angelica ;
Lee, Jae Min ;
Dimantas, Marcelo I. ;
Prata, Tiago Santos .
CLINICAL OPHTHALMOLOGY, 2011, 5 :1333-1337
[2]   Macular thickness decreases with age in normal eyes: a study on the macular thickness map protocol in the Stratus OCT [J].
Eriksson, U. ;
Alm, A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (11) :1448-1452
[3]   Comparison of data analysis tools for detection of glaucoma with the Heidelberg Retina Tomograph [J].
Ford, BA ;
Artes, PH ;
McCormick, TA ;
Nicolela, MT ;
LeBlanc, RP ;
Chauhan, BC .
OPHTHALMOLOGY, 2003, 110 (06) :1145-1150
[4]   Diagnostic accuracy of nerve fibre layer, macular thickness and optic disc measurements made with the RTVue-100 optical coherence tomograph to detect glaucoma [J].
Garas, A. ;
Vargha, P. ;
Hollo, G. .
EYE, 2011, 25 (01) :57-65
[5]   Macular segmentation with optical coherence tomography [J].
Ishikawa, H ;
Stein, DM ;
Wollstein, G ;
Beaton, S ;
Fujimoto, JG ;
Schuman, JS .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (06) :2012-2017
[6]   Comparing the ganglion cell complex and retinal nerve fibre layer measurements by Fourier domain OCT to detect glaucoma in high myopia [J].
Kim, Na Rae ;
Lee, Eun Suk ;
Seong, Gong Je ;
Kang, Sung Yong ;
Kim, Ji Hyun ;
Hong, Samin ;
Kim, Chan Yun .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2011, 95 (08) :1115-1121
[7]   Determinants of Perimacular Inner Retinal Layer Thickness in Normal Eyes Measured by Fourier-Domain Optical Coherence Tomography [J].
Kim, Na Rae ;
Kim, Ji Hyun ;
Lee, Junwon ;
Lee, Eun Suk ;
Seong, Gong Je ;
Kim, Chan Yun .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (06) :3413-3418
[8]   Structure-Function Relationship and Diagnostic Value of Macular Ganglion Cell Complex Measurement Using Fourier-Domain OCT in Glaucoma [J].
Kim, Na Rae ;
Lee, Eun Suk ;
Seong, Gong Je ;
Kim, Ji Hyun ;
An, Hyong Gin ;
Kim, Chan Yun .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (09) :4646-4651
[9]   Three-Dimensional Imaging of Macular Inner Structures in Glaucoma by Using Spectral-Domain Optical Coherence Tomography [J].
Kotera, Yuriko ;
Hangai, Masanori ;
Hirose, Fumitaka ;
Mori, Satoshi ;
Yoshimura, Nagahisa .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (03) :1412-1421
[10]   Regional variations in the relationship between macular thickness measurements and myopia [J].
Lam, Dennis Shun Chiu ;
Leung, King Sai ;
Mohamed, Shaheeda ;
Chan, Wai-man ;
Palanivelu, Mahesh Shanmugam ;
Cheung, Carol Yim Lui ;
Li, Emmy Yuen Mei ;
Lai, Ricky Yiu Kwong ;
Leung, Christopher Kai-shun .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (01) :376-382