A Cadaver Study to Compare Vertebral Augmentation With a High-viscosity Cement to Augmentation With Conventional Lower-viscosity Cement

被引:30
作者
Lador, Ran [1 ]
Liberman, Shari [1 ]
Ben-Galim, Peleg [1 ]
Dreiangel, Niv [1 ]
Reitman, Charles A. [1 ]
Hipp, John A. [1 ]
机构
[1] Baylor Coll Med, Dept Orthoped Surg, Spine Res Lab, Houston, TX 77030 USA
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2013年 / 26卷 / 02期
关键词
vertebral augmentation; cement viscosity; extravsation; cadaver study; high-resolution imaging; PERCUTANEOUS TRANSPEDICULAR VERTEBROPLASTY; OSTEOPOROTIC COMPRESSION FRACTURES; BONE-CEMENT; PULMONARY-EMBOLISM; RANDOMIZED-TRIAL; POLYMETHYLMETHACRYLATE; LEAKAGE; VOLUME; BIOMECHANICS; KYPHOPLASTY;
D O I
10.1097/BSD.0b013e3182337d69
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: Comparison of extravasations in fractured cadaver vertebrae augmented with commercial low-viscosity versus high-viscosity cements. Objective: Use of high-resolution, 3-dimensional (3D) imaging to test the hypothesis that high-viscosity cements can reduce the type and severity of extravasations after vertebral augmentation procedures. Summary of Background Data: Cement extravasations are one of the primary complications of vertebral augmentation procedures. There is some evidence that high-viscosity cements might reduce extravasations, but additional data are needed to confirm the early findings. Methods: A range of vertebral fractures were created in fresh human cadavers. One group was then augmented with a low-viscosity polymethylmethacrylate (PMMA)-based cement and the other group injected with high-viscosity PMMA-based cement. High-resolution computerized tomography exams were obtained, and extravasations were assessed using 3D volume renderings. The type and severity of extravasations were recorded and analyzed Results: The proportion of vertebrae with any type of extravasation through the posterior wall to the spinal canal, into small vessels laterally or anteriorly, through the endplates, or anywhere around the body was not significantly different between the high-viscosity and low-viscosity groups. There was significantly less severe extravasation through the endplates (P = 0.02), and a trend toward less severe extravasation through vessels (P = 0.06) with the high versus low-viscosity cements. Conclusions: In agreement with previous research, high-viscosity PMMA-based cement may help to reduce the more severe forms of extravasations after vertebral augmentation procedures in newly fractured vertebrae.
引用
收藏
页码:68 / 73
页数:6
相关论文
共 35 条
[1]   Percutaneous vertebroplasty and bone cement leakage: Clinical experience with a new high-viscosity bone cement and delivery system for vertebral augmentation in benign and malignant compression fractures [J].
Anselmetti, Giovanni Carlo ;
Zoarski, Gregg ;
Manca, Antonio ;
Masala, Salvatore ;
Eminefendic, Haris ;
Russo, Filippo ;
Regge, Daniele .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 31 (05) :937-947
[2]   High-viscosity cement significantly enhances uniformity of cement filling in vertebroplasty: An experimental model and study on cement leakage [J].
Baroud, Gamal ;
Crookshank, Meghan ;
Bohner, Marc .
SPINE, 2006, 31 (22) :2562-2568
[3]   The biomechanics of vertebroplasty - The effect of cement volume on mechanical behavior [J].
Belkoff, SM ;
Mathis, JM ;
Jasper, LE ;
Deramond, H .
SPINE, 2001, 26 (14) :1537-1541
[4]   A Randomized Trial of Vertebroplasty for Painful Osteoporotic Vertebral Fractures [J].
Buchbinder, Rachelle ;
Osborne, Richard H. ;
Ebeling, Peter R. ;
Wark, John D. ;
Mitchell, Peter ;
Wriedt, Chris ;
Graves, Stephen ;
Staples, Margaret P. ;
Murphy, Bridie .
NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (06) :557-568
[5]  
Burton AW., 2005, Neurosurg Focus, V18, P1, DOI DOI 10.3171/FOC.2005.18.3.2
[6]   A lethal pulmonary embolism during percutaneous vertebroplasty [J].
Chen, HL ;
Wong, CS ;
Ho, ST ;
Chang, FL ;
Hsu, CH ;
Wu, CT .
ANESTHESIA AND ANALGESIA, 2002, 95 (04) :1060-1062
[7]   Temperature elevation caused by bone cement polymerization during vertebroplasty [J].
Deramond, H ;
Wright, NT ;
Belkoff, SM .
BONE, 1999, 25 (02) :17S-21S
[8]   Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients [J].
Fourney, DR ;
Schomer, DF ;
Nader, R ;
Chlan-Fourney, J ;
Suki, D ;
Ahrar, K ;
Rhines, LD ;
Gokaslan, ZL .
JOURNAL OF NEUROSURGERY, 2003, 98 (01) :21-30
[9]   Pulmonary embolism caused by polymethylmethacrylate during percutaneous vertebroplasty in orthopaedic surgery [J].
Freitag, M ;
Gottschalk, A ;
Schuster, M ;
Wenk, W ;
Wiesner, L ;
Standl, TG .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2006, 50 (02) :248-251
[10]  
GALIBERT P, 1987, NEUROCHIRURGIE, V33, P166