Multicommuted flow system for the determination of glucose in animal blood serum exploiting enzymatic reaction and chemiluminescence detection

被引:0
|
作者
Pires, CK
Martelli, PB
Reis, BF
Lima, JLFC
Saraiva, MLMFS
机构
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400870 Piracicaba, SP, Brazil
[2] Univ Fed Sao Carlos, Dept Quim, BR-13560 Sao Carlos, SP, Brazil
[3] Univ Fed Sao Joao Del Rei, Dept Ciencias Nat, Sao Joao Del Rei, MG, Brazil
[4] Univ Porto, Fac Farm, Dept Quim Fis, REQUIMTE, P-4100 Oporto, Portugal
关键词
D O I
10.1155/S1463924603000191
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automatic flow procedure based on multicommutation dedicated for the determination of glucose in animal blood serum using glucose oxidase with chemiluminescence detection is described. The flow manifold consisted of a set of three- way solenoid valves assembled to implement multicommutation. A microcomputer furnished with an electronic interface and software written in Quick BASIC 4.5 controlled the manifold and performed data acquisition. Glucose oxidase was immobilized on porous silica beads ( glass aminopropyl) and packed in a minicolumn ( 15 x 5 mm). The procedure was based on the enzymatic degradation of glucose, producing hydrogen peroxide, which oxidized luminol in the presence of hexacyanoferrate( III), causing the chemiluminescence. The system was tested by analysing a set of serum animal samples without previous treatment. Results were in agreement with those obtained with the conventional method ( LABTEST Kit) at the 95% confidence level. The detection limit and variation coefficient were estimated as 12.0 mg l(-1) ( 99.7% confidence level) and 3.5% ( n = 20), respectively. The sampling rate was about 60 determinations h(-1) with sample concentrations ranging from 50 to 600 mg l(-1) glucose. The consumptions of serum sample, hexacyanoferrate( III) and luminol were 46 mul, 10.0 mg and 0.2 mg/ determination, respectively.
引用
收藏
页码:109 / 114
页数:6
相关论文
共 50 条