Blood compatibility and adhesion of collagen/heparin multilayers coated on two titanium surfaces by a layer-by-layer technique

被引:17
作者
Chou, Chau-Chang [1 ,2 ]
Zeng, Hong-Jhih [1 ]
Yeh, Chi-Hsiao [3 ]
机构
[1] Natl Taiwan Ocean Univ, Dept Mech & Mechatron Engn, Keelung 202, Taiwan
[2] Natl Taiwan Ocean Univ, CMMS, Keelung 202, Taiwan
[3] Chang Gung Mem Hosp, Div Thorac & Cardiovasc Surg, Keelung 204, Taiwan
关键词
Heparin; Collagen; Titanium; Self-assembly; Multilayers; Blood compatibility; Adhesion; DIAMOND-LIKE CARBON; MECHANICAL-PROPERTIES; BIOMEDICAL ALLOY; FILMS; DLC; BIOCOMPATIBILITY; DEPOSITION; TI;
D O I
10.1016/j.tsf.2013.09.093
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper investigates the blood compatibility and adhesion of collagen/heparin multilayers coated on cp-Ti substrates with a layer-by-layer self-assembly technique. Two surface polishing processes were used for the titanium samples: one was mechanical polishing (MP) and the other, electropolishing (EP). These samples were pretreated by being immersed in NaOH solution to obtain a negatively charged surface with hydroxyl groups and then positively charged in poly-L-lysine solution. The repeated treatment of the samples by applying heparin and collagen alternately determined the number and thickness of the multilayers. The surface topography, chemical composition, and hydrophilicity of the films were investigated by atomic force microscopy, scanning electron microscopy, Fourier transform infrared spectroscopy, and water contact angle measurement. The study of the adhesion of the multilayer was conducted by a nano-scratch test. The blood compatibility was evaluated by measuring the hemolysis ratio and platelet-covered area in vitro. The uncoated titanium surface was used as the benchmark. The results indicated that the anticoagulation performance of collagen/heparin multilayers on the titanium surface was superior to that of the uncoated titanium surface. The hemolysis ratios of samples with an EP Ti substrate, a relatively rougher one, were essentially lower than those of samples with an MP substrate. The increase in the multilayers' thickness enhanced their adhesion to the Ti substrate. (C) 2013 Elsevier B. V. All rights reserved.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 19 条
[1]   Human micro-vascular endothelial cell seeding on Cr-DLC thin films for mechanical heart valve applications [J].
Ali, N. ;
Kousar, Y. ;
Okpalugo, T. I. ;
Singh, V. ;
Pease, M. ;
Ogwu, A. A. ;
Gracio, J. ;
Titus, E. ;
Meletis, E. I. ;
Jackson, M. J. .
THIN SOLID FILMS, 2006, 515 (01) :59-65
[2]   Improving blood-compatibility of titanium by coating collagen-heparin multilayers [J].
Chen, J. L. ;
Li, Q. L. ;
Chen, J. Y. ;
Chen, C. ;
Huang, N. .
APPLIED SURFACE SCIENCE, 2009, 255 (15) :6894-6900
[3]   Collagen/heparin coating on titanium surface improves the biocompatibility of titanium applied as a blood-contacting biomaterial [J].
Chen, Jialong ;
Chen, Cheng ;
Chen, Zhuoyue ;
Chen, Junying ;
Li, Quanli ;
Huang, Nan .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2010, 95A (02) :341-349
[4]   Characterization and haemocompatibility of fluorinated DLC and Si interlayer on Ti6Al4V [J].
Chou, Chau-Chang ;
Wu, Yi-Yang ;
Lee, Jyh-Wei ;
Yeh, Chi-Hsiao ;
Huang, Jen-Ching .
SURFACE & COATINGS TECHNOLOGY, 2013, 231 :418-422
[5]   Mechanical properties of fluorinated DLC and Si interlayer on a Ti biomedical alloy [J].
Chou, Chau-Chang ;
Wu, Yi-Yang ;
Lee, Jyh-Wei ;
Huang, Jen-Ching ;
Yeh, Chi-Hsiao .
THIN SOLID FILMS, 2013, 528 :136-142
[6]  
Crespilho FN, 2006, INT J ELECTROCHEM SC, V1, P194
[7]   A review of investigations on biocompatibility of diamond-like carbon and carbon nitride films [J].
Cui, FZ ;
Li, DJ .
SURFACE & COATINGS TECHNOLOGY, 2000, 131 (1-3) :481-487
[8]   Fuzzy nanoassemblies: Toward layered polymeric multicomposites [J].
Decher, G .
SCIENCE, 1997, 277 (5330) :1232-1237
[9]   Characterization and blood compatibility of TiCxN1-x hard coating prepared by plasma electrolytic carbonitriding [J].
Dong, Yan-Xia ;
Chen, Ya-Shao ;
Chen, Qiang ;
Liu, Bo ;
Song, Zhong-Xiao .
SURFACE & COATINGS TECHNOLOGY, 2007, 201 (21) :8789-8795
[10]   Film growth and relationship between microstructure and mechanical properties of a-C:H:F films deposited by PECVD [J].
Freire, FL ;
da Costa, MEHM ;
Jacobsohn, LG ;
Franceschini, DF .
DIAMOND AND RELATED MATERIALS, 2001, 10 (02) :125-131