Comparison and evaluation of Abbott chemiluminescent microparticle immunoassay and ChIVD light-initiated chemiluminescent assay in the detection ofTreponema pallidumantibody

被引:9
作者
Chen, Xiaohui [1 ]
Yang, Ruifeng [2 ]
Liang, Yongming [1 ]
Yuan, Teng [1 ]
Zhou, Jiansuo [1 ]
Wang, Tiancheng [1 ]
Cui, Liyan [1 ]
机构
[1] Peking Univ Third Hosp, Dept Lab Med, 49 North Garden Rd, Beijing 100191, Peoples R China
[2] Peking Univ, Beijing Key Lab Hepatitis C & Immunotherapy Liver, Inst Hepatol, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
Abbott chemiluminescent microparticle immunoassay; ChIVD light-initiated chemiluminescent assay; recombinant immunoblot assay; syphilis; Treponema pallidum; Treponema pallidumparticle agglutination; SYPHILIS; DISEASE;
D O I
10.1002/jcla.23275
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Laboratory tests play an important role in the diagnosis of syphilis. This study aimed to compare and assess the performance of the Abbott chemiluminescent microparticle immunoassay (CMIA) and the ChIVD light-initiated chemiluminescent assay (LICA) in the detection ofTreponema pallidum(TP) antibody. Methods A total of 10 498 serum samples were detected with two assays, and theTreponema pallidumparticle agglutination assay (TPPA) and recombinant immunoblot assay (RIBA) methods were used for confirmation. The sensitivity, specificity, positive predictive value, and negative predictive value of the Abbott CMIA and ChIVD LICA were calculated. The coincidence rate between two assays was also evaluated. The causes of false positive and false negative of two assays were studied. Results For the Abbott CMIA and ChIVD LICA, the sensitivity was 94.44% and 98.15%, the specificity was 99.89% and 99.81%, the positive predictive value was 93.29% and 88.83%, and the negative predictive value was 99.91% and 99.97%, respectively. The coincidence rate between Abbott CMIA and ChIVD LICA was 99.26%, and kappa value was .790. The disease of infertility, hypertensive disease, liver disease, and cancer were the common causes of false positive in both assays, while infertility was also the main reason lead to false negative. Conclusion Our results demonstrated that the Abbott CMIA and ChIVD LICA generally had high sensitivity and specificity and therefore may be suitable for the detection ofTPantibody and screening for syphilis.
引用
收藏
页数:6
相关论文
共 19 条
  • [1] Development of a light-initiated chemiluminescent assay for the quantitation of sIgE against egg white allergens based on component-resolved diagnosis
    Bian, Ying
    Liu, Changshan
    She, Tiantian
    Wang, Mengjuan
    Yan, Juanjuan
    Wei, Dianjun
    Li, Huiqiang
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2018, 410 (05) : 1501 - 1510
  • [2] Competitive light-initiated chemiluminescent assay: using 5--dihydrotestosterone-BSA as competitive antigen for quantitation of total testosterone in human sera
    Cui, Yaqiong
    She, Tiantian
    Zhao, Hong
    Li, Jiuzhi
    Li, Liuxu
    Gao, Weizhen
    Li, Huiqiang
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (03) : 745 - 754
  • [3] Evaluate performance of the Abbott chemiluminescent microparticle immunoassay assay for detection of syphilis infection in Chinese blood donors
    Fu, Ping
    Devare, Sushi
    Liu, Jian-fang
    Lv, Xiao-ting
    Yin, Peng
    Wu, Bing-ting
    Ke, Ling
    Liu, Yu
    Shan, Hua
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2020, 34 (01)
  • [4] Biological false-positive tests comprise a high proportion of Venereal Disease Research Laboratory reactions in an analysis of 300,000 sera
    Geusau, A
    Kittler, H
    Hein, U
    Dangl-Erlach, E
    Stingl, G
    Tschachler, E
    [J]. INTERNATIONAL JOURNAL OF STD & AIDS, 2005, 16 (11) : 722 - 726
  • [5] 2014 European guideline on the management of syphilis
    Janier, M.
    Hegyi, V.
    Dupin, N.
    Unemo, M.
    Tiplica, G. S.
    Potocnik, M.
    French, P.
    Patel, R.
    [J]. JOURNAL OF THE EUROPEAN ACADEMY OF DERMATOLOGY AND VENEREOLOGY, 2014, 28 (12) : 1581 - 1593
  • [6] Quantitation of estradiol by competitive light-initiated chemiluminescent assay using estriol as competitive antigen
    Li, Jiuzhi
    Li, Liuxu
    Bian, Ying
    Yu, Yang
    Qiang, Zhonghua
    Zhang, Yuexiang
    Li, Huiqiang
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2020, 34 (01)
  • [7] A light-initiated chemiluminescent assay for rapid quantitation of allergen-specific IgG4 in clinical samples
    Li, Junpu
    Li, Shaoshen
    Huang, Lunhui
    Cui, Yaqiong
    She, Tiantian
    Bian, Ying
    Li, Huiqiang
    [J]. CLINICA CHIMICA ACTA, 2019, 489 : 83 - 88
  • [8] Consistency Between Treponema pallidum Particle Agglutination Assay and Architect Chemiluminescent Microparticle Immunoassay and Characterization of Inconsistent Samples
    Li Zhiyan
    Wang Meiling
    Liu Ping
    Dong Jinhua
    Yan Zhenlin
    Feng Zhenru
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2015, 29 (04) : 281 - 284
  • [9] Characterization of the classical biological false-positive reaction in the serological test for syphilis in the modern era
    Liu, Fan
    Liu, Li-Li
    Guo, Xiao-Jing
    Xi, Ya
    Lin, Li-Rong
    Zhang, Hui-Lin
    Huang, Song-Jie
    Chen, Yu-Yan
    Zhang, Ya-Feng
    Zhang, Qiao
    Huang, Ge-Ling
    Tong, Man-Li
    Jiang, Jie
    Yang, Tian-Ci
    [J]. INTERNATIONAL IMMUNOPHARMACOLOGY, 2014, 20 (02) : 331 - 336
  • [10] Comparison of Traditional and Reverse Syphilis Screening Algorithms in Medical Health Checkups
    Nah, Eun-Hee
    Cho, Seon
    Kim, Suyoung
    Cho, Han-Ik
    Chai, Jong-Yil
    [J]. ANNALS OF LABORATORY MEDICINE, 2017, 37 (06) : 511 - 515