Miniaturized and automated sample pretreatment for determination of PCBs in environmental aqueous samples using an on-line microporous membrane liquid-liquid extraction-gas chromatography system

被引:42
作者
Barri, T
Bergström, S
Norberg, J
Jönsson, JA
机构
[1] Lund Univ, Dept Analyt Chem, SE-22100 Lund, Sweden
[2] Biotage AB, SE-75318 Uppsala, Sweden
关键词
D O I
10.1021/ac035089g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new, fast, and automated sample pretreatment technique for determination of lipophilic organic compounds in aqueous samples has been developed and applied to the determination of polychlorinated biphenyls (PCBs) in environmental river water. It is based on miniaturized microporous membrane liquid-liquid extraction coupled on-line to gas chromatography (GC) with electron capture detection. The heart of the system that simultaneously connects the sample pretreatment step to the final GC analysis has been named the extracting syringe (ESy). The ESy carries a miniaturized membrane extraction card attached to an electrically and mechanically designed installment and is mounted directly over a GC injector for fully automated injection of the extract. A method was developed to extract 10 PCB congeners from 1-mL water samples (after addition of 40% acetonitrile) with an extraction time of 10 min. The optimized methodology showed good linearity (in the dynamic concentration range of 5 ng L-1-1 mug L-1), enrichment factors of 33-40 times, repeatable extractions (RSD 2-5%, n = 4), and low detection limits (2-3 ng L-1). Acetonitrile had to be added to the samples in order to overcome the influence of PCB adsorption on the repeatability of extraction and enrichment and to minimize the overall memory effect (OME). OME and carryover depended not only on the concentration of the organic solvent added to the sample and that used in the washing procedure but also on whether the extracting card was changed or not. When an optimized washing procedure was applied, the OME was similar to0.2% at high concentrations (i.e., 1 mug L-1). When each extraction took place in a new extraction card, no OME was detected. Additionally, no significant adsorption onto glass surfaces or a matrix effect on extraction was noticed. The main features of this methodology are good extraction repeatability, low detection limits at short extraction time, and the unsurpassed characteristic of no detectable OME in the entire system when each sample is processed in a new card. The total consumption of organic (nonchlorinated) solvents is less than 5 mL per sample.
引用
收藏
页码:1928 / 1934
页数:7
相关论文
共 42 条
  • [11] MultiSimplex optimisation of the solid-phase microextraction-gas chromatographic-mass spectrometric determination of polycyclic aromatic hydrocarbons, polychlorinated biphenyls and phthalates from water samples
    Cortazar, E
    Zuloaga, O
    Sanz, J
    Raposo, JC
    Etxebarria, N
    Fernández, LA
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 978 (1-2) : 165 - 175
  • [12] Large volume preconcentration of dissolved hydrocarbons and polychlorinated biphenyls from seawater. Intercomparison between C-18 disks and XAD-2 column
    Dachs, J
    Bayona, JM
    [J]. CHEMOSPHERE, 1997, 35 (08) : 1669 - 1679
  • [13] Partitioning of antifouling agents, Irgarol 1051 and Sea Nine 211, to humic organic matter investigated by solid-phase microextraction
    Lambropoulou, DA
    Sakkas, VA
    Albanis, TA
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2004, 84 (1-3) : 47 - 54
  • [14] Erickson M.D., 1997, ANAL CHEM PCBS
  • [15] Evaluation of liquid-solid extraction with a new sorbent and liquid-liquid extraction for multiresidue pesticides. Determination in raw and finished drinking waters
    Fernandez, MJ
    Garcia, C
    GarciaVillanova, RJ
    Gomez, JA
    [J]. JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1996, 44 (07) : 1790 - 1795
  • [16] Current developments in the analysis of water pollution by polychlorinated biphenyls
    Font, G
    Manes, J
    Molto, JC
    Pico, Y
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1996, 733 (1-2) : 449 - 471
  • [17] Determination of pesticide residues in red wines with microporous membrane liquid-liquid extraction and gas chromatography
    Hyötyläinen, T
    Tuutijärvi, T
    Kuosmanen, K
    Riekkola, ML
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 372 (5-6) : 732 - 736
  • [18] Jonsson JÅ, 2001, ADV CHROMATOGR, V41, P53
  • [19] Liquid membrane extraction in analytical sample preparation I.: Principles
    Jönsson, JÅ
    Mathiasson, L
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (05) : 318 - 325
  • [20] Liquid membrane extraction in analytical sample preparation II.: Applications
    Jönsson, JÅ
    Mathiasson, L
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (05) : 325 - 334