Magnetic susceptibility as a B0 field strength independent MRI biomarker of liver iron overload

被引:36
|
作者
Hernando, Diego [1 ]
Cook, Rachel J. [2 ]
Diamond, Carol [3 ]
Reeder, Scott B. [1 ,2 ,4 ,5 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Med, Madison, WI 53792 USA
[3] Univ Wisconsin, Dept Pediat, Madison, WI 53792 USA
[4] Univ Wisconsin, Dept Med Phys, Madison, WI 53792 USA
[5] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53792 USA
关键词
susceptibility; iron overload; chemical shift encoded imaging; field map; magnetic resonance imaging; liver; RESONANCE-IMAGING MEASUREMENT; WATER-FAT SEPARATION; SICKLE-CELL-DISEASE; NONINVASIVE MEASUREMENT; QUANTIFICATION; BIOPSY; VALIDATION; HEMOCHROMATOSIS; RECONSTRUCTION; THALASSEMIA;
D O I
10.1002/mrm.24848
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: MR-based quantification of liver magnetic susceptibility may enable field strength-independent measurement of liver iron concentration (LIC). However, susceptibility quantification is challenging, due to nonlocal effects of susceptibility on the B-0 field. The purpose of this work is to demonstrate feasibility of susceptibility-based LIC quantification using a fat-referenced approach. Methods: Phantoms consisting of vials with increasing iron concentrations immersed between oil/water layers, and 27 subjects (9 controls/18 subjects with liver iron overload) were scanned. Ferriscan (1.5 T) provided R2-based reference LIC. Multiecho three-dimensional-SPGR (1.5 T/3 T) enabled fat-water, B-0- and R2*-mapping. Phantom iron concentration (mg Fe L-1) was estimated from B-0 differences (B-0) between vials and neighboring oil. Liver susceptibility and LIC (mg Fe g(-1) dry tissue) was estimated from B-0 between the lateral right lobe of the liver and adjacent subcutaneous adipose tissue. Results: Estimated phantom iron concentrations had good correlation with true iron concentrations (1.5 T:slope = 0.86, intercept = 0.72, r(2) = 0.98; 3 T:slope = 0.85, intercept = 1.73, r(2) = 0.98). In liver, B-0 correlated strongly with R2* (1.5 T:r(2) = 0.86; 3 T:r(2) = 0.93) and B-0-LIC had good agreement with Ferriscan-LIC (slopes/intercepts nearly 1.0/0.0, 1.5 T:r(2) = 0.67, slope = 0.93 +/- 0.13, P approximate to 0.50, intercept = 1.93 +/- 0.78, P approximate to 0.02; 3 T:r(2) = 0.84, slope = 1.01 +/- 0.09, P approximate to 0.90, intercept = 0.23 +/- 0.52, P approximate to 0.68). Discussion: Fat-referenced, susceptibility-based LIC estimation is feasible at both field strengths. This approach may enable improved susceptibility mapping in the abdomen. Magn Reson Med 70:648-656, 2013. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:648 / 656
页数:9
相关论文
共 22 条
  • [1] B0 magnetic field homogeneity and shimming for in vivo magnetic resonance spectroscopy
    Juchem, Christoph
    de Graaf, Robin A.
    ANALYTICAL BIOCHEMISTRY, 2017, 529 : 17 - 29
  • [2] Effect of head motion on MRI B0 field distribution
    Liu, Jiaen
    de Zwart, Jacco A.
    van Gelderen, Peter
    Murphy-Boesch, Joseph
    Duyn, Jeff H.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (06) : 2538 - 2548
  • [3] Patient-to-patient variation of susceptibility-induced B0 field in bilateral breast MRI
    Lee, Seung-Kyun
    Hancu, Ileana
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (04) : 873 - 880
  • [4] Metallic Implant Geometry and Susceptibility Estimation Using Multispectral B0 Field Maps
    Shi, Xinwei
    Yoon, Daehyun
    Koch, Kevin M.
    Hargreaves, Brian A.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (06) : 2402 - 2413
  • [5] B0 field homogeneity recommendations, specifications, and measurement units for MRI in radiation therapy
    Gach, H. Michael
    Curcuru, Austen N.
    Mutic, Sasa
    Kim, Taeho
    MEDICAL PHYSICS, 2020, 47 (09) : 4101 - 4114
  • [6] Measurement of the liver iron concentration in transfusional iron overload by MRI R2*and by high-transition-temperature superconducting magnetic susceptometry
    Sheth, Sujit
    Allen, Christopher J.
    Farrell, David E.
    Tripp, John H.
    Jafari, Ramin
    Wang, Yi
    Brittenham, Gary M.
    CLINICAL IMAGING, 2019, 55 : 65 - 70
  • [7] B0 concomitant field compensation for MRI systems employing asymmetric transverse gradient coils
    Weavers, Paul T.
    Tao, Shengzhen
    Trzasko, Joshua D.
    Frigo, Louis M.
    Shu, Yunhong
    Frick, Matthew A.
    Lee, Seung-Kyun
    Foo, Thomas K-F
    Bernstein, Matt A.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (03) : 1538 - 1544
  • [8] 3D gradient system for two B0 field directions in earth’s field MRI
    Steffen Lother
    Uvo Hoelscher
    Thomas Kampf
    Peter Jakob
    Florian Fidler
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2013, 26 : 565 - 573
  • [9] Measurement of liver iron overload: Noninvasive calibration of MRI-R2*by magnetic iron detector susceptometer
    Gianesin, B.
    Zefiro, D.
    Musso, M.
    Rosa, A.
    Bruzzone, C.
    Balocco, M.
    Carrara, P.
    Bacigalupo, L.
    Banderali, S.
    Rollandi, G. A.
    Gambaro, M.
    Marinelli, M.
    Forni, G. L.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (06) : 1782 - 1786
  • [10] Non-Cartesian sampled centric scan SPRITE imaging with magnetic field gradient and B0(t) field measurements for MRI in the vicinity of metal structures
    Han, Hui
    Green, Derrick
    Ouellette, Matthew
    MacGregor, Rodney
    Balcom, Bruce J.
    JOURNAL OF MAGNETIC RESONANCE, 2010, 206 (01) : 97 - 104