Noninvasive method for measuring local hemoglobin oxygen saturation in tissue using wide gap second derivative near-infrared spectroscopy

被引:140
作者
Myers, DE [1 ]
Anderson, LD
Seifert, RP
Ortner, JP
Cooper, CE
Beilman, GJ
Mowlem, JD
机构
[1] Hutchinson Technol Inc, Hutchinson, MN 55350 USA
[2] Univ Essex, Dept Biol Sci, Colchester CO4 3SQ, Essex, England
[3] Univ Minnesota, Dept Surg, Minneapolis, MN 55455 USA
[4] Hutchinson Community Hosp, Emergency Serv, Hutchinson, MN 55350 USA
关键词
near infrared spectroscopy; StO(2); hemoglobin; second derivative; tissue hemoglobin oxygen saturation; muscle;
D O I
10.1117/1.1925250
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple continuous wave near-infrared algorithm for estimating local hemoglobin oxygen saturation in tissue (% StO(2)) is described using single depth attenuation measurements at 680, 720, 760, and 800 nm. Second derivative spectroscopy was used to reduce light scattering effects, chromophores with constant absorption, baseline/instrumentation drift, and movement artifacts. Unlike previous second derivative methods which focused primarily on measuring deoxyhemoglobin concentration; a wide 40 nm wavelength gap used for calculating second derivative attenuation significantly improved sensitivity to oxyhemoglobin absorption. Scaled second derivative attenuation at 720 nm was correlated to in vitro hemoglobin oxygen saturation to generate a % StO(2) calibration curve. The calibration curve was insensitive to total hemoglobin, optical path length, and optical scattering. Measurement error due to normal levels of carboxyhemoglobin, methemoglobin, and water absorption were less than 10 % StO(2) units. Severe methemoglobinemia or edema combined with low blood volume could cause StO(2) errors to exceed 10 StO2 units. Both a broadband and commercial four-wavelength spectrometer ( InSpectra (TM)) measured % StO(2). The InSpectra tissue spectrometer readily detected limb ischemia on 26 human volunteers for hand, forearm, and leg muscles. A strong linear correlation, r(2) > 0.93, between StO(2) and microvascular % SO2 was observed for isolated animal hind limb, kidney, and heart. (c) 2005 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页数:18
相关论文
共 74 条
[1]  
ANDERSON DL, 1999, Patent No. 5879294
[2]   THE OPTICS OF HUMAN-SKIN [J].
ANDERSON, RR ;
PARRISH, JA .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1981, 77 (01) :13-19
[3]   THE THEORETICAL BASIS FOR THE DETERMINATION OF OPTICAL PATHLENGTHS IN TISSUE - TEMPORAL AND FREQUENCY-ANALYSIS [J].
ARRIDGE, SR ;
COPE, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (07) :1531-1560
[4]  
BAUER JD, 1989, CLIN CHEM THEORY ANA, P515
[5]   Comparison of near-infrared spectroscopy and somatosensory evoked potentials for the detection of cerebral ischemia during carotid endarterectomy [J].
Beese, U ;
Langer, H ;
Lang, W ;
Dinkel, M .
STROKE, 1998, 29 (10) :2032-2037
[6]   Near-infrared and nuclear magnetic resonance spectroscopic assessment of tissue energetics in an isolated, perfused canine hind limb model of dysoxia [J].
Beilman, GJ ;
Myers, D ;
Cerra, FB ;
Lazaron, V ;
Dahms, RA ;
Conroy, MJ ;
Hammer, BE .
SHOCK, 2001, 15 (05) :392-397
[7]   Energy metabolism and interstitial fluid displacement in human gastrocnemius during short ischemic cycles [J].
Binzoni, T ;
Quaresima, V ;
Barattelli, G ;
Hiltbrand, E ;
Gürke, L ;
Terrier, F ;
Cerretelli, P ;
Ferrari, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 85 (04) :1244-1251
[8]  
Binzoni Tiziano, 2003, Journal of Physiological Anthropology and Applied Human Science, V22, P211, DOI 10.2114/jpa.22.211
[9]   Phase measurement of light absorption and scatter in human tissue [J].
Chance, B ;
Cope, M ;
Gratton, E ;
Ramanujam, N ;
Tromberg, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (10) :3457-3481
[10]  
Colier W N, 1995, Acta Anaesthesiol Scand Suppl, V107, P101