The development, validity, and reliability of a manual muscle testing device with integrated limb position sensors

被引:38
作者
Li, RC
Jasiewicz, JM [1 ]
Middleton, J
Condie, P
Barriskill, A
Hebnes, H
Purcell, B
机构
[1] Queensland Univ Technol, Fac Hlth, Sch Human Movement Studies, Ctr Hlth Res, Kelvin Grove, Qld 4059, Australia
[2] Queensland Univ Technol, Inst Hlth Biomed Innovat, Kelvin Grove, Qld 4059, Australia
[3] Univ Sydney, Fac Med, Rehabil Studies Unit, Sydney, NSW 2006, Australia
[4] Cochlear Proprietary Ltd, Lane Cove, NSW, Australia
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 2006年 / 87卷 / 03期
关键词
muscles; rehabilitation; reliability and validity; transducers;
D O I
10.1016/j.apmr.2005.11.011
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Objective: To report the development and validation of a new hand-held muscle strength-testing device that is integrated with orientation sensors and designed to test the strength of major muscle groups at a given limb or joint position. Design: Design description and validation study. Setting: University-based human movement facility. Participants: Twenty-eight able-bodied, healthy subjects. Interventions: Not applicable. Main Outcome Measure: A device was developed based on a hand-held force dynamometer with integrated orientation sensors. The validity and reliability (interrater, intertrial) of 5 maximum isometric contractions of hip flexion, knee extension, and ankle plantarflexion and dorsiflexion were assessed. The results were compared with those from an isokinetic dynamometer (KinCom). Results: The new manual muscle tester was highly reliable and valid in estimating muscle strength of the lower limbs. The coefficient of variation between trials of all movements was low, with a mean less than 10% (range, 3.7%-8.9%). The only significant difference in muscle strength between the new device and the isokinetic dynamometer was found for hip flexion. Conclusions: The new hand-held muscle strength tester appears to be a reliable and valid clinical assessment tool that can be used to objectively assess muscle strength at particular limb positions and/or joint angles. This feature appears to represent a technical advance in portable muscle strength devices, providing comparable information to those obtained by isokinetic dynamometers at a fraction of the cost and size. However. the device needs to be validated in clinical populations, such as patients with spinal cord injury and stroke, in order to demonstrate its general clinical utility.
引用
收藏
页码:411 / 417
页数:7
相关论文
共 30 条
[1]  
ARGE JC, 1987, ARCH PHYS MED REHAB, V68, P454
[2]   Reliability of isokinetic ankle inversion- and eversion-strength measurement in neutral foot position, using the Biodex dynamometer [J].
Aydog, E ;
Aydog, ST ;
Çakci, A ;
Doral, MN .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2004, 12 (05) :478-481
[3]   Inter- and intrarater reliability of the Ashworth Scale and the Disability Assessment Scale in patients with upper-limb poststroke spasticity [J].
Brashear, A ;
Zafonte, R ;
Corcoran, M ;
Galvez-Jimenez, N ;
Gracies, JM ;
Gordon, MF ;
Mcafee, A ;
Ruffing, K ;
Thompson, B ;
Williams, M ;
Lee, CH ;
Turkel, C .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2002, 83 (10) :1349-1354
[4]  
BYL N N, 1988, Journal of Orthopaedic and Sports Physical Therapy, V9, P399
[5]  
Chan KM., 1996, PRINCIPLES PRACTICE
[6]   Measurement of isometric muscle strength: a reproducibility study of maximal voluntary contraction in normal subjects and amyotrophic lateral sclerosis patients [J].
Colombo, R ;
Mazzini, L ;
Mora, G ;
Parenzan, R ;
Creola, G ;
Pirali, I ;
Minuco, G .
MEDICAL ENGINEERING & PHYSICS, 2000, 22 (03) :167-174
[7]   Testing isokinetic muscular fatigue of shoulder internal and external rotation in elite junior tennis players [J].
Ellenbecker, TS ;
Roetert, EP .
JOURNAL OF ORTHOPAEDIC & SPORTS PHYSICAL THERAPY, 1999, 29 (05) :275-281
[8]   Reliability of stabilised commercial dynamometers for measuring hip abduction strength: a pilot study [J].
Fenter, PC ;
Bellew, JW ;
Pitts, TA ;
Kay, RE .
BRITISH JOURNAL OF SPORTS MEDICINE, 2003, 37 (04) :331-334
[9]   Reliability of stationary dynamometer muscle strength testing in community-dwelling older adults [J].
Ford-Smith, CD ;
Wyman, JF ;
Elswick, RK ;
Fernandez, T .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2001, 82 (08) :1128-1132
[10]   CLINICAL RELIABILITY OF MANUAL MUSCLE TESTING - MIDDLE TRAPEZIUS AND GLUTEUS MEDIUS MUSCLES [J].
FRESE, E ;
BROWN, M ;
NORTON, BJ .
PHYSICAL THERAPY, 1987, 67 (07) :1072-1076