All-Semiconductor High-Speed Akinetic Swept-Source for OCT

被引:0
|
作者
Minneman, Michael P. [1 ]
Ensher, Jason [1 ]
Crawford, Michael [1 ]
Derickson, Dennis [2 ]
机构
[1] Insight Photon Solut Inc, 300 S Public Rd, Lafayette, CO 80026 USA
[2] Calif Polytech State Univ San Luis Obispo, Dept Elect Engn, San Luis Obispo, CA 93407 USA
关键词
OCT; Optical Coherence Tomography; swept-source OCT; SS-OCT; tunable laser; VT-DBR; OFDR; FBG; OPTICAL COHERENCE TOMOGRAPHY; LASER;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel swept-wavelength laser for optical coherence tomography (OCT) using a monolithic semiconductor device with no moving parts is presented. The laser is a Vernier-Tuned Distributed Bragg Reflector (VT-DBR) structure exhibiting a single longitudinal mode. All-electronic wavelength tuning is achieved at a 200 kHz sweep repetition rate, 20 mW output power, over 100 run sweep width and coherence length longer than 40 mm. OCT point-spread functions with 4555 dB dynamic range are demonstrated; lasers at 1550 run, and now 1310 run, have been developed. Because the laser's long-term tuning stability allows for electronic sample trigger generation at equal k-space intervals (electronic k-clock), the laser does not need an external optical k-clock for measurement interferometer sampling. The non-resonant, all-electronic tuning allows for continuously adjustable sweep repetition rates from mHz to 100s of kHz. Repetition rate duty cycles are continuously adjustable from single-trigger sweeps to over 99% duty cycle. The source includes a monolithically integrated power leveling feature allowing flat or Gaussian power vs. wavelength profiles. Laser fabrication is based on reliable semiconductor wafer-scale processes, leading to low and rapidly decreasing cost of manufacture.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Coherence transfer in an akinetic swept source OCT laser with optical feedback
    Slepneva, S.
    Kovalev, A.
    Rebrova, N.
    Grigorenko, K.
    Viktorov, E.
    Huyet, G.
    OPTICS LETTERS, 2019, 44 (21) : 5161 - 5164
  • [32] Feasibility of swept-source OCT for active birdshot chorioretinopathy
    Olga Garcia-Garcia
    Sara Jordan-Cumplido
    Olaia Subira-Gonzalez
    Pere Garcia-Bru
    Luis Arias
    Josep M. Caminal-Mitjana
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2017, 255 : 1493 - 1502
  • [33] Wavelength encoded OCT imaging using swept-source
    Wang, Ling
    Ding, Zhihua
    Wu, Tong
    Wang, Kai
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS III, 2008, 6826
  • [34] Feasibility of swept-source OCT for active birdshot chorioretinopathy
    Garcia-Garcia, Olga
    Jordan-Cumplido, Sara
    Subira-Gonzalez, Olaia
    Garcia-Bru, Pere
    Arias, Luis
    Caminal-Mitjana, Josep M.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2017, 255 (08) : 1493 - 1502
  • [35] Akinetic swept source with adjustable coherence length for SS-OCT
    Stancu, Radu F.
    Jackson, David A.
    Podoleanu, Adrian Gh.
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XIX, 2015, 9312
  • [36] Re: Michalewska et al: Swept-Source OCT and Swept-Source OCT Angiography before and after Vitrectomy with Stuffing the Optic Pit REPLY
    Michalewska, Zofia
    Nawrocka, Zofia
    Nawrocki, Jerzy
    OPHTHALMOLOGY RETINA, 2020, 4 (09): : E9 - E9
  • [37] High-Speed, Long-Range Swept-Source Optical Coherence Tomography for the Anterior Segment of the Eye
    Chen, Jyu
    Potsaid, Benjamin
    Lin, Junhong
    Hwang, Yunchan
    Moult, Eric
    Li, Yan
    Huang, David
    Fujimoto, James G.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (11)
  • [38] Akinetic Swept-Source Master-Slave-Enhanced Optical Coherence Tomography
    Marques, Manuel J.
    Cernat, Ramona
    Ensher, Jason
    Bradu, Adrian
    Podoleanu, Adrian
    PHOTONICS, 2021, 8 (05)
  • [39] Swept-Source OCT Imaging of the Argus II Epiretinal Prosthesis
    Rodriguez, Marianeli
    Gregori, Ninel Z.
    Roisman, Luiz
    Lam, Byron L.
    Rosenfeld, Philip J.
    OPHTHALMOLOGY RETINA, 2018, 2 (04): : 380 - 382
  • [40] High-speed, digitally refocused retinal imaging with Line-field parallel swept source OCT
    Fechtig, Daniel J.
    Kumar, Abhishek
    Ginner, Laurin
    Drexler, Wolfgang
    Leitgeb, Rainer A.
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XIX, 2015, 9312