Reproducibility of Parenchymal Blood Volume Measurements Using an Angiographic C-arm CT System

被引:5
作者
Mueller, Kerstin [1 ]
Fahrig, Rebecca [1 ]
Manhart, Michael [2 ]
Deuerling-Zheng, Yu [2 ]
Rosenberg, Jarrett [1 ]
Moore, Teri [3 ]
Ganguly, Arundhuti [1 ]
Kothary, Nishita [1 ,4 ]
机构
[1] Stanford Univ, Dept Radiol, 1201 Welch Rd, Stanford, CA 94305 USA
[2] Siemens Healthcare GmbH, Forchheim, Germany
[3] Siemens Med Solut Inc, Malvern, PA USA
[4] Stanford Univ, Med Ctr, Dept Radiol, Stanford, CA 94305 USA
关键词
Parenchymal blood volume (PBV); CBCT; porcine study; chemoembolization; FLAT-DETECTOR CT; TRANSARTERIAL CHEMOEMBOLIZATION; HEPATOCELLULAR-CARCINOMA; COMPUTED-TOMOGRAPHY; LIVER-DISEASE; PERFUSION; TUMOR; SUITE; INTERVENTIONS; SOFTWARE;
D O I
10.1016/j.acra.2016.08.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Intra-procedural measurement of hepatic perfusion following liver embolization continues to be a challenge. Blood volume imaging before and after interventional procedures would allow identifying the treatment end point or even allow predicting treatment outcome. Recent liver oncology studies showed the feasibility of parenchymal blood volume (PBV) imaging using an angiographic C-arm system. This study was done to evaluate the reproducibility of PBV measurements using cone beam computed tomography (CBCT) before and after embolization of the liver in a swine model. Materials and,Methods: CBCT imaging was performed before and after partial bland embolization of the left lobe of the liver in five adult pigs. Intra-arterial injection of iodinated contrast with a 6-second x-ray delay was used with a two-sweep 8-second rotation imaging protocol. Three acquisitions, each separated by 10 minutes to allow for contrast clearance, were obtained before and after embolization in each animal. Post-processing was carried out using dedicated software to generate three-dimensional (3D) PBV maps. Two region of-interest measurements were placed on two views within the right and left lobe on each CBCT 3D PBV map. Variation in PBV for scans acquired within each animal was determined by the coefficient of variation and intraclass correlation. A Wilcoxon signed-rank test was used to test post-procedure reduction in PBV. Results: The CBCT PBV maps showed mean coefficients of variation of 7% (range: 2%-16%) and 25% (range: 13%-34%) for baseline and embolized PBV maps, respectively. The intraclass correlation for PBV measurements was 0.89, demonstrating high reproducibility, with measurable reduction in PBV displayed after embolization (P = 0.007). Conclusions: Intra-procedural acquisition of 3D PBV maps before and after liver embolization using CBCT is highly reproducible and shows promising application for obtaining intra-procedural PBV maps during locoregional therapy.
引用
收藏
页码:1441 / 1445
页数:5
相关论文
共 50 条
  • [11] Image quality of arterial phase and parenchymal blood volume (PBV) maps derived from C-arm computed tomography in the evaluation of transarterial chemoembolization
    Zitzelsberger, Tanja
    Syha, Roland
    Groezinger, Gerd
    Partovi, Sasan
    Nikolaou, Konstantin
    Grosse, Ulrich
    CANCER IMAGING, 2018, 18
  • [12] Technical Feasibility of CT Perfusion using a C-Arm CBCT System
    Xu, Min
    Tang, Jie
    Royalty, Kevin
    Chen, Guang-Hong
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [13] Visualization of hypervascular liver lesions during TACE: Comparison of angiographic C-arm CT and MDCT
    Meyer, Bernhard C.
    Frericks, Bernd B.
    Voges, Maerthe
    Borchert, Michael
    Martus, Peter
    Justiz, Joern
    Wolf, Karl-Juergen
    Wacker, Frank K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (04) : W263 - W269
  • [14] C-Arm CT Measurement of Cerebral Blood Volume in Ischemic Stroke: An Experimental Study in Canines
    Bley, T.
    Strother, C. M.
    Pulfer, K.
    Royalty, K.
    Zellerhoff, M.
    Deuerling-Zheng, Y.
    Bender, F.
    Consigny, D.
    Yasuda, R.
    Niemann, D.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (03) : 536 - 540
  • [15] C-arm computed tomography parenchymal blood volume measurement in evaluation of hepatocellular carcinoma before transarterial chemoembolization with drug eluting beads
    Roland Syha
    Gerd Grözinger
    Ulrich Grosse
    Michael Maurer
    Lars Zender
    Marius Horger
    Konstantin Nikolaou
    Dominik Ketelsen
    Cancer Imaging, 15
  • [16] Quantitative Evaluation of C-Arm CT Cerebral Blood Volume in a Canine Model of Ischemic Stroke
    van der Born, I. M. J.
    Mehra, M.
    Walvick, R. P.
    Chueh, J. -Y.
    Gounis, M. J.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (02) : 353 - 358
  • [17] Dynamic Measurement of Arterial Liver Perfusion With an Interventional C-Arm System
    Datta, Sanjit
    Mueller, Kerstin
    Moore, Terilyn
    Molvin, Lior
    Gehrisch, Sonja
    Rosenberg, Jarrett
    Saenz, Yamil
    Manhart, Michael
    Deuerling-Zheng, Yu
    Kothary, Nishita
    Fahrig, Rebecca
    INVESTIGATIVE RADIOLOGY, 2017, 52 (08) : 456 - 461
  • [18] Image quality of arterial phase and parenchymal blood volume (PBV) maps derived from C-arm computed tomography in the evaluation of transarterial chemoembolization
    Tanja Zitzelsberger
    Roland Syha
    Gerd Grözinger
    Sasan Partovi
    Konstantin Nikolaou
    Ulrich Grosse
    Cancer Imaging, 18
  • [19] C-Arm CT Measurement of Cerebral Blood Volume and Cerebral Blood Flow Using a Novel High-Speed Acquisition and a Single Intravenous Contrast Injection
    Royalty, K.
    Manhart, M.
    Pulfer, K.
    Deuerling-Zheng, Y.
    Strother, C.
    Fieselmann, A.
    Consigny, D.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (11) : 2131 - 2138
  • [20] Comparison of air kerma between C-arm CT and 64-multidetector-row CT using a phantom
    Hosokawa, Seiki
    Kawai, Nobuyuki
    Sato, Morio
    Minamiguchi, Hiroki
    Nakai, Motoki
    Murotani, Kazuhiro
    Nishioku, Tadayoshi
    Shirai, Shintaro
    Sonomura, Tetsuo
    JAPANESE JOURNAL OF RADIOLOGY, 2014, 32 (05) : 302 - 307