Accuracy of three-dimensional (3D) craniofacial cephalometric landmarks on a low-dose 3D computed tomograph

被引:31
|
作者
Olszewski, R. [1 ]
Reychler, H. [1 ]
Cosnard, G. [2 ]
Denis, J. M. [3 ]
Vynckier, S. [3 ]
Zech, F. [4 ]
机构
[1] Catholic Univ Louvain, Clin Univ St Luc, Dept Oral & Maxillofacial Surg, B-1200 Brussels, Belgium
[2] Catholic Univ Louvain, Clin Univ St Luc, Dept Med Imaging, B-1200 Brussels, Belgium
[3] Catholic Univ Louvain, Clin Univ St Luc, Dept Radiotherapy & Oncol, B-1200 Brussels, Belgium
[4] Catholic Univ Louvain, Clin Univ St Luc, Dept Internal Med, B-1200 Brussels, Belgium
关键词
three-dimensional; computed tomography; cephalometry; dose reduction;
D O I
10.1259/dmfr/33343444
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim of this paper is to compare the accuracy of cephalometric landmark identification using three-dimensional CT (3D-CT) surface rendering with "high-dose" (200 mAs) and "low-dose" (35 mAs) CT protocols. The absorbed dose levels for radiosensitive organs in the maxillofacial region during the exposure for both 3D-CT protocols were also measured. Methods: The study population consisted of 15 human dry skulls examined with spiral 3D-CT. 12 cephalometric anatomical landmarks at 7 sites were identified on the 3D-CT surface renderings by 2 observers independently, twice each, using high-dose and low-dose protocols. In total, 1440 imaging measurements were made. Thermoluminescent dosemeters (TLDs) were placed at ten sites around the thyroid and submandibular glands and the eyes in an Alderson phantom for measuring the absorbed dose levels. Results: The intraobserver mean distances between 3D landmarks were smaller for all sites with the high-dose protocol (P = 0.37). There was a significant difference among the observers (P = 0.000004). Interobserver mean distances between 3D landmarks were smaller for four of the seven sites with the low-dose protocol. However, the global interobserver mean distances between 3D landmarks for all sites were smaller with the high-dose protocol (P = 0.028). The low-dose protocol reduced the radiation dose to the thyroid by 6.12, to the submandibular salivary glands by 5.91 and to the eye by 5.44, resulting in a global reduction factor of 5.71. Conclusions: The accuracy in the landmark's identification was maintained when the milliampere-second values were reduced from 200 mAs to 35 mAs. We recommend use of the low-dose protocol for clinical 3D-CT cephalometric applications.
引用
收藏
页码:261 / 267
页数:7
相关论文
共 50 条
  • [1] A study on the reproducibility of cephalometric landmarks when undertaking a three-dimensional (3D) cephalometric analysis
    Zamora, Natalia
    Llamas, Jose-Maria
    Cibrian, Rosa
    Gandia, Jose-Luis
    Paredes, Vanessa
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2012, 17 (04): : E678 - E688
  • [2] Reproducibility of three-dimensional cephalometric landmarks in cone-beam and low-dose computed tomography
    R. Olszewski
    L. Frison
    M. Wisniewski
    J. M. Denis
    S. Vynckier
    G. Cosnard
    F. Zech
    H. Reychler
    Clinical Oral Investigations, 2013, 17 : 285 - 292
  • [3] Reproducibility of three-dimensional cephalometric landmarks in cone-beam and low-dose computed tomography
    Olszewski, R.
    Frison, L.
    Wisniewski, M.
    Denis, J. M.
    Vynckier, S.
    Cosnard, G.
    Zech, F.
    Reychler, H.
    CLINICAL ORAL INVESTIGATIONS, 2013, 17 (01) : 285 - 292
  • [4] Three-dimensional (3D) facial recognition and prediction
    Paul, Okuwobi Idowu
    Lu, Yonghua
    International Journal of Electrical Engineering, 2015, 22 (02): : 63 - 72
  • [5] Three-dimensional (3D) facial recognition and prediction
    Okuwobi, Idowu Paul
    Chen, Qiang
    Niu, Sijie
    Bekalo, Loza
    SIGNAL IMAGE AND VIDEO PROCESSING, 2016, 10 (06) : 1151 - 1158
  • [6] Three-dimensional (3D) facial recognition and prediction
    Idowu Paul Okuwobi
    Qiang Chen
    Sijie Niu
    Loza Bekalo
    Signal, Image and Video Processing, 2016, 10 : 1151 - 1158
  • [7] A 3D cephalometric protocol for the accurate quantification of the craniofacial symmetry and facial growth
    Pinheiro, Manuel
    Ma, Xinhui
    Fagan, Michael J.
    McIntyre, Grant T.
    Lin, Ping
    Sivamurthy, Gautham
    Mossey, Peter A.
    JOURNAL OF BIOLOGICAL ENGINEERING, 2019, 13
  • [8] A 3D cephalometric protocol for the accurate quantification of the craniofacial symmetry and facial growth
    Manuel Pinheiro
    Xinhui Ma
    Michael J. Fagan
    Grant T. McIntyre
    Ping Lin
    Gautham Sivamurthy
    Peter A. Mossey
    Journal of Biological Engineering, 13
  • [9] Three-dimensional (3D) area and volume measurements for rhinoplasty
    Celikoyar, M. Mazhar
    Topsakal, Oguzhan
    Sawyer, Philip
    JOURNAL OF PLASTIC RECONSTRUCTIVE AND AESTHETIC SURGERY, 2023, 83 : 189 - 197
  • [10] 3D ultraslinography is as accurate as low-dose CT in thyroid volumetry
    Licht, K.
    Darr, A.
    Opfermann, T.
    Winkens, T.
    Freesmeyer, M.
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2014, 53 (03): : 99 - 104