Diagnostic accuracy of LAMP assay for HBV infection

被引:13
作者
Chen, Chu-Mao [1 ]
Ouyang, Shi [2 ]
Lin, Li-Ying [1 ,3 ]
Wu, Li-Juan [4 ]
Xie, Tian-Ao [1 ,3 ]
Chen, Juan-Jiang [1 ]
Li, Zhen-Xing [5 ]
Zhu, Guo-Dong [6 ]
Ji, Tian-Xing [7 ]
Pan, Zhi-Yong [1 ,3 ]
Xia, Yong [1 ,3 ]
Guo, Xu-Guang [1 ,3 ,8 ,9 ]
机构
[1] Guangzhou Med Univ, Affiliated Hosp 3, Dept Clin Lab Med, Guangzhou, Guangdong, Peoples R China
[2] Guangzhou Med Univ, Affiliated Hosp 5, Dept Infect Dis, Guangzhou, Guangdong, Peoples R China
[3] Guangzhou Med Univ, Clin Sch 3, Dept Clin Med, Guangzhou, Guangdong, Peoples R China
[4] Jinan Univ, Baoan Maternal & Child Hlth Hosp, Shenzhen, Guangdong, Peoples R China
[5] Guangzhou Med Univ, Affiliated Hosp 3, Dept Resp Med, Guangzhou, Guangdong, Peoples R China
[6] South China Univ Technol, Dept Geriatr, Guangzhou Peoples Hosp 1, Sch Med, Guangzhou, Guangdong, Peoples R China
[7] Guangzhou Med Univ, Affiliated Hosp 2, Dept Clin Med, Guangzhou, Guangdong, Peoples R China
[8] Guangzhou Med Univ, Affiliated Hosp 3, Key Lab Major Obstet Dis Guangdong Prov, Guangzhou, Guangdong, Peoples R China
[9] Guangzhou Med Univ, Affiliated Hosp 3, Key Lab Reprod & Genet, Guangdong Higher Educ Inst, Guangzhou, Guangdong, Peoples R China
关键词
HBV; LAMP; PCR; rapid diagnosis accuracy; systematic evaluation; HEPATITIS-B-VIRUS; MEDIATED ISOTHERMAL AMPLIFICATION; SYSTEMATIC REVIEWS; RAPID DETECTION; BLOOD; QUANTIFICATION; HETEROGENEITY; METAANALYSIS; GENOTYPES; ELISA;
D O I
10.1002/jcla.23281
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Detection of hepatitis B virus (HBV) is vital for the diagnosis of hepatitis B infection. A novel test loop-mediated isothermal amplification (LAMP) has been successfully applied to detect various pathogens. However, the accuracy of LAMP in diagnosing HBV remains unclear. Therefore, in the present study, the accuracy of LAMP for HBV detection was evaluated systematically. Methods Embase, Cochrane Library, and PubMed databases were searched for studies using LAMP to detect HBV. Then, two researchers extracted data and assessed the quality of literature using the QUADAS-2 tool independently. I(2)statistic and chi-square test were analyzed to investigate the heterogeneity, and Deek's funnel plot assessed the publication bias. The pooled sensitivity (SEN), specificity (SPE), positive LR (PLR), negative LR (NLR), diagnostic odds ratio (DOR), and 95% confidence intervals were displayed in forest plots. We calculated the area under the curve (AUC) to assess the overall efficiency of LAMP for HBV detection. Results A total of nine studies with 1298 samples were finally included in this evaluation. The pooled sensitivity and specificity of HBV detection were 0.91 (95% CI: 0.89 ~ 0.92) and 0.97 (95% CI: 0.94 ~ 0.99), respectively. The PLR, NLR, and DOR were 16.93 (95% CI: 6.15 ~ 46.55), 0.08 (95% CI: 0.05 ~ 0.14), and 397.57 (95% CI: 145.41 ~ 1087.07). Besides, the AUC was 0.9872, and Deek's plot suggested that there existed publication bias in the studies. Conclusion Compared with PCR, LAMP is a simple, rapid, and effective assay to diagnose HBV. However, additional evidence is essential to confirm that LAMP can replace other methods in diagnosing HBV infection.
引用
收藏
页数:9
相关论文
共 43 条
[21]   Systematic Reviews of Diagnostic Test Accuracy [J].
Leeflang, Mariska M. G. ;
Deeks, Jonathan J. ;
Gatsonis, Constantine ;
Bossuyt, Patrick M. M. .
ANNALS OF INTERNAL MEDICINE, 2008, 149 (12) :889-+
[22]  
Mansouri N, 2014, IRAN J PEDIATR HEMAT, V4, P178
[23]   Preferred Reporting Items for a Systematic Review and Meta-analysis of Diagnostic Test Accuracy Studies The PRISMA-DTA Statement [J].
McInnes, Matthew D. F. ;
Moher, David ;
Thombs, Brett D. ;
McGrath, Trevor A. ;
Bossuyt, Patrick M. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2018, 319 (04) :388-396
[24]   Rapid detection of the New Delhi metallo-b-lactamase 1 (NDM-1) gene by loop-mediated isothermal amplification (LAMP) [J].
Moreira, Mirna Giselle ;
Barreto, Lidia Miranda ;
dos Santos, Vera Lucia ;
Monteiro, Andrea Souza ;
Nobre, Vandack ;
dos Santos, Simone Goncalves .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (04)
[25]  
Moslemi E, 2009, AFR J MICROBIOL RES, V3, P439
[26]   Closed tube loop-mediated isothermal amplification assay for rapid detection of hepatitis B virus in human blood [J].
Nguyen Bao Quoc ;
Nguyen Doan Nguyen Phuong ;
Nguyen Ngoc Bao Chau ;
Do Thi Phuong Linh .
HELIYON, 2018, 4 (03)
[27]   Loop-mediated isothermal amplification of DNA [J].
Notomi, Tsugunori ;
Okayama, Hiroto ;
Masubuchi, Harumi ;
Yonekawa, Toshihiro ;
Watanabe, Keiko ;
Amino, Nobuyuki ;
Hase, Tetsu .
NUCLEIC ACIDS RESEARCH, 2000, 28 (12)
[28]   Rapid Detection of Hepatitis B Virus in Blood Plasma by a Specific and Sensitive Loop-Mediated Isothermal Amplification Assay [J].
Nyan, Dougbeh-Chris ;
Ulitzky, Laura E. ;
Cehan, Nicoleta ;
Williamson, Phillip ;
Winkelman, Valerie ;
Rios, Maria ;
Taylor, Deborah R. .
CLINICAL INFECTIOUS DISEASES, 2014, 59 (01) :16-23
[29]   The use of fresh whole blood in massive transfusion [J].
Repine, Thomas B. ;
Perkins, Jeremy G. ;
Kauvar, David S. ;
Blackborne, Lorne .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 60 (06) :S59-S66
[30]   Detection of Acute HIV-1 Infection by RT-LAMP [J].
Rudolph, Donna L. ;
Sullivan, Vickie ;
Owen, S. Michele ;
Curtis, Kelly A. .
PLOS ONE, 2015, 10 (05)