A 13.56MHz Wireless Power Transfer System with Dual-Output Regulated Active Rectifier for Implantable Medical Devices

被引:0
|
作者
Yang, Fu-Bin [1 ]
Fuh, Jonathan [1 ]
Chen, Po-Hung [1 ]
机构
[1] Natl Chiao Tung Univ, Inst Elect, Hsinchu, Taiwan
来源
2018 IEEE 61ST INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS) | 2018年
关键词
Active rectifier; wireless power transfer; resonant coupling; pulse skip modulation; automatic digital offset compensation; implantable medical devices;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this paper, a 13.56 MHz wireless power transfer system with dual-output regulated active rectifier is developed for implantable medical devices. The rectifier employs pulse-skip modulation (PSM) to regulate two output voltages without using power consuming low dropout regulators (LDOs). The proposed automatic digital offset compensation adjusts both turn-on timing and turn-off timing to compensate the turn-on and turn-off delays of the comparator. The simulation results show that the proposed active rectifier achieves 94% efficiency. The entire wireless power transmission system including power amplifier, resonant tanks, and rectifier, has 67.2% maximum power conversion efficiency (PCE). Compared to the conventional approach with ideal LDOs, 19.2% efficiency improvement is obtained.
引用
收藏
页码:440 / 443
页数:4
相关论文
共 50 条