Neutron production from a mobile linear accelerator operating in electron mode for intraoperative radiation therapy

被引:30
作者
Loi, G
Dominietto, M
Cannillo, I
Ciocca, M
Krengli, M
Mones, E
Negri, E
Brambilla, M
机构
[1] Azienda Osped Maggiore Carita, Dept Med Phys, I-28100 Novara, Italy
[2] European Inst Oncol, Dept Med Phys, I-20141 Milan, Italy
[3] Azienda Osped Maggiore Carita, Dept Radiotherapy, I-28100 Novara, Italy
关键词
D O I
10.1088/0031-9155/51/3/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intraoperative electron beam radiotherapy is increasingly performed using mobile linac delivering therapeutic radiation doses ill unshielded operating rooms. While no special neutron-shielding problem Should arise for operation at 10 MeV or less, it is not clear whether this holds true for operation at higher energies. This paper reports the measured neutron production from a Mobetron mobile electron linac, operated at 12 MeV, and compares the results with those from,I conventional linac, also operated at 12 MeV in electron mode. Neutron leakage measurements were performed by means of passive bubble detectors in the scattering foil, patient and floor planes. Neutron dose equivalent rates per unit of electron close delivered by the Mobetron at its normal treatment distance (50 cm SSD) were 0.33 mu Sv Gy(-1) at the accelerator head, 0.18 mu Sv Gy(-1) ill the patient plane at 15 cm from the beam axis and 0.31 mu Sv Gy(-1) at the floor plane, on the beam axis and under the beam stopper. For a weekly workload of 250 Gy, the weekly neutron dose equivalents at 12 MeV for the Mobetron at a distance of 300 cm from the scattering foil were 14.3 and 1.7 mu Sv/week for floor below and adjoining areas on the same floor, respectively. Neutron dose equivalent rates generated from Mobetron are at least one order of magnitude lower than ones produced by a conventional linac operated at the same energy in electron mode. Mobetron call be used at 12 MeV in all unshielded operating room for,I weekly workload of up to 250 Gy if the bremsstrahlung x-rays are shielded to negligible levels.
引用
收藏
页码:695 / 702
页数:8
相关论文
共 8 条
[1]  
Daves J L, 2001, J Appl Clin Med Phys, V2, P165, DOI 10.1120/1.1385129
[2]   Characteristics of the photoneutron contamination present in a high-energy radiotherapy treatment room [J].
Garnica-Garza, HM .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (03) :531-539
[3]   PHOTONEUTRONS FROM HIGH-ENERGY MEDICAL LINEAR ACCELERATORS - MEASUREMENT OF THE SPECTRUM AND DOSE USING A MINIATURE PROPORTIONAL COUNTER [J].
KLIAUGA, P ;
AMOLS, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (03) :629-633
[4]   NEUTRON CONTAMINATION OF X-RAY-BEAMS PRODUCED BY THE VARIAN CLINAC-1800 [J].
MCGINLEY, PH ;
LANDRY, JC .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) :777-783
[5]  
Mills M D, 2001, J Appl Clin Med Phys, V2, P121, DOI 10.1120/1.1385128
[6]   SHIELDING CONSIDERATIONS FOR AN OPERATING-ROOM BASED INTRAOPERATIVE ELECTRON RADIOTHERAPY UNIT [J].
MILLS, MD ;
ALMOND, PR ;
BOYER, AL ;
OCHRAN, TG ;
MADIGAN, W ;
RICH, TA ;
DALLY, EB .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05) :1215-1221
[7]  
NCRP, 1984, 79 NCRP
[8]   Analysis of photoneutron spectra produced in medical accelerators [J].
Ongaro, C ;
Zanini, A ;
Nastasi, U ;
Ródenas, J ;
Ottaviano, G ;
Manfredotti, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) :L55-L61