Comparison of stroke volume measurements during hemodialysis using bioimpedance cardiography and echocardiography

被引:15
|
作者
Germain, Michael J. [1 ,2 ]
Joubert, Jyovani [3 ]
O'Grady, Daniel [4 ]
Nathanson, Brian H. [5 ]
Chait, Yossi [6 ]
Levin, Nathan W. [7 ]
机构
[1] Baystate Med Ctr, Springfield, MA USA
[2] Univ Massachusetts, Sch Med, Worcester, MA USA
[3] Kidney Care & Transplant Associates New England, Springfield, MA USA
[4] Mercy Med Ctr, Springfield, MA USA
[5] OptiStatim LLC, Longmeadow, MA USA
[6] Univ Massachusetts, Amherst, MA 01003 USA
[7] Mt Sinai Ichan Sch Med, New York, NY USA
基金
美国国家卫生研究院;
关键词
Stroke volume; echocardiography; bioimpedance cardiography; hemodialysis; CARDIAC-OUTPUT; BLOOD-VOLUME; IMPEDANCE CARDIOGRAPHY; DIALYSIS; HYPOTENSION; THERMODILUTION; CARDIOMYOPATHY; MORTALITY; RISK;
D O I
10.1111/hdi.12589
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: Fluid management remains a major challenge of hemodialysis (HD) care, with serious implications for morbidity and mortality. Intradialytic fluid management is typically guided by blood pressure, an indirect resultant of hemodynamics status. Direct measurements of hemodynamic parameters may improve cardiovascular outcomes by providing rational bases for intervention. We compare stroke volume (SV) measurements using a noninvasive, regional biompedance cardiography device (NiCaS) with Doppler echocardiography (Echo) in HD setting. Methods: Stroke volumes were simultaneously measured using the devices in 17 patients receiving maintenance HD. Measurements were made during 2 weekly HD treatments, and twice within each HD treatment during the first and last hour of each treatment, for a total of 64 SV measurements. Agreement between devices was assessed using linear regression, a Pearson's correlation coefficient, and a Bland-Altman plot all adjusted for repeated measures within patients. Results: Echo and NiCaS SV mean and 95% CIs were 58.0 (50.1, 65.8) and 56.7 (49.4, 64.0) mL, respectively. NiCaS SV correlated strongly with Echo SV during the first and last hours of treatments (r=0.93, P<0.001 and r=0.92, P<0.001, respectively). Linear regression of NiCaS on Echo showed a slope of 0.97, 95% CI (0.91, 1.02) which did not differ from 1, P=0.20. A Bland-Altman plot and 4-Quadrant plot confirmed that the 2 methods produced comparable measurements. Conclusion: NiCaS SV measurements are similar to and strongly correlated with Echo SV measurements. This suggests that noninvasive NiCaS technology may be a practical method for measuring SV during HD.
引用
收藏
页码:201 / 208
页数:8
相关论文
共 50 条
  • [1] Stroke volume assessment by impedance cardiography: Comparative analysis with transthoracic echocardiography
    Silluta, S.
    Pilt, K.
    Bischler, E.
    Koots, K.
    Meigas, K.
    Viigimaa, M.
    EMBEC & NBC 2017, 2018, 65 : 398 - 401
  • [2] Impedance cardiography derived cardiac output in hemodialysis patients: A study of reproducibility and comparison with echocardiography
    Karakitsos, D. N.
    Patrianakos, A. P.
    Paraskevopoulos, A.
    Parthenakis, F. I.
    Tzenakis, N.
    Fourtounas, C.
    Daphnis, E. K.
    Vardas, P. E.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (06): : 564 - 572
  • [3] Beat-to-beat estimation of stroke volume using impedance cardiography and artificial neural network
    Naidu, S. M. M.
    Pandey, Prem C.
    Bagal, Uttam R.
    Hardas, Suhas P.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2018, 56 (06) : 1077 - 1089
  • [4] Stroke volume and cardiac output measurement in cardiac patients during a rehabilitation program: comparison between tonometry, impedancemetry and echocardiography
    Gonzalez-Represas, Alicia
    Mourot, Laurent
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2020, 36 (03): : 447 - 455
  • [5] Comparison of Impedance Cardiography and Thermodilution-Derived Measurements of Stroke Volume and Cardiac Output at Rest and During Exercise Testing
    A. Scherhag
    J. J. Kaden
    E. Kentschke
    T. Sueselbeck
    M. Borggrefe
    Cardiovascular Drugs and Therapy, 2005, 19 : 141 - 147
  • [6] Comparison of impedance cardiography and thermodilution-derived measurements of stroke volume and cardiac output at rest and during exercise testing
    Scherhag, A
    Kaden, JK
    Kentschke, E
    Sueselbeck, T
    Borggrefe, M
    CARDIOVASCULAR DRUGS AND THERAPY, 2005, 19 (02) : 141 - 147
  • [7] Comparison of whole body versus thoracic bioimpedance in relation to ultrafiltration volume and systolic blood pressure during hemodialysis
    Schoutteten, Melanie K.
    Lindeboom, Lucas
    Brys, Astrid
    Lanssens, Dorien
    Smeets, Christophe J. P.
    De Canniere, Helene
    De Moor, Bart
    Peeters, Jacques
    Heylen, Line
    Van Hoof, Chris
    Groenendaal, Willemijn
    Kooman, Jeroen P.
    Vandervoort, Pieter M.
    JOURNAL OF APPLIED PHYSIOLOGY, 2023, 135 (06) : 1330 - 1338
  • [8] Reliability of the Measurement of Stroke Volume Using Impedance Cardiography During Acute Cold Exposure
    Muller, Matthew D.
    Ryan, Edward J.
    Kim, Chul-Ho
    Bellar, David M.
    Blankfield, Robert P.
    Glickman, Ellen L.
    AVIATION SPACE AND ENVIRONMENTAL MEDICINE, 2010, 81 (02): : 120 - 124
  • [9] Standardization of non-invasive impedance cardiography for assessment of stroke volume: Comparison with thermodilution
    Woltjer, HH
    Bogaard, HJ
    Scheffer, GJ
    vanderSpoel, HI
    Huybregts, MAJM
    deVries, PMJM
    BRITISH JOURNAL OF ANAESTHESIA, 1996, 77 (06) : 748 - 752
  • [10] Stroke volume and cardiac output measurement in cardiac patients during a rehabilitation program: comparison between tonometry, impedancemetry and echocardiography
    Alicia Gonzalez-Represas
    Laurent Mourot
    The International Journal of Cardiovascular Imaging, 2020, 36 : 447 - 455