Reduction of Blood Glucose Measurements to Calibrate Subcutaneous Glucose Sensors: A Bayesian Multiday Framework

被引:16
作者
Acciaroli, Giada [1 ]
Vettoretti, Martina [1 ]
Facchinetti, Andrea [1 ]
Sparacino, Giovanni [1 ]
Cobelli, Claudio [1 ]
机构
[1] Univ Padua, Dept Informat Engn, I-35131 Padua, Italy
关键词
Continuous glucose monitoring; deconvolution; modeling; parameter estimation; sensor accuracy; MONITORING-SYSTEM; PLASMA-GLUCOSE; TIME; ACCURACY; PERFORMANCE; ALGORITHM;
D O I
10.1109/TBME.2017.2706974
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: In most continuous glucose monitoring (CGM) devices used for diabetes management, the electrical signal measured by the sensor is transformed to glucose concentration by a calibration function whose parameters are estimated using self-monitoring of blood glucose (SMBG) samples. The calibration function is usually a linear model approximating the nonlinear relationship between electrical signal and glucose concentration in certain time intervals. Thus, CGM devices require frequent calibrations, usually twice a day. The aim here is to develop a new method able to reduce the frequency of calibrations. Methods: The algorithm is based on amultiple-day model of sensor time-variability with second-order statistical priors on its unknown parameters. In an online setting, these parameters are numerically determined by the Bayesian estimation exploiting SMBG sparsely collected by the patient. The method is assessed retrospectively on 108 CGM signals acquired for 7 days by the Dexcom G4 Platinum sensor, testing progressively less-calibration scenarios. Results: Despite the reduction of calibration frequency (on average from 2/day to 0.25/day), the method shows a statistically significant accuracy improvement compared to manufacturer calibration, e.g., mean absolute relative difference when compared to a laboratory reference decreases from 12.83% to 11.62% (p-value of 0.006). Conclusion: The methodology maintains (sometimes improves) CGM sensor accuracy compared to that of the original manufacturer, while reducing the frequency of calibrations. Significance: Reducing the need of calibrations facilitates the adoption of CGM technology both in terms of ease of use and cost, an obvious prerequisite for its use as replacement of traditional SMBG devices.
引用
收藏
页码:587 / 595
页数:9
相关论文
共 39 条
[1]   From Two to One Per Day Calibration of Dexcom G4 Platinum by a Time-Varying Day-Specific Bayesian Prior [J].
Acciaroli, Giada ;
Vettoretti, Martina ;
Facchinetti, Andrea ;
Sparacino, Giovanni ;
Cobelli, Claudio .
DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 (08) :472-479
[2]  
[Anonymous], 2012, DIABETES CARE, V35, pS11, DOI [10.2337/dc35-S011, 10.2337/dc12-s004]
[3]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[4]   New Features and Performance of a Next-Generation SEVEN-Day Continuous Glucose Monitoring System with Short Lag Time [J].
Bailey, Timothy ;
Zisser, Howard ;
Chang, Anna .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (12) :749-755
[5]  
Bailey Timothy S, 2015, J Diabetes Sci Technol, V9, P209, DOI 10.1177/1932296814559746
[6]   Can Calibration-Free Sensors Be Realized? [J].
Bakker, Eric .
ACS SENSORS, 2016, 1 (07) :838-841
[7]   Adaptive Calibration Algorithm for Plasma Glucose Estimation in Continuous Glucose Monitoring [J].
Barcelo-Rico, Fatima ;
Diez, Jose-Luis ;
Rossetti, Paolo ;
Vehi, Josep ;
Bondia, Jorge .
IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2013, 17 (03) :530-538
[8]   A Multiple Local Models Approach to Accuracy Improvement in Continuous Glucose Monitoring [J].
Barcelo-Rico, Fatima ;
Bondia, Jorge ;
Luis Diez, Jose ;
Rossetti, Paolo .
DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (01) :74-82
[9]  
Basu A., 2014, J DIABETES SCI TECHN, V9, P63, DOI DOI 10.1177/1932296814554797
[10]  
Bequette B Wayne, 2010, J Diabetes Sci Technol, V4, P404