Computational and experimental investigation into mechanical performances of Poly-L-Lactide Acid (PLLA) coronary stents

被引:74
|
作者
Wang, Qian [1 ]
Fang, Gang [2 ]
Zhao, Yinghong [3 ]
Wang, Guohui [3 ]
Cai, Tao [3 ]
机构
[1] Tsinghua Univ, Dept Mech Engn, Beijing 100084, Peoples R China
[2] State Key Lab Tribol, Beijing 100084, Peoples R China
[3] Shanghai Bioheart Biol Technol Co Ltd, Suite 302,Bldg 9,590 Ruiqing Rd, Shanghai 201201, Peoples R China
基金
美国国家科学基金会; 北京市自然科学基金;
关键词
Biodegradable stent; PLLA; Finite element simulation; Radial property; Bending flexibility; BEHAVIOR; ARTERY; OPTIMIZATION; IMPLANTATION; DEFORMATION; DESIGN; METAL;
D O I
10.1016/j.jmbbm.2016.08.033
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Poly-L-lactide Acid (PLLA), as a credible biodegradable polymer-based material, can provide a promising amount of degradation time for vessel remodeling. Served as a sort of reliable intravascular implants, PLLA stents are expected to provide sufficient scaffolding to the target arteries without generating too much recoil after deployment. Besides, the stress and strain distribution should be as homogeneous as possible, and the stent conformability in fitting to the nature curvature of the vessels needs to be guaranteed. In the present study, mechanical performances of a stent made of PLLA material were investigated based on 3-D finite element method (FEM) and experiment verification. Simulations contained several deformation steps: crimping, spring-back after crimping, expanding and spring-back after expanding. The stent's deformation and stress/strain distributions were analyzed. Several indexes including the radial recoil ratio after crimping and expanding to different sizes, the radial properties including radial strength, the radial stiffness and the collapse pressure were established. In vitro static loading experiments of the stent were conducted as the verification of the FEM results, and a good agreement between them was obtained. Moreover, simulation of three-point bending was performed to assess the bending flexibility of the stent, and bending stiffness was defined as a measurement of structure resistance to the bending deformation. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:415 / 427
页数:13
相关论文
共 50 条
  • [1] Degradation modeling of poly-L-lactide acid (PLLA) bioresorbable vascular scaffold within a coronary artery
    Lin, Shengmao
    Dong, Pengfei
    Zhou, Changchun
    Dallan, Luis Augusto P.
    Zimin, Vladislav N.
    Pereira, Gabriel T. R.
    Lee, Juhwan
    Gharaibeh, Yazan
    Wilson, David L.
    Bezerra, Hiram G.
    Gu, Linxia
    NANOTECHNOLOGY REVIEWS, 2020, 9 (01) : 1217 - 1226
  • [2] Long-term in vivo degradation of poly-L-lactide (PLLA) in bone
    Walton, Mark
    Cotton, Nicholas J.
    JOURNAL OF BIOMATERIALS APPLICATIONS, 2007, 21 (04) : 395 - 411
  • [3] Biological evaluation of hydroxyapatite/poly-L-lactide (HAp/PLLA) composite biomaterials with poly-L-lactide of different molecular weights intraperitoneally implanted into mice
    Najman, S
    Savic, V
    Djordjevic, L
    Ignjatovic, N
    Uskokovic, D
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2004, 14 (01) : 61 - 70
  • [4] STRAIN-BASED DEGRADATION MODEL WITH APPLICATION TO POLY-L-LACTIDE ACID (PLLA) ARTERY STENT
    Lin, Shengmao
    Dong, Pengfei
    Gu, Linxia
    PROCEEDINGS OF ASME 2021 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION (IMECE2021), VOL 5, 2021,
  • [5] TUMORIGENICITY OF POLY-L-LACTIDE (PLLA) PLATES COMPARED WITH MEDICAL-GRADE POLYETHYLENE
    NAKAMURA, T
    SHIMIZU, Y
    OKUMURA, N
    MATSUI, T
    HYON, SH
    SHIMAMOTO, T
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (01): : 17 - 25
  • [6] Coating of stents with poly-L-lactide via fluidized-bed technology
    Wentzlaff, M.
    Seidlitz, A.
    Harder, C.
    Grabow, N.
    Weitschies, W.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S841 - +
  • [7] Cellular compatibility of β-calcium metaphosphate (β-CMP)/poly-L-lactide (PLLA) fracture innerfixation composites
    Liao, Li
    Yin, Guangfu
    Xie, Kenan
    Kang, Yunqing
    Long, Qin
    Lai, Xuefei
    Fuhe Cailiao Xuebao/Acta Materiae Compositae Sinica, 2009, 26 (04): : 59 - 62
  • [8] Bioabsorbable self-reinforced poly-L-lactide, metallic, and silicone stents in the management of experimental tracheal stenosis
    Korpela, A
    Aarnio, P
    Sariola, H
    Törmälä, P
    Harjula, A
    CHEST, 1999, 115 (02) : 490 - 495
  • [9] Optimization of the property profile of poly-L-lactide by synthesis of PLLA-polystyrene-block copolymers
    Dorff, Giselher
    Hahn, Mathias
    Laschewsky, Andre
    Lieske, Antje
    JOURNAL OF APPLIED POLYMER SCIENCE, 2013, 127 (01) : 120 - 126
  • [10] Hydroxyapatite/poly-L-lactide acid screws have better biocompatibility and femoral burr hole closure than does poly-L-lactide acid alone
    Akagi, Hiroyuki
    Iwata, Munetaka
    Ichinohe, Tomu
    Amimoto, Hirokazu
    Hayashi, Yuuta
    Kannno, Nobuo
    Ochi, Hiroki
    Fujita, Yukihiro
    Harada, Yasuji
    Tagawa, Masahiro
    Hara, Yasushi
    JOURNAL OF BIOMATERIALS APPLICATIONS, 2014, 28 (06) : 954 - 962