Homemade computed tomography setup with FDK reconstruction software

被引:2
|
作者
Kussainov, A. S. [1 ,2 ]
Saduev, N. O. [1 ,2 ]
机构
[1] Al Farabi Kazakh Natl Univ, Phys & Technol Dept, 71 Al Farabi Ave, Alma Ata 050040, Kazakhstan
[2] Natl Nanotechnol Lab Open Type, 71 Al Farabi Ave, Alma Ata 050040, Kazakhstan
来源
8TH INTERNATIONAL CONFERENCE ON MATHEMATICAL MODELING IN PHYSICAL SCIENCE | 2019年 / 1391卷
关键词
D O I
10.1088/1742-6596/1391/1/012086
中图分类号
O1 [数学];
学科分类号
0701 ; 070101 ;
摘要
X-ray source, detector and rotation stage as the sample holder were used to built the hardware part of the homemade CBCT scanner. The full scale numerical simulation of the CBCT experiment was modified to complete the scanner. This C/C++ software package is capable to process an arbitrary CT experimental or synthesized data for 3D reconstruction using the FDK method. The high degree of the package's flexibility allows different CBCT recording and reconstruction geometries with different source and detector positions and their spatial trajectory. The classical computed tomography arrangement, when the gantry, rotating around the sample, is loaded with the opposed sources and detectors, was replaced by the rotating sample and fixed source and detector. Mismatches between these two schemes in reconstruction procedure such as misplacement of the rotation axis from the expected location and other misalignments are compensated by introducing the corresponding variables in the reconstruction software.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Cochlear linear reconstruction with multidetector computed tomography
    Puyalto, P.
    Miro, N.
    Guerrero, D.
    Perez, M.
    Cuadras, P.
    Sanchez, J. J.
    CLINICAL OTOLARYNGOLOGY, 2017, 42 (06) : 1435 - 1437
  • [42] Differential reconstruction for planar object in computed tomography
    Liu, Tong
    Xu, Jian
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2009, 17 (02) : 101 - 114
  • [43] Regularization of Image Reconstruction in Ultrasound Computed Tomography
    Perez-Liva, Mailyn
    Herraiz, Joaquin L.
    Medina-Valdes, Luis
    Camacho, Jorge
    Fritsch, Carlos
    Treeby, Bradley E.
    Udias, Jose M.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [44] Applicability and potential of monitored reconstruction in computed tomography
    Gilmanov, Marat
    Bulatov, Konstantin
    Bugai, Oleg
    Ingacheva, Anastasia
    Chukalina, Marina
    Nikolaev, Dmitrii
    Arlazarov, Vladimir
    PLOS ONE, 2024, 19 (07):
  • [45] A FDK-based reconstruction method for off-centered circular trajectory cone beam tomography
    Valton, S.
    Peyrin, F.
    Sappey-Marinier, D.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) : 2736 - 2745
  • [46] Analysis of Daily Setup Variation with Tomotherapy Megavoltage Computed Tomography (MVCT)
    Zhou, J.
    Young, M.
    Uhl, B.
    Dewitt, K.
    Lo, Y.
    Fei, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [47] Realization of a computed tomography setup to achieve resolutions below 1 μm
    Salamon, M.
    Hanke, R.
    Krueger, P.
    Uhlmann, N.
    Voland, V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 591 (01): : 50 - 53
  • [48] Computed tomography algorithm based on diffraction-enhanced imaging setup
    Zhu, PP
    Wang, JY
    Yuan, QX
    Huang, WX
    Shu, H
    Gao, B
    Hu, TD
    Wu, ZY
    APPLIED PHYSICS LETTERS, 2005, 87 (26) : 1 - 3
  • [49] Validation and Application of a New Image Reconstruction Software toolbox (TIGRE) for Breast Cone-Beam Computed Tomography
    Kazemi, Shada
    Diaz, Oliver
    Elangovan, Premkumar
    Wells, Kevin
    Lohstroh, Annika
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [50] Measurement of proptosis using computed tomography based three-dimensional reconstruction software in patients with Graves’ orbitopathy
    Jung Huh
    Sang Joon Park
    Jeong Kyu Lee
    Scientific Reports, 10