Electrical tuning for sensitivity enhancement of a piezo-electric ultrasonic transducer: Simulation and fabrication

被引:6
作者
Kim, Geonwoo [1 ,2 ]
Hwang, Young-In [3 ]
Seo, Mu-Kyung [3 ]
Kim, Ki-Bok [2 ,3 ]
机构
[1] ARS, Environm Microbial & Food Safety Lab, USDA, Powder Mill Rd,Bldg 303, Beltsville, MD USA
[2] Univ Sci & Technol, Dept Sci Measurement, 217 Gajeong Ro, Daejeon, South Korea
[3] Korea Res Inst Stand & Sci, Ctr Safety Measurement, 267 Gajeong Ro, Daejeon, South Korea
关键词
KLM model; Peizo-electric ultrasonic transducer; Electrical impedance tuning; PT SINGLE-CRYSTAL; PMN-PT; EQUIVALENT-CIRCUITS; 1-3; COMPOSITE; KLM MODEL; DESIGN; PZT;
D O I
10.1007/s12206-020-0707-1
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
In this study, two simple electrical impedance tuning methods derived from Krimholtz, Leedom, and Matthaei (KLM) modeling were proposed for increasing the sensitivity of a piezo-electric material-based ultrasonic transducer. In the first method, the reactance components of the ultrasonic transducer were removed, and in the other method, the electrical impedance of an ultrasonic transducer was matched to that of the ultrasonic system. To these ends, electrical circuits composed of inductors and capacitors were de-signed and KLM simulations of the two methods were performed. To verify the simulation results, ultrasonic transducers were fabricated and operated under the same conditions as in the simulations, and the experimental results were compared with the simulated results. In conclusion, the electrical tuning method (elimination of the reactance compo-nents) was more effective and the transducer circuit structure was simpler. The experi-mental results showed good agreement with the KLM simulation results.
引用
收藏
页码:3155 / 3164
页数:10
相关论文
共 29 条
[21]   Development and application of piezoelectric materials for ultrasound generation and detection [J].
Manbachi, Amir ;
Cobbold, Richard S. C. .
ULTRASOUND, 2011, 19 (04) :187-196
[22]  
Petersen G., 2006, L-matching the output of a RITEC gated amplifier to an arbitrary load
[23]  
San Emeterio J. L., 2002, Ferroelectrics, V273, P297
[24]  
Sherrit S, 1999, ULTRASON, P921, DOI 10.1109/ULTSYM.1999.849139
[25]   2-DIMENSIONAL ARRAYS FOR MEDICAL ULTRASOUND [J].
SMITH, SW ;
TRAHEY, GE ;
VONRAMM, OT .
ULTRASONIC IMAGING, 1992, 14 (03) :213-233
[26]   The piezo-electric resonator and its equivalent network [J].
Van Dyke, KS .
PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1928, 16 (06) :742-764
[27]  
Wang Q., 2020, MATER TECHNOL, P1
[28]   Zero-Bending Piezoelectric Micromachined Ultrasonic Transducer (pMUT) With Enhanced Transmitting Performance [J].
Wang, Tao ;
Lee, Chengkuo .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2015, 24 (06) :2083-2091
[29]  
Yang Y., 2017, SENSORS, V17