Flow injection determination of mercury with preconcentration by amalgamation on a gold-platinum gauze by inductively coupled plasma mass spectrometry

被引:43
作者
Debrah, E [1 ]
Denoyer, ER [1 ]
机构
[1] UNIV MASSACHUSETTS,DEPT CHEM,AMHERST,MA 01003
关键词
preconcentration; amalgamation; flow injection; inductively coupled plasma mass spectrometry;
D O I
10.1039/ja9961100127
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Trace concentrations of mercury in water samples mere determined by a method involving a preconcentration procedure and inductively coupled plasma mass spectrometry (ICP-MS) detection. Mercury vapour, generated by tin(II) chloride as the reductant, was trapped by amalgamation on a gold-platinum gauze, released by controlled heating and detected by ICP-MS. A how injection sample introduction system with time-based injection was used and the sensitivity was found to be proportional to the mass of mercury introduced. Quantitative recoveries of mercury from open ocean sea-water, coastal sea-water and fresh water reference materials were obtained. A detection limit of 200 pg l(-1) for a 25 ml sample was obtained. The precision of the method was 1% (relative standard deviation) at 1 mu g l(-1).
引用
收藏
页码:127 / 132
页数:6
相关论文
共 20 条
[1]   DETERMINATION OF TRACE-METALS IN MARINE BIOLOGICAL REFERENCE MATERIALS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
WILLIE, SN ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1988, 60 (07) :687-691
[2]   DETERMINATION OF URANIUM IN ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BOOMER, DW ;
POWELL, MJ .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2810-2813
[3]   COLD VAPOR ATOMIC-ABSORPTION DETERMINATION OF MERCURY BY FLOW-INJECTION ANALYSIS USING A TEFLON MEMBRANE PHASE SEPARATOR COUPLED TO THE ABSORPTION CELL [J].
DEANDRADE, JC ;
PASQUINI, C ;
BACCAN, N ;
VANLOON, JC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1983, 38 (10) :1329-1338
[4]   AN IMPROVED INTERFACE FOR INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY (ICP-MS) [J].
DOUGLAS, DJ ;
FRENCH, JB .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (03) :197-204
[5]  
ERLER W, TSAA9A
[6]  
FANG ZL, 1988, ANAL CHIM ACTA, V214, P41
[7]   COMBINATION OF FLOW-INJECTION TECHNIQUES WITH ATOMIC SPECTROMETRY IN AGRICULTURAL AND ENVIRONMENTAL-ANALYSIS [J].
FANG, ZL ;
XU, SK ;
WANG, X ;
ZHANG, SC .
ANALYTICA CHIMICA ACTA, 1986, 179 :325-340
[8]   DETERMINATION OF MERCURY IN URINE BY FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
GUO, TZ ;
BAASNER, J .
ANALYTICA CHIMICA ACTA, 1993, 278 (01) :189-196
[9]   ONLINE MICROWAVE SAMPLE PRETREATMENT FOR THE DETERMINATION OF MERCURY IN BLOOD BY FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY [J].
GUO, TZ ;
BAASNER, J .
TALANTA, 1993, 40 (12) :1927-1936
[10]   EXAMINATION OF SEPARATION EFFICIENCIES OF MERCURY-VAPOR FOR DIFFERENT GAS-LIQUID SEPARATORS IN FLOW-INJECTION COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY WITH AMALGAM PRECONCENTRATION [J].
HANNA, CP ;
HAIGH, PE ;
TYSON, JF ;
MCINTOSH, S .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (04) :585-588