In vitro evaluation of ureteral stent compression

被引:37
作者
Hendlin, K
Vedula, K
Horn, C
Monga, M
机构
[1] Univ Minnesota, Dept Urol, Minneapolis, MN 55455 USA
[2] Univ Minnesota, Dept Biomed Engn, Minneapolis, MN USA
[3] Vet Affairs Healthcare Syst, Dept Urol Surg, Minneapolis, MN USA
关键词
D O I
10.1016/j.urology.2005.10.038
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objectives. To evaluate the force required to radially compress ureteral stents and the impact of compression on stent performance across a broad range of stent sizes and manufacturers. Methods. Fourteen ureteral stents ranging in size from 6F to 10.3F were evaluated for radial compression using an MTS Micro Bionix Testing System using Testworks II software, 5N load cell, vibration isolation table, and socket set screws with rounded caps. Radial compression was exerted in 0.2-mm increments. Cycles of compression and a 100-second hold time were repeated for each stent to maximal compression. Three trials were completed for each stent. The Young's Modulus, E, was calculated from each trial using the engineering stress. Simultaneous flow data using normal saline were collected for the Amplatz stent. Results. The Cook C-Flex (E = 30,355 +/- 910 Pa), Microvasive Percuflex Plus 6F (E = 28,973 +/- 250 Pa), Cook EncloSof (E = 28,403 +/- 471 Pa), and Fossa Open Lumen (E = 28,627 +/- 5,338) were the most resistant to compression and the Cook Amplatz 10.2F (E = 11,922 +/- 220 Pa) and Microvasive Percuflex Plus 10.3F (E = 8,247 +/- 286 Pa) were the least resistant to compression. Stress relaxation occurred similarly in all stents. Flow through the Cook Amplatz stent decreased at a rate of 5 mL/s/mm of compression linearly (R-2 = 0.95). Conclusions. Stents resistant to radial compression are critical for function in the face of extrinsic ureteral obstruction. The Cook C-Flex ureteral stent resisted extrinsic compressive forces best. Although large-lumen stents are often used in this clinical situation, they were the most susceptible to compressive forces.
引用
收藏
页码:679 / 682
页数:4
相关论文
共 10 条
  • [1] 15-year experience with the management of extrinsic ureteral obstruction with indwelling ureteral stents
    Chung, SY
    Stein, RJ
    Landsittel, D
    Davies, BJ
    Cuellar, DC
    Hrebinko, RL
    Tarin, T
    Averch, TD
    [J]. JOURNAL OF UROLOGY, 2004, 172 (02) : 592 - 595
  • [2] HIGH FAILURE RATE OF INDWELLING URETERAL STENTS IN PATIENTS WITH EXTRINSIC OBSTRUCTION - EXPERIENCE AT 2 INSTITUTIONS
    DOCIMO, SG
    DEWOLF, WC
    [J]. JOURNAL OF UROLOGY, 1989, 142 (02) : 277 - 279
  • [3] THE DOUBLE-J URETERAL STENT - INVIVO AND INVITRO FLOW STUDIES
    HUBNER, WA
    PLAS, EG
    STOLLER, ML
    [J]. JOURNAL OF UROLOGY, 1992, 148 (02) : 278 - 280
  • [4] The use of 2 ipsilateral ureteral stents for relief of ureteral obstruction from extrinsic compression
    Liu, JS
    Hrebinko, RL
    [J]. JOURNAL OF UROLOGY, 1998, 159 (01) : 179 - 181
  • [5] MARDIS HK, 1982, UROL CLIN N AM, V9, P95
  • [6] METALLIC WALLSTENTS - A NEW THERAPY FOR EXTRINSIC URETERAL OBSTRUCTION
    PAUER, W
    LUGMAYR, H
    [J]. JOURNAL OF UROLOGY, 1992, 148 (02) : 281 - 284
  • [7] RESNICK MI, 1998, CAMPBELLS UROLOGY, V1, P387
  • [8] Management of malignant extrinsic compression of the ureter by simultaneous placement of two ipsilateral ureteral stents
    Rotariu, P
    Yohannes, P
    Alexianu, M
    Rosner, D
    Lee, BR
    Lucan, M
    Smith, AD
    [J]. JOURNAL OF ENDOUROLOGY, 2001, 15 (10) : 979 - 983
  • [9] New endoureteral double-J stent resists extrinsic ureteral compression
    Schlick, RW
    Seidl, EM
    Kalem, T
    Volkmer, B
    Planz, K
    [J]. JOURNAL OF ENDOUROLOGY, 1998, 12 (01) : 37 - 40
  • [10] SPIRAL-REINFORCED URETERAL STENT - AN ALTERNATIVE FOR INTERNAL URINARY-DIVERSION
    TSCHADA, RK
    HENKEL, TO
    JUNEMANN, KP
    RASSWEILER, J
    ALKEN, P
    [J]. JOURNAL OF ENDOUROLOGY, 1994, 8 (02) : 119 - 123