Thermoelectrically cooled quantum-cascade-laser-based sensor for the continuous monitoring of ambient atmospheric carbon monoxide

被引:63
作者
Kosterev, AA
Tittel, FK
Köhler, R
Gmachl, C
Capasso, F
Sivco, DL
Cho, AY
Wehe, S
Allen, MG
机构
[1] Rice Univ, Rice Quantum Inst, Houston, TX 77251 USA
[2] Bell Labs, Lucent Technol, Murray Hill, NJ 07974 USA
[3] Phys Sci Inc, Andover, MA 01810 USA
关键词
D O I
10.1364/AO.41.001169
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report the first application of a thermoelectrically cooled, distributed-feedback quantum-cascade laser for continuous spectroscopic monitoring of CO in ambient air at a wavelength of 4.6 mum. A noise-equivalent detection limit of 12 parts per billion was demonstrated experimentally with a 102-cm optical pathlength and a 2.5-min data acquisition time at a 10-kHz pulsed-laser repetition rate. This sensitivity corresponds to a standard error in fractional absorbance of 3 x 10(-5). (C) 2002 Optical Society of America.
引用
收藏
页码:1169 / 1173
页数:5
相关论文
共 10 条
  • [1] New frontiers in quantum cascade lasers and applications
    Capasso, F
    Gmachl, C
    Paiella, R
    Tredicucci, A
    Hutchinson, AL
    Sivco, DL
    Baillargeon, JN
    Cho, AY
    Liu, HC
    [J]. IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2000, 6 (06) : 931 - 947
  • [2] DLUGOKENCKY E, NATL OCEANIC ATMOSPH
  • [3] Gmachl C, 2000, IEEE CIRCUITS DEVICE, V16, P10, DOI 10.1109/101.845908
  • [4] Single-mode tunable, pulsed, and continuous wave quantum-cascade distributed feedback lasers at λ congruent 4.6-4.7 μm
    Kohler, Rudeger
    Gmachl, Claire
    Tredicucci, Alessandro
    Capasso, Federico
    Sivco, Deborah L.
    Chu, S.N. George
    Cho, Alfred Y.
    [J]. Applied Physics Letters, 2000, 76 (09)
  • [5] Trace-gas detection in ambient air with a thermoelectrically cooled, pulsed quantum-cascade distributed feedback laser
    Kosterev, AA
    Tittel, FK
    Gmachl, C
    Capasso, F
    Sivco, DL
    Baillargeon, JN
    Hutchinson, AL
    Cho, AY
    [J]. APPLIED OPTICS, 2000, 39 (36) : 6866 - 6872
  • [6] Transportable automated ammonia sensor based on a pulsed thermoelectrically cooled quantum-cascade distributed feedback laser
    Kosterev, AA
    Curl, RF
    Tittel, FK
    Köhler, R
    Gmachl, C
    Capasso, F
    Sivco, DL
    Cho, AY
    [J]. APPLIED OPTICS, 2002, 41 (03) : 573 - 578
  • [7] Cavity ringdown spectroscopic detection of nitric oxide with a continuous-wave quantum-cascade laser
    Kosterev, AA
    Malinovsky, AL
    Tittel, FK
    Gmachl, C
    Capasso, F
    Sivco, DL
    Baillargeon, JN
    Hutchinson, AL
    Cho, AY
    [J]. APPLIED OPTICS, 2001, 40 (30) : 5522 - 5529
  • [8] Sensitive absorption spectroscopy with a room-temperature distributed-feedback quantum-cascade laser
    Namjou, K
    Cai, S
    Whittaker, EA
    Faist, J
    Gmachl, C
    Capasso, F
    Sivco, DL
    Cho, AY
    [J]. OPTICS LETTERS, 1998, 23 (03) : 219 - 221
  • [9] The HITRAN molecular spectroscopic database and HAWKS (HITRAN Atmospheric Workstation): 1996 edition
    Rothman, LS
    Rinsland, CP
    Goldman, A
    Massie, ST
    Edwards, DP
    Flaud, JM
    Perrin, A
    Camy-Peyret, C
    Dana, V
    Mandin, JY
    Schroeder, J
    McCann, A
    Gamache, RR
    Wattson, RB
    Yoshino, K
    Chance, KV
    Jucks, KW
    Brown, LR
    Nemtchinov, V
    Varanasi, P
    [J]. JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 1998, 60 (05) : 665 - 710
  • [10] Room-temperature mid-infrared laser sensor for trace gas detection
    Topfer, T
    Petrov, KP
    Mine, Y
    Jundt, D
    Curl, RF
    Tittel, FK
    [J]. APPLIED OPTICS, 1997, 36 (30): : 8042 - 8049