1.5-D high intensity focused ultrasound array for non-invasive prostate cancer surgery

被引:47
作者
Curiel, L [1 ]
Chavrier, F [1 ]
Souchon, R [1 ]
Birer, A [1 ]
Chapelon, JY [1 ]
机构
[1] INSERM, U556, F-69424 Lyon 03, France
关键词
D O I
10.1109/58.985707
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The aim of this study is to demonstrate the feasibility of a new spherically curved 1.5-D phased array for the treatment of localized prostatic cancer. The device is designed to conform to the Ablatherm(R) machine (EDAP-Technomed, France), a commercially available machine in which high intensity focused ultrasound (HIFU) treatment for prostate cancer is administered transrectally. It uses high intensity electronically focused ultrasound to steer a beam along two axes, allowing enough depth to be reached to treat large prostates and eliminating two degrees of mechanical movement. Through computer simulation, it was determined that a curved 1.5-D configuration offered the optimal design. Two configurations were then proposed, and their ability to steer a beam within a target volume centered on the geometric focus of the transducer was simulated. An eight-element prototype was constructed to test the piezo-composite material and its electro-acoustical efficiency. Then, an array was constructed, and a multichannel amplifier and control system were added, to permit remote operation. Acoustical and electrical measurements were made to verify performance. Finally, the 1.5-D array was tested in vitro on samples of pig liver to confirm the ability to induce lesions.
引用
收藏
页码:231 / 242
页数:12
相关论文
共 23 条
[1]  
[Anonymous], 2012, ESSENTIALS MED ULTRA
[2]   CHARACTERISTICS OF A PVDF MEMBRANE HYDROPHONE FOR USE IN THE RANGE 1-100 MHZ [J].
BACON, DR .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1982, 29 (01) :18-25
[3]   ANALYSIS OF A MULTIELEMENT ULTRASOUND HYPERTHERMIA APPLICATOR [J].
BENKESER, PJ ;
FRIZZELL, LA ;
GOSS, SA ;
CAIN, CA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1989, 36 (03) :319-325
[4]   DESIGN AND EXPERIMENTAL EVALUATION OF ARM INTRACAVITARY ULTRASOUND PHASED-ARRAY SYSTEM FOR HYPERTHERMIA [J].
BUCHANAN, MT ;
HYNYNEN, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (12) :1178-1187
[5]  
Cathignol D, 1997, ACUSTICA, V83, P410
[6]  
Chapelon J Y, 1999, Eur J Ultrasound, V9, P31, DOI 10.1016/S0929-8266(99)00005-1
[7]  
CHAPELON JY, 1992, EUR UROL, V22, P147
[8]  
CHAPELON JY, 1993, IEEE ULTR S, V2, P1211
[9]  
CURIEL L, 2000, IEEE ULTR S P, V2, P1444
[10]   A 256-element ultrasonic phased array system for the treatment of large volumes of deep seated tissue [J].
Daum, DR ;
Hynynen, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (05) :1254-1268