Adverse events related to extraoperative invasive EEG monitoring with subdural grid electrodes: A systematic review and meta-analysis

被引:170
作者
Arya, Ravindra [1 ]
Mangano, Francesco T. [2 ]
Horn, Paul S. [1 ,3 ]
Holland, Katherine D. [1 ]
Rose, Douglas F. [1 ]
Glauser, Tracy A. [1 ]
机构
[1] Cincinnati Childrens Hosp Med Ctr, Comprehens Epilepsy Ctr, Div Neurol, Cincinnati, OH USA
[2] Cincinnati Childrens Hosp Med Ctr, Div Neurosurg, Cincinnati, OH USA
[3] Univ Cincinnati, Dept Math Sci, Cincinnati, OH 45221 USA
关键词
Subdural grids; Intracranial EEG; Invasive EEG; Adverse events; Systematic review; EPILEPSY SURGERY PROCEDURES; SINGLE INSTITUTION; COMPLICATIONS; MORBIDITY; EXPERIENCE; CHILDREN; IMPLANTATION;
D O I
10.1111/epi.12073
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Implantation of subdural grids and invasive electroencephalography (EEG) monitoring is important to define the ictal-onset zone and eloquent cortex in selected patients with medically refractory epilepsy. The objective of this systematic review is to summarize data about adverse events related to this procedure. Methods English-language studies published up to July 2012, reporting such adverse events were reviewed. Outcome measures included demographic variables; surgical protocol including number of subdural electrodes implanted per patient, duration of monitoring, antibiotic, and steroid prophylaxis; and adverse events. Key Findings Twenty-one studies were identified including a total of 2,542 patients. The reported mean number of electrodes per patient and duration of monitoring varied from 52 to 95 and 5 to 17days, respectively. There is a trend toward more uniform use of antibiotics and steroids in the perioperative period. Neurologic infections (pooled prevalence2.3%, 95% confidence interval 1.53.1), superficial infections (3.0%, 1.94.1), intracranial hemorrhage (4.0%, 3.24.8), and elevated intracranial pressure (2.4%, 1.53.3) were found to be the most common adverse events. Up to 3.5% of patients required additional surgical procedure(s) for management of these adverse events. Increased number of electrodes (67) was found to be independently associated with increased incidence of adverse events. Significance Although providing critical information for patients with medically refractory epilepsy, subdural grids implantation and invasive EEG monitoring entails risks of infection, hemorrhage, and elevated intracranial pressure. The prevalence estimates, likely to be conservative due to selective reporting, are expected to be helpful in counseling patients.
引用
收藏
页码:828 / 839
页数:12
相关论文
共 25 条
  • [1] ADELSON PD, 1995, PEDIATR NEUROSURG, V22, P174
  • [2] Craniotomy-related complications in a paediatric neurosurgery unit - a prospective study
    Allibone, JB
    Harkness, WJF
    Hayward, RD
    [J]. BRITISH JOURNAL OF NEUROSURGERY, 1999, 13 (02) : 148 - 153
  • [3] [Anonymous], EPILEPSY COMPREHENSI
  • [4] Surgical and neurological complications in a series of 708 epilepsy surgery procedures
    Behrens, E
    Schramm, J
    Zentner, J
    Konig, R
    [J]. NEUROSURGERY, 1997, 41 (01) : 1 - 9
  • [5] Adverse Events Occurring During Invasive Electroencephalogram Recordings in Children
    Blauwblomme, Thomas
    Ternier, Jessica
    Romero, Carla
    Pier, Kelly S. T.
    D'Argenzio, Luigi
    Pressler, Ronit
    Cross, Helen
    Harkness, William
    [J]. NEUROSURGERY, 2011, 69 : 169 - 175
  • [6] Morbidity associated with the use of intracranial electrodes for epilepsy surgery
    Burneo, JG
    Steven, DA
    McLachlan, RS
    Parrent, AG
    [J]. CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 2006, 33 (02) : 223 - 227
  • [7] Subdural electrode-associated complications: A 20-year experience
    Fountas, Kostas N.
    Smith, Joseph R.
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (06) : 264 - 272
  • [8] Complications of invasive video-EEG monitoring with subdural grid electrodes
    Hamer, HM
    Morris, HH
    Mascha, EJ
    Karafa, MT
    Bingaman, WE
    Bej, MD
    Burgess, RC
    Dinner, DS
    Foldvary, NR
    Hahn, JF
    Kotagal, P
    Najm, I
    Wyllie, E
    Lüders, HO
    [J]. NEUROLOGY, 2002, 58 (01) : 97 - 103
  • [9] Complications of invasive subdural electrode monitoring at St. Louis Children's Hospital, 1994-2005
    Johnston, James M., Jr.
    Mangano, Francesco T.
    Ojemann, Jeffrey G.
    Park, Tae Sung
    Trevathan, Edwin
    Smyth, Matthew D.
    [J]. JOURNAL OF NEUROSURGERY, 2006, 105 (05) : 343 - 347
  • [10] Complications and results of subdural grid electrode implantation in epilepsy surgery
    Lee, WS
    Lee, JK
    Lee, SA
    Kang, JK
    Ko, TS
    [J]. SURGICAL NEUROLOGY, 2000, 54 (05): : 346 - 351